EXPEDIUM 196791455Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM PRE-CONTOURED NORMAL LARGE 5.5 · Class 2 · Implant
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10705034295312), received between Apr 3, 2018 and Mar 7, 2019 — classified as 4 injuries.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Break4
Patient problems
- Spinal Column Injury3
- Not Applicable2
- No Code Available1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 7, 2019Break; Break
THE FOLLOWING WAS REPORTED: REVISION T4-PELVIS & BROKEN PEDICLE SCREWS: DR (B)(6), (B)(6) HOSPITAL, (B)(6) 2019. PT HAD A POSTERIOR SPINAL FUSION FOR SPINAL DEFORMITY, T4-PELVIS ((B)(6) 2016). IT WAS QUITE A SIGNIFICANT DEFORMITY. IT WAS NOTED THAT THE RODS BROKE AT L1 (2018) WHERE THERE IS A NON-UNION. THIS IS STAGE 1 OF 2. THE RODS WERE REMOVED ON BOTH SIDES, BROKEN SCREWS REMOVED, NEW SCREWS IMPLANTED AND THEN NEW COCR RODS. TWO NEW SFX CROSS CONNECTORS WERE ALSO USED. IMPACTED PRODUCT: UNKNOWN SPD= 5.5MM ROD, TI, CODE: 2797.62.300, NOT AVAILABLE.
Patient problems: No Code Available
MAUDE report 8400607 ↗InjuryJun 5, 2018Break; Break
REV T3-PELVIS, BROKEN RODS AT L5/S1; (B)(6) 2018. PT IN FOR SURGERY FOR BROKEN RODS AT L5/S1 ON BOTH SIDES. EXISTING CONSTRUCT - T3-PELVIS ((B)(6) 2015). SURGEON STATED THAT THERE WAS NO OBVIOUS NON UNION. REPLACED SCREWS AND RODS, AT L4-PELVIS. ALSO NOTED THAT WHEN HE WENT TO REMOVE THE ILIAC SCREW ON THE L) SIDE, THE SCREW HEAD BROKE OFF FROM THE SHAFT. THE SURGEON LEFT THIS IN THE PATIENT AND PLACED ANOTHER SCREW IN THE PELVIS. THE RODS WERE REMOVED AND REPLACED AND JOINED TO THE REMAINING CONSTRUCT WITH AXIAL CONNECTORS.
Patient problems: Not Applicable; Spinal Column Injury
MAUDE report 7568696 ↗InjuryJun 5, 2018Break; Break
REV T3-PELVIS, BROKEN RODS AT L5/S1; (B)(6) HOSPITAL (B)(6) 2018. PT IN FOR SURGERY FOR BROKEN RODS AT L5/S1 ON BOTH SIDES. EXISTING CONSTRUCT - T3-PELVIS ((B)(6) 2015). SURGEON STATED THAT THERE WAS NO OBVIOUS NON UNION. REPLACED SCREWS AND RODS, AT L4-PELVIS. ALSO NOTED THAT WHEN HE WENT TO REMOVE THE ILIAC SCREW ON THE L) SIDE, THE SCREW HEAD BROKE OFF FROM THE SHAFT. THE SURGEON LEFT THIS IN THE PATIENT AND PLACED ANOTHER SCREW IN THE PELVIS. THE RODS WERE REMOVED AND REPLACED AND JOINED TO THE REMAINING CONSTRUCT WITH AXIAL CONNECTORS.
Patient problems: Not Applicable; Spinal Column Injury
MAUDE report 7568639 ↗InjuryApr 3, 2018Break; Break
SURGEON REQUESTED HARDWARE REMOVAL INSTRUMENTS FOR (B)(6) 18 REVISION OF A (B)(6) 2017 INSTRUMENTED CASE. (B)(6) 2017 CASE WAS AN EXTENSION TO AND KATI GET REVISION,HARDWARE FROM T3-PELVIS. L2 SCREWS WERE SKIPPED, HARDWARE 455MM PRE BENT COCR ROD BROKE ON PATIENTS RIGHT SIDE AT L2 ALONG WITH THE EXPEDIUM 5.5 SFX. SURGEON REMOVED VERSE SCREWS L3-SACRO ILIAC FULLY THREADED SAI. SURGEON STATED SET SCREWS WERE LOOSE IN REMOVAL. SURGEON ALSO STATED THAT THE FORMABLE VIVIGEN WAS COMING OUT LIKE IT WENT IN PATIENT CONSEQUENCE? UNKNOWN. ACTION TAKEN FOR PROCEDURE:SURGEON REVISED CONSTRUCT REMOVING L3-PELVIS DEPUY REVISED WITH INVASIVE RELIGN SCREWS AND END TO END CONNECTORS. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? YES.
Patient problems: Spinal Column Injury
MAUDE report 7393455 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026