EXPEDIUM VIPER 196789600Recalls & reported problems

Depuy Spine, LLC · EXPEDIUM and VIPER 2 SPINE SYSTEM ROD, STRAIGHT 5.5 x 600mm · Class 2 · Implant

All models of EXPEDIUM VIPER (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10705034295282), received between Oct 2, 2017 and May 16, 2024 — classified as 8 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

1 other recall cover devices cleared under the same FDA submission (K090648, K111136); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Device Slipped2
  • Adverse Event Without Identified Device or Use Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Deformity/ Disfigurement2
  • Failure of Implant2
  • Fall2
  • Injury2
  • Necrosis2
  • Pain2
  • Spinal Column Injury2

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 16, 2024Device Slipped; Device Slipped

    IT WAS REPORTED THAT ON OR ABOUT (B)(6), 2020 A DEPUY SPINAL WAS USED IN PATIENT SURGERY. UPON INFORMATION, THE PRODUCT USED IN PATIENT SURGERY INCLUDE DEPUY EXPEDIUM, COBALT CHROMIUM RODS, UNIAXIAL PEDICLE SCREWS, DEPUY SPINE SCREWS, ROD VIPER STRAIGHT AND OR OTHER ASSOCIATED DEVICES. THE PATIENT HAD SOME PAIN POSTOPERATIVELY AND ALSO HAD DESCRIPTIONS OF CRACKING AND POPPING INVOLVING HER LUMBOSACRAL SPINE. X-RAY REVEALED A DISENGAGEMENT OF THE LEFT-SIDED ROD FROM HER L3 PEDICLE SCREW AND THE PATIENT'S CORONAL BALANCE ALSO APPEARED TO BE SIGNIFICANTLY AFFECTED. PATIENT WAS CAUSED TO SUSTAIN SEVERE, SERIOUS, PERMANENT INJURIES THAT HAD RENDERED HER SICK LORE, LAME AND DISABLED. ALSO NOTED THE PATIENT EXPERIENCED NECROSIS. ON (B)(6), 2020 THE PATIENT UNDERWENT REINSERTION OF SPINAL FIXATION DEVICE WITH REVISION OF INSTRUMENTATION DUE TO POSTERIOR INSTRUMENTED SPINAL FUSION HARDWARE FAILURE. ON (B)(6), 2020 THE PATIENT UNDERWENT EXCISIONAL DEBRIDEMENT OF SKIN INCLUDING THE SUBCUTANEOUS LEVEL AND FASCIA, RIGHT SIDE AND LEFT SIDE. THIS REPORT INVOLVES ONE EXPEDIUM AND VIPER 2 SPINE SYSTEM ROD, STRAIGHT 5.5 X 600MM. THIS IS REPORT 2 OF 10 FOR (B)(4). ADDITIONAL IMPACTED PRODUCTS RELATED

    Patient problems: Deformity/ Disfigurement; Necrosis; Pain; Spinal Cord Injury

    MAUDE report 19338148 ↗
  • InjuryMay 16, 2024Device Slipped; Device Slipped

    IT WAS REPORTED THAT ON OR ABOUT (B)(6) 2020 A DEPUY SPINAL WAS USED IN PATIENT SURGERY. UPON INFORMATION, THE PRODUCT USED IN PATIENT SURGERY INCLUDE DEPUY EXPEDIUM, COBALT CHROMIUM RODS, UNIAXIAL PEDICLE SCREWS, DEPUY SPINE SCREWS, ROD VIPER STRAIGHT AND OR OTHER ASSOCIATED DEVICES. THE PATIENT HAD SOME PAIN POSTOPERATIVELY AND ALSO HAD DESCRIPTIONS OF CRACKING AND POPPING INVOLVING HER LUMBOSACRAL SPINE. X-RAY REVEALED A DISENGAGEMENT OF THE LEFT-SIDED ROD FROM HER L3 PEDICLE SCREW AND THE PATIENT'S CORONAL BALANCE ALSO APPEARED TO BE SIGNIFICANTLY AFFECTED. PATIENT WAS CAUSED TO SUSTAIN SEVERE, SERIOUS, PERMANENT INJURIES THAT HAD RENDERED HER SICK LORE, LAME AND DISABLED. ALSO NOTED THE PATIENT EXPERIENCED NECROSIS. ON (B)(6) 2020 THE PATIENT UNDERWENT REINSERTION OF SPINAL FIXATION DEVICE WITH REVISION OF INSTRUMENTATION DUE TO POSTERIOR INSTRUMENTED SPINAL FUSION HARDWARE FAILURE. ON (B)(6) 2020 THE PATIENT UNDERWENT EXCISIONAL DEBRIDEMENT OF SKIN INCLUDING THE SUBCUTANEOUS LEVEL AND FASCIA, RIGHT SIDE AND LEFT SIDE. THIS REPORT INVOLVES ONE EXPEDIUM AND VIPER 2 SPINE SYSTEM ROD, STRAIGHT 5.5 X 600MM. THIS IS REPORT 1 OF 10 FOR (B)(4). ADDITIONAL IMPACTED PRODUCTS RELATED TO

    Patient problems: Deformity/ Disfigurement; Necrosis; Pain; Spinal Cord Injury

    MAUDE report 19338147 ↗
  • InjuryJan 9, 2024Break; Break

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: PATIENT ORIGINALLY UNDERWENT SPINE SURGERY ON (B)(6) 2005. ON (B)(6) 2015, PATIENT UNDERWENT REVISION SURGERY DUE TO ROD BREAKAGE. THE RE-FUSION WAS DONE WITH SCREWS. ON (B)(6) 2023, PATIENT UNDERWENT A SECOND REVISION PROCEDURE DUE TO ROD RE-BREAKAGE. DURING THE PROCEDURE, A SET SCREW AND CONNECTOR BROKE. PATIENT WAS STABLE. THIS REPORT IS FOR A VIPER2 STRAIGHT ROD-600MM, COC. THIS IS REPORT 2 OF 2 FOR (B)(4) AND CAPTURES THE FIRST REVISION SURGERY. THE SECOND REVISION SURGERY IS CAPTURED UNDER (B)(4). THE INTRAOPERATIVE INCIDENT IS CAPTURED UNDER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18479497 ↗
  • InjuryJan 9, 2024Break; Break

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: PATIENT ORIGINALLY UNDERWENT SPINE SURGERY ON (B)(6) 2005. ON (B)(6) 2015, PATIENT UNDERWENT REVISION SURGERY DUE TO ROD BREAKAGE. THE RE-FUSION WAS DONE WITH SCREWS. ON (B)(6) 2023, PATIENT UNDERWENT A SECOND REVISION PROCEDURE DUE TO ROD RE-BREAKAGE. DURING THE PROCEDURE, A SET SCREW AND CONNECTOR BROKE. PATIENT WAS STABLE. THIS REPORT IS FOR A VIPER2 STRAIGHT ROD-600MM, COC. THIS IS REPORT 1 OF 2 FOR (B)(4) AND CAPTURES THE FIRST REVISION SURGERY. THE SECOND REVISION SURGERY IS CAPTURED UNDER (B)(4). THE INTRAOPERATIVE INCIDENT IS CAPTURED UNDER (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18479369 ↗
  • MalfunctionNov 19, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, THE VIPER2 STRAIGHT ROD-600MM, COC WERE BROKEN DURING REVERSE LOGISTICS AUDIT OF RETURNED DEVICE AT MILLSTONE. THERE WAS NO PATIENT INVOLVEMENT, AND NO ADDITIONAL INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES TWO (2) NUMBER OF DEVICES. THIS REPORT IS FOR (1) VIPER2 STRAIGHT ROD-600MM, COC. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12848979 ↗
  • MalfunctionNov 19, 2021Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2021, THE VIPER2 STRAIGHT ROD-600MM, COC WERE BROKEN DURING REVERSE LOGISTICS AUDIT OF RETURNED DEVICE AT (B)(4). THERE WAS NO PATIENT INVOLVEMENT, AND NO ADDITIONAL INFORMATION IS AVAILABLE. THIS COMPLAINT INVOLVES TWO (2) NUMBER OF DEVICES. THIS REPORT IS FOR (1) VIPER2 STRAIGHT ROD-600MM, COC. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12848974 ↗
  • InjuryJul 31, 2018Break; Break

    FURTHER TO (B)(4), ON (B)(6) 2018, WHEN THE PATIENT VISITED THE HOSPITAL DUE TO A LOWER BACK PAIN, THE BREAKAGE OF THE ROD (196789600) WAS FOUND UNDER THE X-RAY BETWEEN L4 AND L5 (RIGHT), AND BETWEEN L5 AND S (LEFT). ON (B)(6) 2018, THE RE-OPERATION WAS RESCHEDULED DUE TO THE PATIENT¿S HIGH CRP (C-REACTIVE PROTEIN) LEVEL. ON (B)(6) 2018, THE RE-OPERATION WAS PERFORMED TO FIND THE CAUSE OF THE PATIENT¿S HIGH CRP (C-REACTIVE PROTEIN) LEVEL, CONSIDERING THE SUSPECTED INFECTIONS. THE PROCEDURES WERE CONDUCTED IN THE FOLLOWING WAY. - SCREWS AT L4, L5 AND S1 (RIGHT) WERE EXTRACTED AND NEW SCREWS WERE INSERTED AT L5 AND S1 TO CONNECT WITH UPPER RODS USING A CONNECTOR. - A SCREW AT L5 (LEFT) WAS EXTRACTED TO CONNECT WITH UPPER RODS USING A CONNECTOR. - SI SET SCREWS AT L4 AND OVER (BOTH SIDES) WERE REPLACED WITH NEW ONES, AND RODS AT FAR END WERE ALSO REPLACED WITH NEW ONES. ALSO, THE FOLLOWING WERE COMMENTED BY THE SURGEON. - LOOSENING OF ALL THE SCREWS WERE OBSERVED WHEN EXTRACTED - TORQUE WAS NOT PROPERLY APPLIED TO SI SET SCREWS WHEN EXTRACTED. - BLACK-COLORED SUBSTANCES WERE SEEN AROUND THE SCREW AND THE ROD, WHICH MIGHT HAVE BEEN CAUSED BY THE METAL CORROSION. - THE PATIENT¿S CRP LEV

    Patient problems: Spinal Column Injury; Unspecified Infection

    MAUDE report 7736313 ↗
  • InjuryJul 31, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2016, THE PRIMARY SURGERY WAS PERFORMED TO FIX T10 ¿ SAI. ON (B)(6) 2017, DUE TO UNKNOWN REASONS, THE EXTENSION SURGERY WAS PERFORMED TO FIX T7 ¿ SAI. NO FURTHER INFORMATION HAS BEEN PROVIDED BY THE HOSPITAL.

    Patient problems: Spinal Column Injury

    MAUDE report 7734790 ↗
  • InjuryOct 2, 2017Break; Break

    REVISION SCOLIOSIS CORRECTION ON (B)(6) 2017. PATIENT UNDERWENT PRIMARY CORRECTION (B)(6) 2010. 3 YEARS AGO THE PATIENT HAD A FALL AND THE SCREWS BROKE JUST UNDER THE TULIP AND THE COCR RODS ALSO BROKE. X-RAY PROVIDED. CASE IS LINKED TO 5109967 - THIS CASE CAPTURES THE INSTRUMENT ISSUES THAT OCCURRED DURING THE REVISION PROCEDURE.

    Patient problems: Failure of Implant; Fall; Injury

    MAUDE report 6906915 ↗
  • InjuryOct 2, 2017Break; Break

    REVISION SCOLIOSIS CORRECTION ON (B)(6) 2017. PATIENT UNDERWENT PRIMARY CORRECTION (B)(6) 2010. THREE (3) YEARS AGO THE PATIENT HAD A FALL AND THE SCREWS BROKE JUST UNDER THE TULIP AND THE COCR RODS ALSO BROKE. X-RAY PROVIDED. CASE IS LINKED TO (B)(4) - THIS CASE CAPTURES THE INSTRUMENT ISSUES THAT OCCURRED DURING THE REVISION PROCEDURE.

    Patient problems: Failure of Implant; Fall; Injury

    MAUDE report 6906896 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K090648, K111136. One submission often covers many models, so check the product and lot details in each record.

  • Z-1674-2015 · Apr 21, 2015 · Terminated
    EXPEDIUM Spine System/Synapse System; MODEL: EXPEDIUM Offset Conical Lock Nut 8mm, CATALOG NO. : 1754-91-150, UDI NUMBER (if applicable): GTIN: 10705034032467

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →