The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034509419), received between Jun 4, 2019 and Oct 11, 2024 — classified as 3 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Loose or Intermittent Connection” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loose or Intermittent Connection2
- Migration2
- Break1
- Compatibility Problem1
- Crack1
Patient problems
- Insufficient Information3
- No Consequences Or Impact To Patient2
- Failure of Implant1
- No Code Available1
- Pain1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 11, 2024Crack; Compatibility Problem; Crack; Compatibility Problem
IT WAS REPORTED THAT PATIENT UNDERWENT A & P LUMBAR FUSION ON (B)(6) 2024. DURING THE SURGERY, THE BAR IS NOT LOCKED WITH THE SLEEVE BECAUSE ONE OF THE PERCUTANEOUS SCREWS, HAD A BAD THREAD, AND IT HAS TO BE CHANGED TO OPEN SURGERY TO CHANGE IT. PROCEDURE WAS COMPLETED SUCCESSFULLY WITH A NINETY MINUTE DELAY. THIS REPORT IS FOR A VIPER PRIME CFXFEN XTAB 6X50MM.
Patient problems: Insufficient Information
MAUDE report 20427319 ↗InjuryJan 10, 2024Migration; Migration
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF. THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: IMPLANT MALPOSITION, KIDNEY / URINARY. COMPLICATION TIME/INTERVENTION POSTOPERATIVE (BEFORE DISCHARGE) THIS REPORT IS FOR ONE (1) VIPER PRIME CFXFEN XTAB 6X50MM. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18483718 ↗InjuryJan 5, 2024Migration; Migration
THIS REPORT IS BEING FILED AFTER THE REVIEW OF A CLINICAL EVALUATION REPORT (CER) FROM A DATABASE RELATED RESEARCH ACTIVITY (DRRA): SPINE TANGO SOA REPORT - TLIF THE FOLLOWING COMPLICATIONS HAVE BEEN IDENTIFIED: IMPLANT MALPOSITION, KIDNEY / URINARY, COMPLICATION TIME/INTERVENTION. POSTOPERATIVE (BEFORE DISCHARGE). THIS REPORT IS FOR ONE (1) VIPER PRIME CFXFEN XTAB 6X50MM. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).
Patient problems: Insufficient Information
MAUDE report 18460091 ↗InjurySep 29, 2020Break; Break
THE PREVIOUS HARDWARE WAS PARTIALLY REMOVED AND NEW HARDWARE WAS INSERTED TO TWO ADDITION LEVELS BILATERALLY. PORTION OF LEFT SIDED SCREWS WERE IMPACTED IN BONE AND COULD NOT BE RETRIEVED. THE BROKEN SCREWS WERE ENCASED IN BONE TOO ANTERIOR IN THE VERTEBRAL BODY; RETRIEVAL WOULD HAVE BEEN DAMAGING TO THE INTEGRITY OF THE ANATOMY. IT WAS ALSO REPORTED THE L4/L5 REGION WAS NOT FUSED AND STILL MOBILE AND THE PATIENT EXPERIENCED A LITTLE PAIN. IT WAS NOTED THE REVISION PROCEDURE WAS COMPLETED SUCCESSFULLY. THIS COMPLAINT ((B)(4)) CAPTURES THE POST-OPERATIVELY BROKEN SCREWS. RELATED COMPLAINT ((B)(4)) CAPTURES THE INTRA-OPERATIVE EVENT OF THE SCREWS IMPACTED IN BONE NOT BEING ABLE TO BE REMOVED.
Patient problems: Failure of Implant; No Code Available; Pain
MAUDE report 10600385 ↗MalfunctionJun 4, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ON (B)(6) 2019, THE THREE (3) VIPER PRIME SCREWS HAD TO BE REPLACED INTRAOPERATIVELY DUE TO LACK OF PULL OUT STRENGTH OR INAPPROPRIATE SCREW LENGTH. AFTER REPLACEMENT OF THE SCREWS, THE OR NURSE WANTED THEM TO BE STERILIZED FOR DEMONSTRATION PURPOSE. AFTER STERILIZATION, TWO (2) OF THE THREE (3) SCREWS FELL APART IN 3 PIECES. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THERE WERE NO PATIENT CONSEQUENCES. . THIS COMPLAINT INVOLVES THREE (3) DEVICES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8666958 ↗MalfunctionJun 4, 2019Loose or Intermittent Connection; Loose or Intermittent Connection
IT WAS REPORTED THAT ON (B)(6) 2019, THE THREE (3) VIPER PRIME SCREWS HAD TO BE REPLACED INTRAOPERATIVELY DUE TO LACK OF PULL OUT STRENGTH OR INAPPROPRIATE SCREW LENGTH. AFTER REPLACEMENT OF THE SCREWS, THE OR NURSE WANTED THEM TO BE STERILIZED FOR DEMONSTRATION PURPOSE. AFTER STERILIZATION, TWO (2) OF THE THREE (3) SCREWS FELL APART IN 3 PIECES. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. THERE WERE NO PATIENT CONSEQUENCES. THIS COMPLAINT INVOLVES THREE (3) DEVICES.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8666826 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗