MOUNTAINEER 188362025Recalls & reported problems
Medos International Sàrl · MOUNTAINEER OCT SPINAL SYSTEM OCCIPITAL PLATE INNER SCREW 3.5 · Class 2 · Implant
All models of MOUNTAINEER (51)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034158853), received between Aug 4, 2017 and Mar 9, 2022 — classified as 2 injuries, 4 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Deformation4
- Migration2
Patient problems
- No Known Impact Or Consequence To Patient4
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 9, 2022Migration; Migration
IT WAS REPORTED THAT PATIENT UNDERWENT REVISION SURGERY OF A POSTERIOR SPINAL FUSION ON (B)(6) 2022. THE ORIGINAL PROCEDURE WAS DONE ON (B)(6) 2022. ON AN UNKNOWN DATE FOLLOWING THE INITIAL PROCEDURE, THE MOUNTAINEER OC PLATE, TWO (2) RODS, AND TWO (2) SET SCREWS POPPED OUT. BOTH RODS POPPED OUT BUT WERE STILL IN THE CERVICAL PORTION. IT WAS NOTED THAT THE PATIENT DID NOT WEAR A NECK BRACE ALL THE TIME AND THE NECK MAY HAVE BEEN HYPEREXTENDED IN THE ORIGINAL CASE. THE DOCTOR HAD ONLY PUT TWO SCREWS ON THE PLATE. THE ANGLE WAS BAD, AND THE THIRD SCREW COULD NOT BE ADDED. PATIENT WAS REVISED TO AN OCCIPITAL TO T2 FUSION. IN THE REVISION, THE SURGEON PUT IN ALL THREE SCREWS. THIS REPORT IS FOR A MOUNTAINEER OCCIPITAL SCREW. THIS IS REPORT 4 OF 5 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13712052 ↗InjuryMar 9, 2022Migration; Migration
IT WAS REPORTED THAT PATIENT UNDERWENT REVISION SURGERY OF A POSTERIOR SPINAL FUSION ON (B)(6) 2022. THE ORIGINAL PROCEDURE WAS DONE ON (B)(6) 2022. ON AN UNKNOWN DATE FOLLOWING THE INITIAL PROCEDURE, THE MOUNTAINEER OC PLATE, TWO (2) RODS, AND TWO (2) SET SCREWS POPPED OUT. BOTH RODS POPPED OUT BUT WERE STILL IN THE CERVICAL PORTION. IT WAS NOTED THAT THE PATIENT DID NOT WEAR A NECK BRACE ALL THE TIME AND THE NECK MAY HAVE BEEN HYPEREXTENDED IN THE ORIGINAL CASE. THE DOCTOR HAD ONLY PUT TWO SCREWS ON THE PLATE. THE ANGLE WAS BAD, AND THE THIRD SCREW COULD NOT BE ADDED. PATIENT WAS REVISED TO AN OCCIPITAL TO T2 FUSION. IN THE REVISION, THE SURGEON PUT IN ALL THREE SCREWS. THIS REPORT IS FOR A MOUNTAINEER OCCIPITAL SCREW. THIS IS REPORT 3 OF 5 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13711973 ↗MalfunctionAug 4, 2017Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING FINAL TIGHTENING OF OCCIPITAL PLATE THE SET SCREWS WOULD NOT LOCK IN AND APPEARED TO BE CROSS THREADED. SEVERAL SET SCREWS WERE TRIED, IN THE END A NEW 31MM PLATE, SET SCREWS AND FINAL TIGHTENER WERE USED AND THE CASE WAS ABLE TO BE COMPLETED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6768851 ↗MalfunctionAug 4, 2017Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING FINAL TIGHTENING OF OCCIPITAL PLATE THE SET SCREWS WOULD NOT LOCK IN AND APPEARED TO BE CROSS THREADED. SEVERAL SET SCREWS WERE TRIED, IN THE END A NEW 31MM PLATE, SET SCREWS AND FINAL TIGHTENER WERE USED AND THE CASE WAS ABLE TO BE COMPLETED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6768830 ↗MalfunctionAug 4, 2017Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING FINAL TIGHTENING OF OCCIPITAL PLATE THE SET SCREWS WOULD NOT LOCK IN AND APPEARED TO BE CROSS THREADED. SEVERAL SET SCREWS WERE TRIED, IN THE END A NEW 31MM PLATE, SET SCREWS AND FINAL TIGHTENER WERE USED AND THE CASE WAS ABLE TO BE COMPLETED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6768803 ↗MalfunctionAug 4, 2017Material Deformation; Material Deformation
IT WAS REPORTED THAT DURING FINAL TIGHTENING OF OCCIPITAL PLATE THE SET SCREWS WOULD NOT LOCK IN AND APPEARED TO BE CROSS THREADED. SEVERAL SET SCREWS WERE TRIED, IN THE END A NEW 31 MM PLATE, SET SCREWS AND FINAL TIGHTENER WERE USED AND THE CASE WAS ABLE TO BE COMPLETED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6768767 ↗
Other Thoracolumbosacral pedicle screw system devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page