CONFIDENCE SPINAL CEMENT SYSTEMRecalls & reported problems across all models
Medos International Sàrl · 11 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 359
- Deaths reported
- 13
- Injuries reported
- 52
- Recalls
- 0
We found 11 models of CONFIDENCE SPINAL CEMENT SYSTEM by Medos International Sàrl in the FDA's device registry (GUDID). Together they are named in 359 adverse event reports in the FDA's MAUDE database, received between Aug 1, 2017 and Sep 30, 2026 — classified as 13 deaths, 52 injuries, 294 malfunctions.
The most frequently reported specific device problem is “Migration or Expulsion of Device” (109 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported CONFIDENCE SPINAL CEMENT SYSTEM problems
Device problems
- Migration or Expulsion of Device109
- Chemical Problem77
- Break72
- Failure to Eject34
- Adverse Event Without Identified Device or Use Problem29
- Fluid/Blood Leak27
- Appropriate Device Problem Term/Code Not Available21
- Entrapment of Device13
Patient problems
- No Clinical Signs, Symptoms or Conditions206
- No Consequences Or Impact To Patient63
- Foreign Body In Patient28
- No Known Impact Or Consequence To Patient16
- No Code Available11
- Insufficient Information10
- Unspecified Infection9
- Injury5
CONFIDENCE SPINAL CEMENT SYSTEM models
11 models, most reported first. Each model has its own page with its reports, recalls and device details.
- CONFIDENCE SPINAL CEMENT SYSTEM 283910000CEMENT, BONE, VERTEBROPLASTY5 deaths163 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283913000CEMENT, BONE, VERTEBROPLASTY8 deaths146 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283903611NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE17 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283903411NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE15 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283902611NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE6 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283905000CEMENT, BONE, VERTEBROPLASTY6 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283902613NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE2 reports
- CONFIDENCE SPINAL CEMENT SYSTEM 283902411NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE1 report
- CONFIDENCE SPINAL CEMENT SYSTEM 283902413NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE1 report
- CONFIDENCE SPINAL CEMENT SYSTEM 283904413NEEDLE, ASPIRATION AND INJECTION, DISPOSABLE1 report
- CONFIDENCE SPINAL CEMENT SYSTEM 283999002ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1 report
Recent CONFIDENCE SPINAL CEMENT SYSTEM adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Chemical Problem
It was reported that the cement was very tough, could not be mixed as described in the packaging instructions. There was no patient harm or surgical delay.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50948790 ↗MalfunctionSep 24, 2026Migration or Expulsion of Device; Failure to Eject
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50885539 ↗MalfunctionSep 22, 2026Migration or Expulsion of Device
It was reported that when using Confidence granulation, both stirring and assembly during the procedure went smoothly. Bone cement was poured in, and the rotary pressure doctor felt no pressure and noticed the milk bottle was leaking. They immediately requested a mobile video and urgently contacted the supervisor, deciding to reapply a new CONFIDENCE pack. The surgery was successfully completed with the defective product retained.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50859061 ↗MalfunctionSep 21, 2026Migration or Expulsion of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50841885 ↗MalfunctionSep 9, 2026Break; Migration or Expulsion of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50698204 ↗MalfunctionSep 2, 2026Migration or Expulsion of Device; Device-Device Incompatibility
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50572833 ↗MalfunctionSep 2, 2026Migration or Expulsion of Device; Failure to Eject
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50564156 ↗MalfunctionMay 18, 2026Chemical Problem; Failure to Eject
DURING THE VERTEBROPLASTY PROCEDURE, THE BONE CEMENT (CONFIDENCE KIT) WAS OPENED AND PREPARED. AT THE TIME OF TRANSFERRING THE PRODUCT TO THE APPLICATION SYSTEM, SIGNIFICANT DIFFICULTY WAS OBSERVED, AND AN INCREASE IN VISCOSITY WAS IDENTIFIED THAT WAS INCOMPATIBLE WITH THE APPROPRIATE HANDLING PHASE. IN VIEW OF THE IMPOSSIBILITY OF USING THE PREPARED MATERIAL, IT WAS DISCARDED ACCORDING TO THE GUIDANCE OF THE HOSPITAL INSTITUTION. THERE WAS NO DAMAGE REPORTED BY THE PATIENT OR USER. THERE WAS NO SIGNIFICANT DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25202086 ↗MalfunctionMay 15, 2026Migration or Expulsion of Device
IT WAS REPORTED THE CONFIDENCE CEMENT PUMP LEAKED INSTEAD OF PROPELLING CEMENT NO PATIENT HARM AND NO SURGICAL DELAY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25183840 ↗MalfunctionApr 22, 2026Chemical Problem
INSUFFICIENT MIXING LIQUID: AFTER OPENING THE BOTTLE AND STIRRING, IT CANNOT BE MIXED EVENLY; SOME CEMENT REMAINS IN POWDER FORM. ONLY HALF THE NORMAL USAGE AMOUNT IS SQUEEZED INTO THE BOTTLE. A NEW BOTTLE WAS USED. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957902 ↗MalfunctionMar 6, 2026Migration or Expulsion of Device; Failure to Eject
IT WAS REPORTED THAT WHEN THE BONE CEMENT WAS HALFWAY POURED IN, WATER STARTED LEAKING FROM THE CAP. THE CAP WAS CHECKED AND FOUND TO BE SECURELY LOCKED. THE DOCTOR CONTINUED TO TURN THE PRESSURE GUN, BUT WATER BEGAN SPRAYING OUT OF THE CAP. BECAUSE TOO MUCH WATER WAS BEING LOST FROM THE PRESSURE GUN, THE BONE CEMENT COULD NOT BE POURED IN FURTHER, AND THE ATTENDING PHYSICIAN REQUESTED ANOTHER PACKET. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24533483 ↗MalfunctionFeb 26, 2026Migration or Expulsion of Device
IT WAS REPORTED THAT FLUID FROM THE PUMP HAS BEEN LEAKING DURING THE APPLICATION OF CEMENT TO THE SCREWS INTRAOPERATIVELY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24460857 ↗MalfunctionFeb 26, 2026Crack
IT WAS REPORTED THAT ON FEBRUARY 13, 2026, THE CEMENT MIXING BOWL CRACKED WHEN THE BOX WAS OPENED. THERE WERE NO REPORTED PATIENT CONSEQUENCES. IT IS UNKNOWN IF THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24456643 ↗MalfunctionFeb 20, 2026Break
IT WAS REPORTED THAT THE PUMP SYSTEM HAS BURST. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24399904 ↗MalfunctionFeb 9, 2026Chemical Problem
IT WAS REPORTED THAT CONFIDENCE KIT, NO NEEDLES, EXPERIENCED A FAILURE IN WHICH THE CEMENT BECAME LUMPY AFTER MIXING, RESULTING IN AN INABILITY TO MIX THE CEMENT PROPERLY. THE EVENT OCCURRED INTRA-OPERATIVELY. THE PATIENT DID NOT COME INTO CONTACT WITH THE PRODUCT, AND THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. SURGERY WAS PROLONGED BY EIGHT MINUTES, BUT THE PROCEDURE WAS SUCCESSFULLY COMPLETED AFTER A NEW ITEM WAS OPENED. NO FRAGMENTS WERE GENERATED, AND NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24292004 ↗
Other Medos International Sàrl products
- TRUESPAN ORTHOCORD6 models · 4,109 reports
- FMS39 models · 3,588 reports
- EXPEDIUM291 models · 3,312 reports
- EXPEDIUM VERSE90 models · 2,686 reports
- EXPRESSEW6 models · 2,459 reports
- OMNISPAN ORTHOCORD3 models · 1,581 reports
- HEALIX ADVANCE ORTHOCORD21 models · 1,311 reports
- PROWLER SELECT9 models · 1,066 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page