EXPEDIUM 179799745Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM DI FAVORED ANGLE REDUCTION SCREW 5.5 x 7 x 45mm · Class 2
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034083827), received between Sep 27, 2018 and Sep 27, 2018 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- Spinal Column Injury2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjurySep 27, 2018Break; Break
THE FOLLOWING WAS REPORTED: PRODUCT CODE: 179799745. LOT/BATCH/EXP: ATMC7K AND ARKBLR. DID THE EVENT HAPPEN DURING A PROCEDURE? NO. WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? YES. WAS THE PRODUCT BEING USED IN A CLINICAL TRIAL? NO. EVENT OUTCOME/HOW WAS IT MANAGED? THE SCREWS WERE SUCCESSFULLY REMOVED AND REVISED. WAS THERE A PATIENT IMPACT OR WAS THE PROCEDURE EXTENDED GREATER THAN 30 MINUTES DUE TO THE FAILURE? NO. HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. IS THE PRODUCT AVAILABLE FOR RETURN? IF THE ITEM IS RETURNED FROM CSSD FOLLOWING DECOM PLEASE GIVE A DETAILED EXPLANATION OF THE EVENT: THESE SCREWS WERE IMPLANTED AT S1 IN A CONSTRUCT SPANNING L3-S1 WITH CONCORDE CAGES AT L3/4, L4/5 AND L5/S1. BOTH S1 SCREWS BROKE DUE TO THE BREAKAGE AND THE PSEUDOARTHROSIS AT L5/S1 A REVISION SURGERY WAS REQUIRED. THE SCREWS WERE SUCCESSFULLY EXTRACTED (ALL FRAGMENTS) AND REPLACED WITH EXPREDIUM VERSE SCREWS AND THE FUSION EXTENDED TO THE PELVIS. THE CAGE AT L5/S1 WAS ALSO REVISED.
Patient problems: Spinal Column Injury
MAUDE report 7914463 ↗InjurySep 27, 2018Break; Break
THE FOLLOWING WAS REPORTED: PRODUCT CODE : 179799745. LOT/BATCH/EXP: ATMC7K AND ARKBLR . DID THE EVENT HAPPEN DURING A PROCEDURE? NO. WERE YOU IN THE PROCEDURE AT THE TIME OF THE EVENT? YES. WAS THE PRODUCT BEING USED IN A CLINICAL TRIAL? NO. EVENT OUTCOME/HOW WAS IT MANAGED? THE SCREWS WERE SUCCESSFULLY REMOVED AND REVISED. WAS THERE A PATIENT IMPACT OR WAS THE PROCEDURE EXTENDED GREATER THAN 30 MINUTES DUE TO THE FAILURE? NO. HAS THE REPORTER FACILITY INDICATED THERE MAY BE LEGAL ACTION? NO. IS THE PRODUCT AVAILABLE FOR RETURN? IF THE ITEM IS RETURNED FROM CSSD FOLLOWING DECOM PLEASE GIVE A DETAILED EXPLANATION OF THE EVENT: THESE SCREWS WERE IMPLANTED AT S1 IN A CONSTRUCT SPANNING L3-S1 WITH CONCORDE CAGES AT L3/4, L4/5 AND L5/S1. BOTH S1 SCREWS BROKE DUE TO THE BREAKAGE AND THE PSEUDOARTHROSIS AT L5/S1 A REVISION SURGERY WAS REQUIRED. THE SCREWS WERE SUCCESSFULLY EXTRACTED (ALL FRAGMENTS) AND REPLACED WITH EXPREDIUM VERSE SCREWS AND THE FUSION EXTENDED TO THE PELVIS. THE CAGE AT L5/S1 WAS ALSO REVISED.
Patient problems: Spinal Column Injury
MAUDE report 7914269 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.