Skip to content

SYMPHONY 1020-24-002SRecalls & reported problems

Medos International Sàrl · SYMPHONY OCT SYSTEM CROSSCONNECTOR ROD TO ROD 26 to 32mm · Class 2 · Implant

All models of SYMPHONY (113)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034520209), received between May 13, 2022 and Jul 14, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device-Device Incompatibility2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 14, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THIS WAS AN UNKNOWN SURGERY PERFORMED ON (B)(6) 2025. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT ANY SURGICAL DELAY. ON (B)(6) 2025, THE SALES REP WAS INFORMED BY A SURGEON THAT ON (B)(6) SURGERY, THE CROSS CONNECTOR IN QUESTION INTERFERED WITH THE SCREW HEAD AND COULD NOT BE PLACED. ANOTHER CROSS CONNECTOR WAS CUT AND USED AND THE SURGERY WAS COMPLETED WITHOUT PROBLEM. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22502413 ↗
  • MalfunctionMay 13, 2022Device-Device Incompatibility; Material Deformation; Device-Device Incompatibility; Material Deformation

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED INTRAOPERATIVELY ON (B)(6) 2022 FOR A SPINE PCF (C3-5) PROCEDURE ON THAT THIS WAS A PCF (C3-5), A ROD WAS NOT ABLE TO BE DEPLOYED ON THE CONNECTOR. IT SEEMED THAT THE CONNECTOR¿S BASE AREA WHERE THE ROD WAS SUPPOSED TO BE PLACED MIGHT HAVE BEEN DEFORMED. AS A RESULT, THE SETSCREW BECAME IMMOVABLE. THE PROCEDURE WAS COMPLETED WITH REPLACEMENTS LESS THAN 30-MINUTE SURGICAL DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN RODS (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS IS REPORT 1 OF 1 FOR PC- (B)(4). THIS REPORT IS FOR CROSSCONNECTOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14395930 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page