Skip to content

VIPER 186715850Recalls & reported problems

Medos International Sàrl · VIPER SYSTEM POLYAXIAL SCREW 5.5 x 8 x 50mm · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034122755), received between Apr 16, 2018 and Jan 5, 2024 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

8 other recalls cover devices cleared under the same FDA submission (K101993); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Adverse Event Without Identified Device or Use Problem1
  • Device-Device Incompatibility1
  • Disassembly1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Joint Laxity1
  • No Consequences Or Impact To Patient1
  • Not Applicable1
  • Osteolysis1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryJan 5, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: ADJACENT SEGMENT PATHOLOGY, INSTABILITY COMPLICATION TIME/INTERVENTION. REPEAT SURGERY THIS IS FOR DEPUY SYNTHES SPINE EXPEDIUM /OTHER AND MOSS MIAMI SI /OTHER AND VIPER PRODUCTS (VIPER POLY SCREW 8X50MM TI (B)(4).

    Patient problems: Joint Laxity; Osteolysis

    MAUDE report 18453447 ↗
  • MalfunctionMay 24, 2022Break; Break

    IT WAS REPORTED ON (B)(6) 2022, THE SURGEON WAS INSERTING THE SCREW INTO THE VERTEBRAL BODY WHEN THE HEAD OF THE SCREW BROKE OFF AND DISENGAGED. THE FELLOW PROCEEDED TO INSERT HIS SCREW AND THE SAME EXACT THING HAPPENED. IN BOTH INSTANCES, THE PEDICLES THEY WERE INSERTING THE SCREWS INTO HAD PREVIOUS HARDWARE THAT HAD BEEN REMOVED. THE SURGEON/FELLOW DID NOT TAP. THIS REPORT IS FOR ONE (1) VIPER POLY SCREW 8X50 MM TI. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14492019 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT DURING POSTERIOR SPINAL FUSION SURGERY ON (B)(6) 2021, THE SCREW HEAD ¿TULIP¿ PART OF SCREW POPPED OFF AS THE SURGEON WAS INSERTING THE SCREW. THE SURGEON WAS USING 8.0X50MM SCREW AT THE LEFT T11 PEDICLE. THE SURGEON USES CUSTOM AWLS TO ACCESS THE PEDICLE; DOES NOT TAP AND PROCEEDS DIRECTLY FROM THE AWL TO SCREW INSERTION. THERE WERE NO FRAGMENTS GENERATED. THE SURGEON REMOVED THE BROKEN SCREW REPLACED IT WITH A SMALLER DIAMETER SCREW. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO DELAY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT IS FOR A VIPER POLY SCREW. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11739725 ↗
  • MalfunctionApr 28, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021 DURING AN UNKNOWN SURGERY WHILE PLACING A SCREW AT THE L4 VERTEBRAL BODY, THE SURGEON WAS TURNING THE HANDLE AND ENCOUNTERED A HIGHER INSERTION TORQUE THAN NORMAL. THE SCREW HEAD THEN DISENGAGED FROM THE SCREW SHANK AND CAME OFF. IT WAS ADVISED THAT THE SURGEON TAP THE BONE PRIOR TO SCREW PLACEMENT, HOWEVER HE WAS SIMPLY REPLACING A SCREW THAT HAD BEEN REMOVED AT THAT SITE PREVIOUSLY AND DECIDED AGAINST TAPPING. NO FRAGMENTS WERE GENERATED. SCREW SHANK WAS REMOVED USING A DIFFERENT SCREWDRIVER. THERE WAS A SURGICAL DELAY OF TEN (10) MINUTES. THERE WERE NO PATIENT CONSEQUENCES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT IS FOR ONE (1) VIPER SYSTEM POLYAXIAL SCREW 5.5 X 8 X 50MM. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11737277 ↗
  • InjuryFeb 8, 2021Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS FURTHER REPORTED THAT THE SURGEON COMPLETED LAMINECTOMY REVISION SURGERY. REVISION WAS PERFORMED AS THE PATIENT NEEDED A PROCEDURE DONE AND THE IMPLANTS WERE IN THE WAY TO ACHIEVE THE SURGICAL GOAL FOR HIS PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11296329 ↗
  • MalfunctionApr 16, 2018Disassembly; Disassembly

    IT WAS REPORTED DOCTOR WAS IMPLANTING 8X50 VIPER SCREW. HE STOPPED TO TAKE FLUORO PICTURE TO CONFIRM ALL WAS OK WITH PLACEMENT. STARTED TO ADVANCE SCREW FURTHER DOWN AND SCREW HEAD POPPED OFF. THE RING BETWEEN SCREW HEAD AND SCREW SHANK WAS BROKEN. HAD TO RETRIEVE THOSE BROKEN METAL FRAGMENTS AND TAKE SCREW SHAFT OUT. FIFTEEN MINUTES DELAY REPORTED.

    Patient problems: No Consequences Or Impact To Patient; Not Applicable

    MAUDE report 7432336 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K101993. One submission often covers many models, so check the product and lot details in each record.

  • Z-0708-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 9 X70MM Tl Product Code: 1867-15-970 Spinal Implant Component.
  • Z-0710-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 9 X100MM Tl Product Code: 1867-15-999 Spinal Implant Component.
  • Z-0707-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 8 X100MM Tl Product Code: 1867-15-899 Spinal Implant Component.
  • Z-0709-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 9 X 80MM Tl Product Code: 1867-15-980 Spinal Implant Component.
  • Z-0712-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 10 X 80MM Tl Product Code: 1867-15-180 Spinal Implant Component.
  • Z-0705-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 8 X70MM Tl Product Code: 1867-15-870 Spinal Implant Component.
  • Z-0706-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 8 X80MM Tl Product Code: 1867-15-880 Spinal Implant Component.
  • Z-0713-2013 · Nov 7, 2012 · Terminated
    DePuy Spine VIPER POLY SCREW 5.5 10X100MM Tl Product Code: 1867-15-199 Spinal Implant Component.

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources and updates