TEMPO SRD6892Recalls & reported problems

Cordis Corporation · 5 F TEMPO 0.038 INCHES 125CM VERT135 · Class 2

All models of TEMPO (56)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032051774), received between Sep 19, 2017 and Mar 28, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Detachment Of Device Component1
  • Material Puncture/Hole1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2025Crack; Material Puncture/Hole; Crack; Material Puncture/Hole

    AS REPORTED, THERE WERE MULTIPLE HOLES THROUGHOUT THE LENGTH OF A 5F 125CM 135-DEGREE VERTEBRAL (VERT) TEMPO DIAGNOSTIC CATHETER. THE DAMAGE TO THE CATHETER WAS OBSERVED DURING A CEREBRAL ANGIOGRAM, AND THE CATHETER WAS IN THE RIGHT M2 VESSEL WHEN THE DAMAGE WAS OBSERVED. THERE WERE NO REPORTED INJURIES TO THE PATIENT. THE DEVICE WAS STORED AND PREPPED PER THE INSTRUCTIONS FOR USE (IFU) AND THERE WAS NO DAMAGE TO THE DEVICE PACKAGING. THE VESSEL CHARACTERISTICS WERE NON-CALCIFIED, NON-TORTUOUS, AND HAD 0% STENOSIS. THE DEVICE WAS NEVER PLACED IN AN ACUTE BEND DURING ITS USE IN THE PATIENT, AND THERE WAS NO DIFFICULTY REMOVING THE DEVICE FROM THE PATIENT. THE DEVICE HAS NOW BEEN RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21720353 ↗
  • MalfunctionSep 19, 2017Detachment Of Device Component; Detachment Of Device Component

    DURING A STROKE INTERVENTION CASE, A 5F TEMPO VERTEBRAL 135 DEGREE CATHETER WAS BEING PREPPED. THE CATHETER WAS PLACED INSIDE A NON-CORDIS GUIDE CATHETER AND THE MD NOTICED THAT CATHETER RADIOPAQUE TIP WAS NON EXISTENT AND REMOVED CATHETER OUT OF GUIDE CATHETER. BY PLACING A BRAND NEW VERTEBRAL CATHETER, MD FLUORO¿D PRIOR TO TREATMENT AND NOTICED THE BROKEN TIP WAS INTACT AND INSIDE THE GUIDE CATHETER. HE THEN CONTINUED TO REMOVE BOTH CATHETERS OUT SUCCESSFULLY WITHOUT PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6875693 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →