Medos International Sàrl · EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 x 7.0 x 45mm · Class 2 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034074047), received between Jun 29, 2020 and Dec 27, 2023 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Foreign Body In Patient1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 27, 2023Break; Break

    IT WAS REPORTED THAT THE PATIENT HAD AN L4-5 TLIF (TRANSFORAMINAL LUMBAR INTERBODY FUSION) ON (B)(6) 2022 WITH THE EXPEDIUM SCREWS AND A NON-SYNTHES EXPANDABLE CAGE. UPON POST-OP FOLLOWUP ON AUGUST 2, 2023, THE SURGEON DISCOVERED A BROKEN SCREW AT L5 RIGHT. A REVISION SURGERY WAS SCHEDULED FOR (B)(6) 2024 FOR IMPLANT REMOVAL AND CONSTRUCT EXTENSION. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 X 7.0 X 45MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18399543 ↗
  • MalfunctionSep 15, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN CHINA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, WHEN SCREWING THE PEDICLE SCREW DURING SURGERY, THE SCREW ROD SEPARATED FROM THE LONG TAIL OF SCREW CAP. THE SURGEON REMOVED THE DETACHED SCREW ROD AND CAP FROM THE PATIENT, AND CHANGED TO ANOTHER DEVICE TO COMPLETE THE SURGERY. THIS CAUSED THE SURGERY TO BE EXTENDED FOR ABOUT 1.5 HOURS. THE PATIENT IS REPORTED AS STABLE. THIS REPORT INVOLVES ONE EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 X 7.0 X 45MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15427188 ↗
  • InjuryAug 1, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, THE PATIENT UNDERWENT A REVISION PROCEDURE, LUMBAR SPINAL FUSION. AFTER REMOVING A SCREW, THE SURGEON MOVED ON TO INSERTING A CFS 8MM SCREW WITH THE SCREWDRIVER. THEN, THE SCREWDRIVER¿S TIP BROKE OFF. THE FRAGMENTED TIP WAS IMMOVABLE IN THE SCREW¿S SHANK AND HAS BEEN LEFT IN THE LESION. THIS REPORT INVOLVES ONE (1) EXPEDIUM SPINE SYSTEM SI POLYAXIAL EXTEND TAB 5.5 X 7.0 X 45MM. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: Foreign Body In Patient

    MAUDE report 15140685 ↗
  • MalfunctionJun 29, 2020Material Deformation; Material Deformation

    THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10205691 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices