Skip to content

VIPER PRIME 186770440Recalls & reported problems

Medos International Sàrl · VIPER PRIME CFX FEN X-TAB POLYAXIAL SCREW 5.5 7 x 40mm · Class 2 · Implant

All models of VIPER PRIME (60)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034509600), received between Mar 15, 2018 and Mar 15, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Disassembly” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Disassembly1

Patient problems

  • No Consequences Or Impact To Patient1
  • Not Applicable1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 15, 2018Disassembly; Disassembly

    VIPER PRIME CORTICAL FIX SCREW DISASSEMBLED AND PREVENTED ROD FROM REDUCING. THIS REQUIRED REMOVAL OF ROD AND 2 ADDITIONAL SCREWS TO CORRECT THE ISSUE. NEW SCREWS WERE USED. PATIENT CONSEQUENCE? :YES PATIENT CONSEQUENCE DESCRIPTION: ADDITIONAL INCISION SIZE WAS NEEDED AND APPROXIMATELY 45 MIN WAS ADDED TO PROCEDURE. IS THE INFORMATION BEING SUBMITTED FOR THIS COMPLAINT ALL THE DETAILS THAT ARE KNOWN/AVAILABLE REGARDING THIS EVENT? : YES. ADDITIONAL CLARIFICATION FROM SALES REP: THE INTERNAL MECHANISM OF THE POLY FEATURE WAS NO LONGER WHERE IT WAS SUPPOSED TO BE. THE OTHER SCREW HAD PART OF THE THREAD SHEAR OFF PREVENTING A SET SCREW FROM ENGAGING. THIS WAS NOTICED AFTER ROD REMOVAL AND ATTEMPTS TO REPLACE SET SCREW WERE UNSUCCESSFUL.

    Patient problems: No Consequences Or Impact To Patient; Not Applicable

    MAUDE report 7342347 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page