Skip to content

EXPEDIUM VERSE 199723440Recalls & reported problems

Medos International Sàrl · EXPEDIUM VERSE SPINE SYSTEM FENESTRATED CORTICAL FIX POLYAXIAL SCREW 5.5 4.35 x 40mm · Class 2 · Implant

All models of EXPEDIUM VERSE (89)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034403434), received between Nov 14, 2019 and Nov 14, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break1
  • Crack1

Patient problems

  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2019Break; Crack; Break; Crack

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING SPINAL FIXATION SURGERY ON AN UNKNOWN DATE, IT WAS NOTED THAT AFTER COMPLETING THE TOTAL SPINAL CONSTRUCT WITH THE EXPEDIUM VERSE ADVANCED SYSTEM AND TIGHTENING OF THE SETSCREWS, THE TABS WERE BROKEN OFF UNDER THE TOPNOTCH INSTEAD OF FLUSH WITH THE TOPNOTCH AS DESCRIBED IN THE SURGICAL TECHNIQUE. THIS CAUSED THE TOPNOTCH TO ALSO BREAK OFF THE SCREWHEAD. THE CONSTRUCT WAS LOOSENED, THIS SPECIFIC SCREW WAS REPLACED WITH A NEW ONE, AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THIS REPORT IS FOR AN EXPEDIUM VERSE 5.5 SCREW 4.35 X 40. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9321119 ↗

FDA clearance and classification

Cleared through 510(k) K160879 on Dec 20, 2016 (decision: Substantially Equivalent). Submitted by Medos International Sàrl as “CONFIDENCE™ High Viscosity Spinal Cement, VIPER® and EXPEDIUM® Fenestrated Screw Systems”. The FDA review took 265 days. FDA record ↗

The FDA regulates product code PML (bone cement, posterior screw augmentation) under 21 CFR 888.3027 as a Class II device. 21 CFR 888.3027 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

FDA definition of this device type: The device is intended to augment the fixation of screws in a posterior spinal system construct.

Common questions about the EXPEDIUM VERSE 199723440

Has the EXPEDIUM VERSE 199723440 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034403434), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the EXPEDIUM VERSE 199723440?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Nov 14, 2019. That report lists device problems “Break” (1) and “Crack” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EXPEDIUM VERSE 199723440?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the EXPEDIUM VERSE 199723440 FDA approved?

It was cleared through 510(k) K160879 on Dec 20, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the EXPEDIUM VERSE 199723440 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the EXPEDIUM VERSE 199723440, and is it still sold?

The FDA's device database lists Medos International Sàrl as the company responsible for it (brand EXPEDIUM VERSE, model 199723440). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other bone cement, posterior screw augmentation devices

All bone cement, posterior screw augmentation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page