EXPEDIUM 179712899Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM POLYAXIAL SCREW 5.5 x 8 x 100mm · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034068183), received between Nov 2, 2020 and Nov 2, 2020 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
8 other recalls cover devices cleared under the same FDA submission (K101993); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Code Available1
- Nonunion/Delayed-union Bone Fracture1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryNov 2, 2020Break; Break
IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT HAD HARDWARE FROM T5-ILIUM. THERE WAS NO SCREW IN L1 AND AN EXTENSION CONNECTOR CONNECTING THE UPPER AND LOWER ROD. THE SURGEON THOUGHT THAT THE PREVIOUS SURGERY WAS ON AN UNKNOWN DAY IN (B)(6) 2019. THE SURGEON SAID THAT THE PATIENT HAS A NON UNION AT L5-S1. AT AN UNKNOWN POINT IN TIME THE PATIENT BROKE THE FOLLOWING HARDWARE: LEFT ILIAC SCREW, RIGHT ILIAC SCREW, LEFT S1 SCREW. THESE SCREWS WERE ALL BROKEN RIGHT AT THE BASE OF THE SCREW HEADS. THE SURGEON PERFORMED THE REVISION ON (B)(6) 2020 AND REMOVED THE CAUDAL RODS THAT WERE IN THE EXTENSION CONNECTOR AND THEN REPLACED ALL THE SCREWS FROM L2-ILIUM. THE DEVICES WERE ALL UNKNOWN SIZES EXCEPT THE BROKEN ONES DESCRIBED ABOVE. THE SURGEON WAS ALSO ABLE TO RETRIEVE AND REMOVE THE BROKEN SCREW SHAFTS IN THE ILIAC SCREWS AND LEFT S AND REPLACED ALL THE L2-S1 SCREWS WITH EXPEDIUM SCREWS AND REPLACED THE ILIAC SCREWS WITH VIPER SAI FULLY THREADED SCREWS. THE SURGEON THEN CONTOURED NEW RODS AND PLACED NEW EXTENSION CONNECTORS ON THE CRANIAL ROD AND PLACED AND TIGHTENED THE NEW CAUDAL RODS ONTO THE EXTENSION CONNECTOR AND CONNECTED ALL THE L2-ILIUM SCREWS TO THE NEW EXPEDIUM / SAI SCREWS.
Patient problems: No Code Available; Nonunion/Delayed-union Bone Fracture
MAUDE report 10773410 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K101993. One submission often covers many models, so check the product and lot details in each record.
- Z-0707-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.
- Z-0712-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
- Z-0708-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X70MM Tl Product Code: 1867-15-970 Spinal Implant Component.
- Z-0713-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.
- Z-0709-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.
- Z-0706-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.
- Z-0705-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
- Z-0710-2013 · Nov 7, 2012 · TerminatedDePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page