STABILIZER 527300ERecalls & reported problems

Cordis Corporation · SGW .014 STABILIZER 300CM · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705032056410), received between May 2, 2017 and Mar 31, 2024 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment Of Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment Of Device Component1
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 31, 2024Fracture; Fracture

    AS REPORTED, WHEN THE .014 300CM STABILIZER WAS USED WITH OUTBACK CATHETER, IT WASN'T GOING THROUGH AS SMOOTHLY. THE RADIOLOGIST FELT A LOT OF RESISTANCE. WHEN TAKING THE WIRE OUT, THE DISTAL TIP CAME OFF GOT STUCK INSIDE THE PATIENT. A SNARE WAS THEN USED TO REMOVE IT SUCCESSFULLY. THE DEVICE WILL BE RETURNED FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19009318 ↗
  • InjuryMay 2, 2017Detachment Of Device Component; Detachment Of Device Component

    CUSTOMER REPORTS DURING PERIPHERAL INTERVENTION, A PIECE OF THE DISTAL TIP OF A .014" STABILIZER XS WIRE SHEARED OFF INSIDE THE PATIENT'S DISTAL PERONEAL ARTERY. THE PHYSICIAN USED A 2 X 40 MM SABER TO DILATE THE LESION AND SECURE THE PIECE OF THE WIRE IN PLACE. PATIENT WAS TREATED WITH ANGIOPLASTY BALLOONS TO SEE IF THE WIRE TIP WOULD MIGRATE, BUT OTHERWISE WAS FURTHER IMPLANTED INTO THE INTIMA OF THE ARTERY. THERE WAS DIFFICULTY TRACKING THE WIRE THROUGH THE VESSEL AND LESION AS THERE WAS CALCIUM IN THE TARGET LESION THAT THE WIRE WAS UNABLE TO SUCCESSFULLY NEGOTIATE. THERE WERE NO PATIENT INJURY. THE DISTAL TIP MUST HAVE KINKED WHEN SEVERAL ATTEMPTS TO ADVANCE PAST THE LESION WERE MADE, AND AGAIN WHEN SUBSEQUENT UNSUCCESSFUL WITHDRAWAL ATTEMPTS WERE MADE. THERE WAS A FOCAL STENOSIS APPROXIMATELY 20 MM IN LENGTH IN THE DISTAL PERONEAL ARTERY THAT WAS CALCIFIED. THE PHYSICIAN BELIEVES THAT WHEN HE WAS ATTEMPTING TO ADVANCE THE WIRE BEYOND THE LESION, THE WIRE MAY HAVE TRAVELED SUBINTIMALLY AND PROLAPSED SLIGHTLY, CAUSING A KINK THAT CAUGHT ON CALCIUM IN THE SUBINTIMAL LAYER. WHEN HE ATTEMPTED TO WITHDRAW AND ADVANCE THE WIRE, IT WAS STUCK AND THE DISTAL TIP, APPROXIMATELY 15 MM IN

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 6536511 ↗

Other Wire, guide, catheter devices