EXPEDIUM 299704190Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM CORONAL IN-SITU BENDER, RIGHT 5.5 · Class 1
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705034466903), received between Nov 27, 2023 and Apr 20, 2026 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Material Deformation” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Material Deformation5
- Break4
- Corroded1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 20, 2026Material Deformation
IT WAS REPORTED THAT BENDER WAS DAMAGED, METAL BENT. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24928020 ↗MalfunctionApr 20, 2026Material Deformation
IT WAS REPORTED THAT BENDER WAS DAMAGED, METAL BENT. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24928013 ↗MalfunctionApr 20, 2026Material Deformation
IT WAS REPORTED THAT BENDER WAS DAMAGED, METAL BENT. THERE WAS NO PATIENT CONSEQUENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24927995 ↗MalfunctionApr 20, 2026Material Deformation
IT WAS REPORTED THAT POLY SHAFT DRIVERS ARE STRIPPED. THE HANDLES ARE NOT RATCHING. THE TAPS ARE DULL AND THE BENDERS ARE DAMAGED, METAL BENT. ROD HOLDERS WORN. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24925266 ↗MalfunctionMar 20, 2026Break
IT WAS REPORTED THAT THE EXPEDIUM VERSE CORONAL BENDER SNAPPED AT THE DISTAL END WHILST BENDING A 5.5 COCR ROD. THERE WAS NO PATIENT IMPACT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649146 ↗MalfunctionNov 11, 2025Material Deformation; Material Deformation
IT WAS REPORTED THAT PRODUCTS ARE BENT DUE TO USE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23523917 ↗MalfunctionApr 23, 2025Break; Break
IT WAS REPORTED THAT THE FRENCH BENDER CENTER PORTION BROKEN. CORONAL BENDERS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21892967 ↗MalfunctionApr 23, 2025Break; Break
IT WAS REPORTED THAT THE FRENCH BENDER CENTER PORTION BROKEN. CORONAL BENDERS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21892233 ↗MalfunctionNov 27, 2023Break; Corroded; Break; Corroded
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN DENMARK AS FOLLOWS: IT WAS REPORTED THAT AFTER SURGERY, THE STERILE DEPARTMENT FOUND SIGNS OF SMALL ¿HOLES¿ AND RUST ON THE SURFACE OF THE INSTRUMENTS. THE INSTRUMENTS ARE RARELY USED. IT LOOKS LIKE A ¿NON-MECHANICAL¿ REASON OF THE SURFACE CHANGES. INSTRUMENTS AND DECONTAMINATION CERTIFICATION WILL BE PICKED UP NEXT WEEK AT CLINIC. SAME LIKE PRODUCT WAS USED TO MANAGE THE PROBLEM. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) CORONAL IN-SITU BENDER, RIGHT THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18212174 ↗
Other BENDER devices
- MONARCH 277030000Depuy Spine, LLC · BENDER15 reports
- EXPEDIUM 299704195Medos International Sàrl · BENDER9 reports
- EXPEDIUM 299704200Medos International Sàrl · BENDER9 reports
- ALTALYNE 5161-04-1700Medos International Sàrl · BENDER8 reports
- EXPEDIUM 299704185Medos International Sàrl · BENDER8 reports
- NA 03.161.057Synthes GmbH · BENDER5 reports
- NA 03.221.251Synthes GmbH · BENDER4 reports
- EXPEDIUM 279763860Medos International Sàrl · BENDER4 reports