EXPEDIUM 179702635Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM MONOAXIAL SCREW 5.5 x 6 x 35mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034064864), received between Nov 21, 2017 and Nov 9, 2022 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
- Loose or Intermittent Connection1
Patient problems
- Failure of Implant1
- Injury1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 9, 2022Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A SPINAL FUSION FOR NEUROMUSCULAR SCOLIOSIS ON (B)(6) 2022. AFTER TEMPORARY FIXATION WITH SET SCREWS, THE PROCESS OF TEMPORARILY TIGHTENING AND LOOSENING WAS REPEATED IN ORDER TO APPLY COMPRESSION AND DISTRACTION OPERATIONS FOR ROD INSERTION. ALTHOUGH THE SURGEON FIXED THE ROD AND THE SET SCREW, THEY LOOSENED. THE SCREW HEAD MOVED ON THE ROD. THE SAME OPERATION WAS REPEATED SEVERAL TIMES. THEN, IT WAS CONFIRMED THAT THE THREADS INSIDE THE SCREW HEAD WERE BROKEN. THE SURGEON USED A REPLACEMENT SCREW, AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS A 30-MINUTE DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM MONOAXIAL SCREW 5.5 X 6 X 35MM. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15761527 ↗InjuryNov 21, 2017Loose or Intermittent Connection; Loose or Intermittent Connection
ON (B)(6) 2016, SURGERY FOR SCOLIOSIS WAS PERFORMED USING THE EXPEDIUM SYSTEM. THE FIXED AREA WAS T11 ¿ L4. RECENTLY, SURGEON CONFIRMED THROUGH AN X-RAY THAT SET SCREWS (PART#: UNK) FIXED AT CAUDAL END WERE LOOSENED AND THAT RODS HAD MOVED IN UPPER POSITION, RESULTING IN LOSS OF THE CORRECTIVE FORCE. THE SURGEON DECIDED TO CONDUCT RE-OPERATION NEXT WEEK BECAUSE SUCH LOOSENED SET SCREWS MIGHT CONTACT THE HEART OF THE PATIENT. HE WILL USE NON-J&J REPLACEMENTS AT THE RE-OPERATION SCHEDULED NEXT WEEK.
Patient problems: Failure of Implant; Injury
MAUDE report 7049012 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.