VIPER 186727770Recalls & reported problems
Medos International Sàrl · VIPER SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 x 7 x 70mm · Class 2 · Implant
All models of VIPER (121)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034282725), received between Jun 11, 2019 and Jun 11, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Code Available1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 11, 2019Break; Break
IT WAS REPORTED THAT ON (B)(6) 2019, DURING AN UNKNOWN PROCEDURE, THE CORTICAL FIX FENESTRATED SCREW MIS TI CFX FEN POLY 8X80 WAS BEING INSERTED INTO THE S2-ALAR POSITION AND BECAME TIGHT. AT THE POINT OF NO POSSIBLE FURTHER FORWARD MOMENTUM, THE DECISION WAS TAKEN TO REMOVE THE SCREW AND UTILIZE A SMALLER DIAMETER SCREW. THE SCREW CAME DETACHED FROM THE SCREW INSERTION INSTRUMENT (286725300) AS THE TORQUE REQUIRED WAS TOO LARGE. MANY ALTERNATIVE INSTRUMENTS WERE USED TO TRY AND REMOVE THE SCREW WITH COUNTERCLOCKWISE ROTATIONS HOWEVER THEY DIDN¿T WORK. DURING THIS SCREW REMOVAL ATTEMPT, A SCREW MIS TI CFX FEN POLY 7X70 PLACED INTRADISCALLY THOUGH S1-L5 WAS TO BE REVERSED SLIGHTLY IN ORDER TO GAIN FURTHER ROOM AND LEVERAGE ON MIS TI CFX FEN POLY 8X80. AT THE FIRST ATTEMPT TO REVERSE THIS SCREW, THE POLYAXIAL HEAD EVENTUALLY PULLED OVER THE HEAD OF THE SHANK OF THE SCREW WHICH HAD FATIGUED AND BROKEN. THERE WAS SIGNIFICANT PROUDNESS OF THE SCREW SHANK AND THEREFORE THERE WAS SIGNIFICANT TIME TAKEN TO TRY TO REMOVE THIS. EVENTUALLY, SEVERAL BURRS WERE USED TO ¿CUT THROUGH¿ THE SCREW AND GET IT FLUSH TO THE PATIENTS BONE WHERE IT WAS LEFT IN SITU. THERE WAS A SURGICAL DELAY ON AN UNKNO
Patient problems: No Code Available
MAUDE report 8688834 ↗
FDA clearance and classification
Cleared through 510(k) K160879 on Dec 20, 2016 (decision: Substantially Equivalent). Submitted by Medos International Sàrl as “CONFIDENCE High Viscosity Spinal Cement, VIPER® and EXPEDIUM® Fenestrated Screw Systems”. The FDA review took 265 days. FDA record ↗
The FDA regulates product code PML (bone cement, posterior screw augmentation) under 21 CFR 888.3027 as a Class II device. 21 CFR 888.3027 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
FDA definition of this device type: The device is intended to augment the fixation of screws in a posterior spinal system construct.
Common questions about the VIPER 186727770
Has the VIPER 186727770 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034282725), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the VIPER 186727770?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jun 11, 2019. That report lists device problem “Break” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VIPER 186727770?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the VIPER 186727770 FDA approved?
It was cleared through 510(k) K160879 on Dec 20, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the VIPER 186727770 MRI safe?
According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.
Who makes the VIPER 186727770, and is it still sold?
The FDA's device database lists Medos International Sàrl as the company responsible for it (brand VIPER, model 186727770). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other bone cement, posterior screw augmentation devices
- VIPER 186727645Medos International Sàrl · bone cement, posterior screw augmentation3 deaths27 reports
- VIPER 186727745Medos International Sàrl · bone cement, posterior screw augmentation23 reports
- EXPEDIUM VERSE 199723645Medos International Sàrl · bone cement, posterior screw augmentation21 reports
- EXPEDIUM VERSE 199723745Medos International Sàrl · bone cement, posterior screw augmentation13 reports
- VIPER 186727640Medos International Sàrl · bone cement, posterior screw augmentation2 deaths11 reports
- VIPER 186727850Medos International Sàrl · bone cement, posterior screw augmentation9 reports
- EXPEDIUM VERSE 199723435Medos International Sàrl · bone cement, posterior screw augmentation6 reports
- EXPEDIUM VERSE 199723655Medos International Sàrl · bone cement, posterior screw augmentation4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page