EXPEDIUM 186100001Recalls & reported problems
Depuy Spine, LLC · EXPEDIUM SPINE SYSTEM SINGLE INNIE SET SCREW 4.5 · Class 2 · Implant
All models of EXPEDIUM (29)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 13
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 10705034103846), received between Oct 16, 2017 and Mar 10, 2020 — classified as 13 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (8 reports).
1 other recall cover devices cleared under the same FDA submission (K071495, K111136); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn8
- Torn Material3
- Loose or Intermittent Connection2
- Material Twisted/Bent2
- Break1
Patient problems
- No Known Impact Or Consequence To Patient9
- No Code Available3
- No Consequences Or Impact To Patient2
- Local Reaction1
- Not Applicable1
- Spinal Column Injury1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryMar 10, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON (B)(6) 2020, DURING GROWING ROD LENGTHENING PROCEDURE, AFTER EXPOSURE OF THE ORIGINAL EXISTING HARDWARE, THE SURGEON DETERMINED THAT THE INTERFACE BETWEEN THE WEDDING BAND CONNECTOR AND THE ROD WAS LOOSE. THE SURGEON COULD IMMEDIATELY TELL THAT THE SINGLE INNIE SET SCREW LOCKED INTO THE CONNECTOR ON THE OPEN SIDE OF THE 4.5 / 4.5 CONNECTOR WAS STRIPPED AND CROSS THREADED. HE THEN ATTEMPTED TO IMPLANT A NEW SINGLE INNIE SET SCREW WHICH IMMEDIATELY CROSS THREADED, REMOVED THAT SET SCREW, AND ALSO THE ORIGINAL 4.5 / 4.5 CONNECTOR. THEN USED IN-SITU BENDERS TO ADJUST THE ROD AND IMPLANTED A NEW CONNECTOR AND DETERMINED THAT THE ORIGINAL WAS DAMAGED. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. THERE WERE NO FRAGMENTS GENERATED. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (PART#: UNKNOWN, LOT#: UNKNOWN, QUANTITY UNKNOWN); EXP 4.5 TI OP/CLSD 4.5X4.5 (PART#: 186172345, LOT#: UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) EXP 4.5 TI SINGLE INNIE. THIS IS REPORT 1 OF 2 FOR (B)(4). THIS COMPLAINT WILL CAPTURE THE INTRA-OP EVENT (ATTEMPTED TO IMPLANT A NEW SINGLE INNIE SET SCREW WHICH IMMED
Patient problems: No Code Available
MAUDE report 9812368 ↗InjuryMar 10, 2020Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON (B)(6) 2020, DURING GROWING ROD LENGTHENING PROCEDURE, AFTER EXPOSURE OF THE ORIGINAL EXISTING HARDWARE, THE SURGEON DETERMINED THAT THE INTERFACE BETWEEN THE WEDDING BAND CONNECTOR AND THE ROD WAS LOOSE. THE SURGEON COULD IMMEDIATELY TELL THAT THE SINGLE INNIE SET SCREW LOCKED INTO THE CONNECTOR ON THE OPEN SIDE OF THE 4.5 / 4.5 CONNECTOR WAS STRIPPED AND CROSS THREADED. HE THEN ATTEMPTED TO IMPLANT A NEW SINGLE INNIE SET SCREW WHICH IMMEDIATELY CROSS THREADED, REMOVED THAT SET SCREW, AND ALSO THE ORIGINAL 4.5 / 4.5 CONNECTOR. THEN USED IN-SITU BENDERS TO ADJUST THE ROD AND IMPLANTED A NEW CONNECTOR AND DETERMINED THAT THE ORIGINAL WAS DAMAGED. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. THERE WERE NO FRAGMENTS GENERATED. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN); EXP 4.5 TI OP/CLSD 4.5X4.5 (PART# 186172345, LOT# UNKNOWN, QUANTITY 1). THIS REPORT IS FOR ONE (1) EXP 4.5 TI SINGLE INNIE. THIS IS REPORT 2 OF 2 FOR (B)(4). THIS COMPLAINT WILL CAPTURE THE INTRA-OP EVENT (ATTEMPTED TO IMPLANT A NEW SINGLE INNIE SET SCREW WHICH IMMEDIATE
Patient problems: No Code Available
MAUDE report 9812366 ↗InjuryAug 27, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING LENGTHENING OF PRIOR GROWING ROD CONSTRUCT, THE ROD WOULD NOT STAY IN THE OPEN SADDLE OF THE CONNECTOR BECAUSE THE SET SCREWS WOULD STRIP OUT AND FAIL. HE FELT THE COBALT CHROME (COCR) ROD NEEDED TO BE CONTOURED SO IT WOULD LAY IN THE CONNECTOR SADDLE WITHOUT RESISTANCE WHEN PLACING A SET SCREW. THERE WAS A 20-30 MINUTES SURGICAL DELAY, PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS PATIENT HARM/CONSEQUENCE. THIS COMPLAINT INVOLVES SIX (6) DEVICES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8938282 ↗InjuryAug 27, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING LENGTHENING OF PRIOR GROWING ROD CONSTRUCT, THE ROD WOULD NOT STAY IN THE OPEN SADDLE OF THE CONNECTOR BECAUSE THE SET SCREWS WOULD STRIP OUT AND FAIL. HE FELT THE COBALT CHROME (COCR) ROD NEEDED TO BE CONTOURED SO IT WOULD LAY IN THE CONNECTOR SADDLE WITHOUT RESISTANCE WHEN PLACING A SET SCREW. THERE WAS A 20-30 MINUTES SURGICAL DELAY, PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS PATIENT HARM/CONSEQUENCE. THIS COMPLAINT INVOLVES SIX (6) DEVICES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8938258 ↗InjuryAug 27, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING LENGTHENING OF PRIOR GROWING ROD CONSTRUCT, THE ROD WOULD NOT STAY IN THE OPEN SADDLE OF THE CONNECTOR BECAUSE THE SET SCREWS WOULD STRIP OUT AND FAIL. HE FELT THE COBALT CHROME (COCR) ROD NEEDED TO BE CONTOURED SO IT WOULD LAY IN THE CONNECTOR SADDLE WITHOUT RESISTANCE WHEN PLACING A SET SCREW. THERE WAS A 20-30 MINUTES SURGICAL DELAY, PROCEDURE OUTCOME WAS UNKNOWN. THERE WAS PATIENT HARM/CONSEQUENCE. THIS COMPLAINT INVOLVES SIX (6) DEVICES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8938209 ↗InjuryAug 12, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT ON (B)(6) 2019, DURING PROXIMAL CONSTRUCT AND DISTAL CONSTRUCT FOR GROWING ROD FINAL FUSION OF T2-L4, WHILE FINAL TIGHTENING, THE UNKNOWN SET SCREW'S THREAD SHEARED OFF. THIS HAPPENED WHILE NOT USING THE ANTI-TORQUE. ONCE THE SURGEON USED THE ANTI-TORQUE THE SET SCREW DID NOT SHEAR OFF. BUT IT HAPPENED THREE TIMES. THE PATIENT WAS EXPOSED TO ADDITIONAL ANESTHESIA FOR TEN (10) MINUTES. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. CONCOMITANT DEVICE REPORTED: POLYAXIAL SCREW (PART # 186112425, LOT# UNKNOWN QUANTITY #3). THIS COMPLAINT INVOLVES THREE (3) DEVICES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8885165 ↗InjuryAug 12, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT ON (B)(6) 2019, DURING PROXIMAL CONSTRUCT AND DISTAL CONSTRUCT FOR GROWING ROD FINAL FUSION OF T2-L4, WHILE FINAL TIGHTENING, THE UNKNOWN SET SCREW'S THREAD SHEARED OFF. THIS HAPPENED WHILE NOT USING THE ANTI-TORQUE. ONCE THE SURGEON USED THE ANTI-TORQUE THE SET SCREW DID NOT SHEAR OFF. BUT IT HAPPENED THREE TIMES. THE PATIENT WAS EXPOSED TO ADDITIONAL ANESTHESIA FOR TEN (10) MINUTES. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. CONCOMITANT DEVICE REPORTED: POLYAXIAL SCREW (PART # 186112425, LOT# UNKNOWN QUANTITY #3) THIS COMPLAINT INVOLVES THREE (3) DEVICES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8884919 ↗InjuryAug 12, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT ON (B)(6) 2019, DURING PROXIMAL CONSTRUCT AND DISTAL CONSTRUCT FOR GROWING ROD FINAL FUSION OF T2-L4, WHILE FINAL TIGHTENING, THE UNKNOWN SET SCREW'S THREAD SHEARED OFF. THIS HAPPENED WHILE NOT USING THE ANTI-TORQUE. ONCE THE SURGEON USED THE ANTI-TORQUE THE SET SCREW DID NOT SHEAR OFF. BUT IT HAPPENED THREE TIMES. THE PATIENT WAS EXPOSED TO ADDITIONAL ANESTHESIA FOR TEN (10) MINUTES. THE PROCEDURE AND PATIENT OUTCOME WERE UNKNOWN. CONCOMITANT DEVICE REPORTED: POLYAXIAL SCREW (PART # 186112425, LOT # UNKNOWN, QUANTITY # 3). THIS COMPLAINT INVOLVES THREE (3) DEVICES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8884845 ↗InjuryMay 22, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT 4.5 TITANIUM EXPEDIUM WITH COBALT CHROME RODS WERE USED. WHEN FINAL TIGHTENING 2 OF THE SET SCREWS WITH THE TORQUE LIMITING DRIVER, THE THREADS SHEARED OFF OF THE SET SCREWS AND THEY HAD TO BE REPLACED. THEY WERE ALREADY DOWN INTO THE SCREW HEAD WHEN THE FINAL TORQUE WAS APPLIED. NO ADVERSE CONSEQUENCE TO THE PATIENT OR ADDITIONAL ANESTHESIA NEEDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8631727 ↗InjuryMay 22, 2019Material Split, Cut or Torn; Material Split, Cut or Torn
IT WAS REPORTED THAT 4.5 TITANIUM EXPEDIUM WITH COBALT CHROME RODS WERE USED. WHEN FINAL TIGHTENING 2 OF THE SET SCREWS WITH THE TORQUE LIMITING DRIVER, THE THREADS SHEARED OFF OF THE SET SCREWS AND THEY HAD TO BE REPLACED. THEY WERE ALREADY DOWN INTO THE SCREW HEAD WHEN THE FINAL TORQUE WAS APPLIED. NO ADVERSE CONSEQUENCE TO THE PATIENT OR ADDITIONAL ANESTHESIA NEEDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8631691 ↗InjuryJan 3, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT DR. (B)(6) DID THE PRIMARY FUSION ON OCTOBER 1ST. THIS FUSION INVOLVED EXPEDIUM 4.5 IN COMBINATION WITH THE SPECIAL ACCESS SYSTEM TROLLEY. DR. (B)(6) MENTIONED THAT HE OVERCORRECTED THE SCOLIOSIS AND THERE WAS A PROXIMAL PULLOUT OF THE CONSTRUCT. BASICALLY THE FIRST SET SCREW ON THE LEFT SIDE OF THE CONSTRUCT AT T3 COMPLETELY CAME OFF AND CONSEQUENTLY THE SCREW AT T4 COMPLETELY PULLED OUT OF THE PEDICLE AS THERE WAS TREMENDOUS STRESS ON THAT SCREW. THIS PARTICULAR SCREW WAS REPLACED WITH A NEW 4.35X30MM AND A NEW ROD AND NEW SET SCREW WAS PLACE ON THE FIRST SCREW. THE PATIENT WAS AFFECTED BY THIS REVISION, INFO IS BELOW. NO REPLACEMENT IS NEEDED. I HAVE NOTHING FURTHER TO ADD. PATIENT CONSEQUENCE? YES. PATIENT CONSEQUENCE DESCRIPTION:THE PATIENT NEEDED REVISION SURGERY IN ORDER TO CORRECT THE POP OFF CAP AND PULLOUT SCREW. ACTION TAKEN FOR PROCEDURE: REVISION SURGERY. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.
Patient problems: No Code Available
MAUDE report 8212469 ↗InjuryNov 28, 2018Loose or Intermittent Connection; Loose or Intermittent Connection
PER REPORTER, MR (B)(6) WAS PERFORMING A DISTRACTION OF GROWING ROD PROCEDURE USING THE CONNECTOR, HE NOTICED THAT THE SET SCREW IN THE CONNECTOR WAS NOT FLUSH AND UPON REMOVING THE SET SCREW AND THE CONNECTOR HE NOTICED SOME METALLOSIS IN THE PATIENTS TISSUE. HE REPLACED THE CONNECTOR AND DISTRACTED THE RODS, WITH NO ISSUES.
Patient problems: Local Reaction
MAUDE report 8111012 ↗InjuryJun 14, 2018Break; Break
THE 4.5 TI EXPEDIUM SET SCREW STRIPPED POST OP IN PATIENT. A REVISION TOOK PLACE AND BROKEN SET SCREW WAS REMOVED AND REPACKED WIT HA DIFFERENCE CONSTRUCT. ADDITIONAL INFORMATION RECEIVED FROM REP 23MAY2018: SET SCREW LOOSENED POST -OP. IT WAS DISCOVERED BECAUSE PATIENT IS A CHILD WITH GROWING RODS AND THEY DID X-RAY AT A LATER DATE. PATIENT CONSEQUENCE? NO. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.
Patient problems: Not Applicable; Spinal Column Injury
MAUDE report 7601946 ↗MalfunctionNov 8, 2017Torn Material; Torn Material
THE SURGEON DISTRACTED THE GROWING ROD CONSTRUCT AND DURING FINAL TIGHTENING, THE SINGLE INNIE SETSCREW STRIPPED OUT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7008963 ↗MalfunctionOct 16, 2017Torn Material; Torn Material
SURGEON HAD INITIALLY TIGHTENED THE SET SCREWS WITH STANDARD SET SCREW DRIVER. WHEN GOING BACK TO FINAL TIGHTEN, THE SET SCREW DID NOT SEEM TO TIGHTEN/SEAT PROPERLY. THE SURGEON THOUGHT THE TORQUE DRIVER SLIPPED, BUT IT APPEARED THAT THE SET SCREW LOOKED BROKEN. AFTER BACKING OUT THE SET SCREW, IT WAS CLEAR THAT THE PART OF THE THREADS HAD BROKEN OFF IN THE PROCESS OF TIGHTENING. THIS OCCURRED ON 2 SET SCREWS IN THE CONSTRUCT DURING FINAL TIGHTENING AND IT IS UNCLEAR WHETHER ANYTHING WAS CROSS-THREADED OR WHAT MAY HAVE BEEN THE PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6952359 ↗MalfunctionOct 16, 2017Torn Material; Torn Material
SURGEON HAD INITIALLY TIGHTENED THE SET SCREWS WITH STANDARD SET SCREW DRIVER. WHEN GOING BACK TO FINAL TIGHTEN, THE SET SCREW DID NOT SEEM TO TIGHTEN/SEAT PROPERLY. THE SURGEON THOUGHT THE TORQUE DRIVER SLIPPED, BUT IT APPEARED THAT THE SET SCREW LOOKED BROKEN. AFTER BACKING OUT THE SET SCREW, IT WAS CLEAR THAT THE PART OF THE THREADS HAD BROKEN OFF IN THE PROCESS OF TIGHTENING. THIS OCCURRED ON 2 SET SCREWS IN THE CONSTRUCT DURING FINAL TIGHTENING AND IT IS UNCLEAR WHETHER ANYTHING WAS CROSS-THREADED OR WHAT MAY HAVE BEEN THE PROBLEM.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6952239 ↗
FDA clearance and classification
Cleared through 510(k) K111136 on Jun 14, 2012. Submitted by Depuy Spine, Inc. as “Expedium Spine System, Viper System, Viper 2 System”. The FDA review took 419 days. FDA record ↗
Cleared through Special 510(k) K071495 on Jun 29, 2007 (decision: Substantially Equivalent). Submitted by Depuy Spine, A Johnson & Johnson Company as “Expedium 4.5mm Spine System”. The FDA review took 29 days. FDA record ↗
The FDA regulates product code KWP (appliance, fixation, spinal interlaminal) under 21 CFR 888.3050 as a Class II device. 21 CFR 888.3050 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K071495, K111136. One submission often covers many models, so check the product and lot details in each record.
- Z-1674-2015 · Apr 21, 2015 · TerminatedEXPEDIUM Spine System/Synapse System; MODEL: EXPEDIUM Offset Conical Lock Nut 8mm, CATALOG NO. : 1754-91-150, UDI NUMBER (if applicable): GTIN: 10705034032467
Common questions about the EXPEDIUM 186100001
Has the EXPEDIUM 186100001 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034103846), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.
What problems have been reported with the EXPEDIUM 186100001?
The FDA's MAUDE database lists 16 adverse event reports for this device, received between Oct 16, 2017 and Mar 10, 2020. In those reports, the most common specific device problems are “Material Split, Cut or Torn” (8), “Torn Material” (3) and “Loose or Intermittent Connection” (2); the most common patient problems are “Local Reaction” (1), “Not Applicable” (1) and “Spinal Column Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the EXPEDIUM 186100001?
None of the reports are classified as deaths, 13 as injuries and 3 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the EXPEDIUM 186100001 FDA approved?
It was cleared through Special 510(k) K071495 on Jun 29, 2007; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the EXPEDIUM 186100001 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the EXPEDIUM 186100001, and is it still sold?
The FDA's device database lists Depuy Spine, LLC as the company responsible for it (brand EXPEDIUM, model 186100001). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL433 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page