VIPER 179704990Recalls & reported problems
Medos International Sàrl · VIPER SYSTEM POLYAXIAL WIDE LAG SCREW 5.5 9 x 90mm · Class 2 · Implant
All models of VIPER (121)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10705034265865), received between Aug 6, 2018 and Apr 22, 2022 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break3
- Loose or Intermittent Connection2
- Device Slipped1
- Material Split, Cut or Torn1
Patient problems
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
- No Clinical Signs, Symptoms or Conditions2
- No Code Available2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 22, 2022Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT DURING THE SURGERY, WHEN ASSEMBLING SCREWS, THE INTERNAL THREAD WAS BROKEN AND DROPPED. ALL THE BROKEN PARTS WERE REMOVED. ANOTHER DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION COULD BE PROVIDED. THIS REPORT IS FOR ONE (1) VIPER TI SAI POLY 9X90MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14187544 ↗InjuryMar 4, 2022Device Slipped; Device Slipped
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THE PATIENT UNDERWENT THE PRIMARY PROCEDURE AND THE VIPER POLYAXIAL SCREW WAS IMPLANTED AT S2. ON AN UNKNOWN POSTOPERATIVE DATE IT WAS FOUND THAT THE SCREW HAD BEEN LOOSE. THE PATIENT UNDERWENT A REVISION PROCEDURE FOR LOOSE SCREW REPLACEMENT ON (B)(6) 2022. NO FURTHER INFORMATION IS AVAILABLE. THIS INCIDENT IS RELATED TO (B)(4) WHICH CAPTURES THE INTRAOPERATIVE SCREWDRIVERS BREAKAGES DURING THE REVISION PROCEDURE. THIS REPORT IS FOR ONE (1) VIPER TI SAI POLY 9X90MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 13668368 ↗MalfunctionFeb 11, 2022Break; Break
IT WAS REPORTED THAT ON (B)(6) 2022, DURING A PROCEDURE DURING FINAL TIGHTENING OF THE SET SCREW THE VIPER SAI SCREW PARTIAL THREAD (9X90MM) HEAD SPLAYED. THE SURGEON HAD TO REMOVE THIS SCREW, AND REPLACE WITH A 10X90MM VIPER SAI SCREW, WHICH WAS THEN ABLE TO FINAL TIGHTEN. NO PATIENT CONSEQUENCES. CONCOMITANT DEVICE REPORTED: UNKNOWN TIGHTENER (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) VIPER TI SAI POLY 9X90MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13515029 ↗InjuryAug 23, 2021Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE PATIENT HAD A SURGERY FOR POSTERIOR SPINAL FUSION IN (B)(6) OF 2020. AT A FOLLOW UP APPOINTMENT IN (B)(6) OF 2020, XRAY IMAGES SHOWED THAT THE SET SCREW HAD LOOSENED FROM THE SAI SCREWS BILATERALLY AND THE ROD HAD LOOSENED DUE TO HEAD SPLAY. THE S1 AND L5 SCREWS HAD FRACTURED AT THE NECK OF THE SCREW JUST BELOW THE TULIP. IT IS UNKNOWN IF FRAGMENTS WERE GENERATED. THE PATIENT DID NOT ACHIEVE A FUSION BELOW L5 DUE TO THIS HARDWARE FAILURE. PROCEDURE OUTCOME IS UNKNOWN. THIS COMPLAINT INVOLVES FIVE (5) DEVICES. THIS REPORT IS FOR (1) VIPER TI SAI POLY 9X90 MM. THIS REPORT IS 2 OF 5 FOR (B)(4).
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12354202 ↗InjuryJul 24, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT REVISION SURGERY DUE TO THE ROD THAT WAS BROKEN BILATERALLY AND THE DISTAL SCREWS WERE LOOSENED. BROKEN ENDS OF ROD CATALOG # 1797-62-480, VERSE SCREW, 1997-21-750 X QTY 1, VERSE SCREW 1997-21-745 X QTY 1 AND 9.0X90MM SAI SCREWS 1797-04-990 X QTY 2 WERE ALL REMOVED TO REVISE. A 12.0X100MM SCREW, 1797-12-399 WAS REMOVED WITH PLIERS DUE TO THE FIRST OF TWO SCREWDRIVER TIPS BREAKING. THIS SCREW WAS NOT RECOVERED TO RETURN. THERE WERE NO FRAGMENTS GENERATED. IT IS UNKNOWN IF THERE WERE SURGICAL DELAY. THE PATIENT AND PROCEDURE OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES SIX (6) DEVICES.
Patient problems: No Code Available
MAUDE report 8821962 ↗InjuryJul 24, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT REVISION SURGERY DUE TO THE ROD THAT WAS BROKEN BILATERALLY AND THE DISTAL SCREWS WERE LOOSENED. BROKEN ENDS OF ROD CATALOG # 1797-62-480, VERSE SCREW, 1997-21-750 X QTY 1, VERSE SCREW 1997-21-745 X QTY 1 AND 9.0X90MM SAI SCREWS 1797-04-990 X QTY 2 WERE ALL REMOVED TO REVISE. A 12.0X100MM SCREW, 1797-12-399 WAS REMOVED WITH PLIERS DUE TO THE FIRST OF TWO SCREWDRIVER TIPS BREAKING. THIS SCREW WAS NOT RECOVERED TO RETURN. THERE WERE NO FRAGMENTS GENERATED. IT IS UNKNOWN IF THERE WERE SURGICAL DELAY. THE PATIENT AND PROCEDURE OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES SIX (6) DEVICES.
Patient problems: No Code Available
MAUDE report 8821841 ↗InjuryAug 6, 2018Material Split, Cut or Torn; Material Split, Cut or Torn
SCREW WAS INSERTED WITH CORRECT SCREW DRIVER AND NO ISSUES. ROD WAS PLACED IN CONSTRUCT AND LOCKING CAPS WERE INSERTED. THE LOCKING CAPS FOR THE SAI SCREW COULD NOT BE TIGHTENED ALTHOUGH ROD SEEMED TO BE SEATING WELL IN SCREW HEAD. A SECOND LOCKING CAP WAS TAKEN TRYING AGAIN TO LOCK IT. THE IT WAS REALIZED THAT THE ISSUE WAS WITH THE SCREW AND NOT THE LOCKING CAP. ACCORDINGLY THE SCREW WAS REPLACED WITH A NEW ONE AND EVERYTHING COULD BE COMPLETED WITHOUT A PROBLEM. PATIENT CONSEQUENCE?: NO. ACTION TAKEN FOR PROCEDURE:NEW SCREW WAS INSERTED. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT?: YES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7752082 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page