EXPEDIUM 299704200Recalls & reported problems
Medos International Sàrl · EXPEDIUM SPINE SYSTEM SAGITTAL IN-SITU BENDER, RIGHT 5.5 · Class 1
All models of EXPEDIUM (291)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 10705034467337), received between Feb 6, 2019 and Dec 2, 2025 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break8
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
- No Consequences Or Impact To Patient3
- Insufficient Information1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 2, 2025Break; Break
IT WAS REPORTED THE SAGITTAL IN-SITU BENDER CHIPPED INTRAOPERATIVELY. THERE WAS NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23694797 ↗MalfunctionDec 2, 2025Break; Break
IT WAS REPORTED THE SAGITTAL IN-SITU BENDER CHIPPED INTRAOPERATIVELY. THERE WAS NO CONSEQUENCES TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23694756 ↗MalfunctionMar 3, 2025Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN-SITU BENDER WAS SNAPPED AND ROD HOLDER BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21503938 ↗MalfunctionMar 3, 2025Break; Break
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN-SITU BENDER WAS SNAPPED AND ROD HOLDER BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21503842 ↗MalfunctionNov 29, 2023Break; Break
DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT DURING A POSTERIOR FUSION FOR HIP PAIN ON 07-NOVEMBER-2023, THE TIP OF THE BENDER IN QUESTION BROKE OFF. THERE WAS NO REPORTED ADVERSE PATIENT IMPACT. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR AN EXPEDIUM SPINE SYSTEM SAGITTAL IN-SITU BENDER, RIGHT 5.5. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: Insufficient Information
MAUDE report 18233384 ↗MalfunctionJan 5, 2022Material Deformation; Material Deformation
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR A SURGERY. DURING THE SURGERY, THE IN-SITU BENDER¿S DISCHARGE PIECES ARE BENT. AS A RESULT, THE OPERATOR ALWAYS SLIPPED OFF WHEN BENDING THE ROD. IT WAS UNKNOWN IF THE SURGERY COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN ROD (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS COMPLAINT INVOLVES (2) DEVICES. THIS REPORT IS FOR (1) SAGITTAL IN-SITU BENDER, RIGHT. THIS REPORT IS 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13171901 ↗MalfunctionNov 5, 2020Break; Break
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2020 DURING AN UNKNOWN PROCEDURE, THE SAGITTAL BENDERS BROKE ON THE TIP. THE X25 INSERTERS WERE ALSO ROUNDED AND CAUSED STRIPPING OF THE SET SCREWS. THERE WAS NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE. THE PROCEDURE OUTCOME WAS NOT REPORTED. THIS REPORT IS FOR ONE (1) SAGITTAL IN-SITU BENDER, RIGHT. THIS IS REPORT 2 OF 4 FOR (B)(4).
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10794382 ↗MalfunctionOct 10, 2019Break; Break
DURING AN AIS CORRECTION WITH EXPEDIUM VERSE DR. (B)(6) PERFORMED SOME IN-SITU BENDING OF THE COCR RODS AFTER THEY WERE FULLY SEATED IN THE SCREWS AND ALL SET SCREWS WERE FINAL TIGHTENED. DURING THE BENDING OF THE SECOND ROD THE IN-SITU BENDER BROCK. NO PATIENT HARM OCCURRED AND THE PROCEDURE COULD BE COMPLETED BY USING THE OTHER SIDE OF THE ROD BENDER. IN ADDITION, DURING THE SCREW LOADING ON THE SCREW DRIVER, IT DISASSEMBLED ITSELF. THE SCREW DRIVER SHAFT UNLOCKED ITSELF FROM THE OUTER SLEEVE WHEN LOADING THE SCREW. SPECIFICALLY THE SCREW DRIVER SHAFT GOT PUSHED OUT OF THE SLEEVE DISTALLY DURING THE LAST TURNS OF LOADING THE SCREW. THE NURSE WAS ABLE TO RE-CREATE THE DISASSEMBLY BY JUST PULLING AND SIMULTANEOUSLY ROTATING THE SCREW DRIVER SHAFT IN THE SLEEVE. IT SEEMS THE RETAINING MECHANISM OF THE OUTER SLEEVE IS DAMAGED / WORN. TO COMPLETE INSERTION OF THE SCREWS WE JUST USED THE 2ND SCREW DRIVER FROM THE SET. IF OTHER, DESCRIBE: REVISION SURGERY OF PREVIOUS AIS CORRECTION WITH EXPEDIUM DIFAR DUE TO PROGRESSING LUMBAR CURVE BELOW THE INITIAL CONSTRUCT., WAS SURGERY DELAYED DUE TO THE REPORTED EVENT?: NO, ACTION TAKEN WHEN EVENT OCCURRED?: USED THE OTHER SIDE OF THE INSTRUMENT,
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9174478 ↗MalfunctionFeb 6, 2019Break; Break
IT WAS REPORTED THAT THE SURGEON HAD 3 INSTRUMENTS BREAK DURING A T3-ILIUM REVISION SPINAL FUSION. ALL INSTRUMENTS BROKE WHILE THE SURGEON WAS USING THEM TO PREFORM VARIOUS FUNCTIONS IN THE CASE. WAS SURGERY DELAYED DUE TO THE REPORTED EVENT? UNKNOWN. WAS PROCEDURE SUCCESSFULLY COMPLETED? UNKNOWN. WERE FRAGMENTS GENERATED? UNKNOWN. IF YES, WERE THEY REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION? UNKNOWN. PATIENT STATUS/ OUTCOME / CONSEQUENCES? NO. WAS OTHER MEDICAL INTERVENTION (E.G. X-RAYS, ADDITIONAL PROCEDURES, PRESCRIPTIONS, OTC, REVISION) REQUIRED? NO. IS THE PATIENT PART OF A CLINICAL STUDY? UNKNOWN.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8309947 ↗
Other BENDER devices
- MONARCH 277030000Depuy Spine, LLC · BENDER15 reports
- EXPEDIUM 299704190Medos International Sàrl · BENDER9 reports
- EXPEDIUM 299704195Medos International Sàrl · BENDER9 reports
- EXPEDIUM 299704185Medos International Sàrl · BENDER8 reports
- ALTALYNE 5161-04-1700Medos International Sàrl · BENDER8 reports
- NA 03.161.057Synthes GmbH · BENDER5 reports
- EXPEDIUM 279763865Medos International Sàrl · BENDER4 reports
- SYMPHONY 2020-00-145Medos International Sàrl · BENDER4 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page