Skip to content

INFINITI SRD5217Recalls & reported problems

Cordis Corporation · 4 F INF 0.038 INCHES 125CM JL4 · Class 2

All models of INFINITI (22)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705032048019), received between Jun 23, 2025 and Jun 23, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Degraded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Degraded1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2025Degraded; Degraded

    AS REPORTED, A 4F 125 CM JUDKINS LEFT (JL4) INFINITI DIAGNOSTIC CATHETER DETERIORATED NEAR THE HUB OF THE CATHETER WHILE STILL IN STERILE PACKAGING. IT IS POSSIBLE THAT UV LIGHT EXPOSURE OCCURRED THROUGH THE CRACK IN THE CABINET. THE ONLY CATHETERS AFFECTED WERE THOSE HANGING ON THE MIDDLE RACK, WHICH WAS ALIGNED WITH THE CABINET OPENING. THERE WAS NO REPORTED PATIENT INJURY. THE DEVICE WAS STORED IN A CABINET WITH ULTRAVIOLET (UV) PROTECTION. THERE WAS NO VISIBLE PACKAGING DAMAGE. NO PHOTOGRAPHS ARE AVAILABLE. THE DEVICE WAS RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22306677 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page