Skip to content

EXPEDIUM 179712555Recalls & reported problems

Medos International Sàrl · EXPEDIUM SPINE SYSTEM SI POLYAXIAL SCREW 5.5 x 5 x 55mm · Class 2 · Implant

All models of EXPEDIUM (291)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034067582), received between Oct 21, 2022 and Oct 21, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 21, 2022Break; Break

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THE PATIENT WAS OPERATED 2 YEARS AGO WITH LUMBAR SPINE FUSION (USING 7 PEDICLE SCREWS). PATIENT CAME BACK TO THE NEUROSURGEON A MONTH AGO WITH A LOT OF PAIN. X-RAY TAKEN ON AN UNKNOWN DATE REVEALED 3 BROKEN SCREWS. SURGEON PERFORMED REVISION SURGERY ON (B)(6) 2022 AND REPLACED 3 BROKEN SCREWS WITH A BIGGER DIAMETER OF 6.0 MM SCREW. ALSO HE CHANGED ANOTHER ONE SCREW OF 5.0 MM WITH A 6.0 MM SCREW. FINALLY HE DECIDED TO ADD 5 MORE SCREWS 4 AT THE THORACIC LEVELS AND ONE SCREW AT THE SACRUM SO AS TO EXPAND THE 2 LEVEL FUSION TO AN ADULT SCOLIOSIS OPERATION. THERE WAS A SURGICAL DELAY OF UNSPECIFIED AMOUNT AS SURGEON HAD TO USE DRILL AND KERISSON TOOL TO REMOVE BROKEN SCREWS. THIS REPORT IS FOR ONE (1) SI POLYAXL SCREW 5 X 50MM. THIS IS REPORT 3 OF 3 FOR COMPLAINT (B)(4).

    Patient problems: Pain

    MAUDE report 15653933 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources and updates