VIPER 186720535Recalls & reported problems

Medos International Sàrl · VIPER SYSTEM CANNULATED UNIPLANAR SCREW 5.5 x 5.0 x 35mm · Class 2 · Implant

All models of VIPER (121)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 10705034125619), received between Apr 13, 2021 and Feb 11, 2022 — classified as 1 injury, 7 malfunctions.

The most frequently reported specific device problem is “Device Slipped” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Slipped7
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Failure of Implant1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 11, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, WHILE DOING THE FINAL LOCK, AFTER THE MANEUVER OVERRIDE, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS WAS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT/DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITHOUT ANY SURGICAL DELAY. NO PATIENT CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) VIPER UNI SCREW 5X35MM TI. THIS IS REPORT 3 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13517753 ↗
  • MalfunctionFeb 11, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, WHILE DOING THE FINAL LOCK, AFTER THE MANEUVER OVERRIDE, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS WAS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT/DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITHOUT ANY SURGICAL DELAY. NO PATIENT CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) VIPER UNI SCREW 5X35MM TI THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13517732 ↗
  • MalfunctionFeb 11, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING A PROCEDURE ON (B)(6) 2022, WHILE DOING THE FINAL LOCK, AFTER THE MANEUVER OVERRIDE, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS WAS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT/DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. THE SURGERY WAS SUCCESSFULLY COMPLETED WITHOUT ANY SURGICAL DELAY. NO PATIENT CONSEQUENCES REPORTED. THIS REPORT IS FOR ONE (1) VIPER UNI SCREW 5X35MM TI. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13517716 ↗
  • MalfunctionDec 15, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, WHILE DOING THE FINAL LOCK, AFTER THE DEROGATION MANEUVER, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT / DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN TORQUE DEVICE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) VIPER UNI SCREW 5 X 35MM TI. THIS REPORT IS 3 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13014268 ↗
  • MalfunctionDec 15, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, WHILE DOING THE FINAL LOCK, AFTER THE DEROGATION MANEUVER, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT / DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN TORQUE DEVICE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) VIPER UNI SCREW 5X35MM TI. THIS REPORT IS 4 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13014237 ↗
  • MalfunctionDec 15, 2021Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, WHILE DOING THE FINAL LOCK, AFTER THE DEROGATION MANEUVER, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT / DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN TORQUE DEVICE (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES FOUR (4) DEVICES. THIS REPORT IS FOR (1) VIPER UNI SCREW 5X35MM TI. THIS REPORT IS 1 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13013395 ↗
  • MalfunctionDec 7, 2021Device Slipped; Device Slipped

    IT WAS REPORTED THAT ON (B)(6) 2021, WHILE DOING THE FINAL LOCK, AFTER THE DEROGATION MANEUVER, THE SURGEON FOUND THAT THE ROD WAS NOT TIGHTLY BLOCKED INTO THE UNIPLANAR SCREW HEADS. THIS IN COMPARISON WITH THE UPPER AND LOWER POLY AXIAL SCREWS WITH THE FIXATION TO THE ROD WAS VERY SOLID. WITH THE UNIPLANAR SCREWS, ONE COULD CLEARLY SEE AN UNEXPECTED MEDIO-LATERAL MOVEMENT / DEGREE OF FREEDOM DESPITE AN AD-HOC TIGHTENING WITH THE TORQUE HANDLE. PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). UNKNOWN TORQUE HANDLE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) VIPER UNI SCREW 5X35MM TI. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12953849 ↗
  • InjuryApr 13, 2021Material Deformation; Material Deformation

    DEVICE REPORT FROM SYNTHES EUROPE REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT UNIPLANAR SCREWS WERE REMOVED FROM THE PATIENT THAT FOUND TO BE SLIGHTLY BENT AT COLLAR LEVEL, IT WAS PERPENDICULAR FROM THE NATURAL SCREWHEAD AXIS. IT WAS ALSO IMPORTANT TO NOTE: THE ABLATION OF THESE SCREWS HAS NO RELATIONSHIP, NOR REASONS, WITH THE BENDING ISSUE ANNOUNCED IN THIS COMPLAINT. THIS REPORT IS FOR ONE (1) VIPER SYSTEM CANNULATED UNIPLANAR SCREW 5.5 X 5.0 X 35MM. THIS REPORT IS 1 OF 2 FOR (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 11655791 ↗

FDA clearance and classification

The FDA regulates product code KWP (appliance, fixation, spinal interlaminal) under 21 CFR 888.3050 as a Class II device. 21 CFR 888.3050 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VIPER 186720535

Has the VIPER 186720535 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034125619), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VIPER 186720535?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Apr 13, 2021 and Feb 11, 2022. In those reports, the most common specific device problems are “Device Slipped” (7) and “Material Deformation” (1); the most common patient problem is “Failure of Implant” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VIPER 186720535?

None of the reports are classified as deaths, 1 as an injury and 7 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the VIPER 186720535 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the VIPER 186720535, and is it still sold?

The FDA's device database lists Medos International Sàrl as the company responsible for it (brand VIPER, model 186720535). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page