Medos International Sàrl · VIPER PRIME CFX X-TAB POLYAXIAL SCREW 5.5 7 x 40mm · Class 2 · Implant

All models of VIPER PRIME (60)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10705034530178), received between Jun 17, 2020 and Mar 12, 2026 — classified as 8 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Device Slipped” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Slipped5
  • Break3
  • Device-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Material Deformation1
  • Migration1
  • Migration or Expulsion of Device1

Patient problems

  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • No Clinical Signs, Symptoms or Conditions4
  • Failure of Implant2
  • Insufficient Information1
  • No Consequences Or Impact To Patient1
  • Unspecified Nervous System Problem1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 12, 2026Device-Device Incompatibility; Migration

    IT WAS REPORTED THAT POST IBF UI H 9MM 4DEG 22/9 AND VIPER PRIME CFX XTAB 7X40MM TI WERE INVOLVED IN A REPORTED EVENT. FOLLOWING A POSTERIOR LUMBAR INTERBODY FUSION FOR LUMBAR SPINAL CANAL STENOSIS, THE CAGE BACKED OUT ONE WEEK AFTER SURGERY, NECESSITATING A REOPERATION. THE CAGE WAS REINSERTED AND ONE SCREW WAS REPLACED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITHOUT SURGICAL DELAY. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR OBSERVABLE CLINICAL SYMPTOMS, AND THE EVENT RESULTED IN A CORRECTIVE INVASIVE PROCEDURE WHERE A DEVICE WAS REPLACED OR REVISED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 24583449 ↗
  • InjuryJan 10, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT UNDERWENT A TRANSFORAMINAL LUMBAR INTERBODY FUSION (TLIF) (L5/S1) FOR LUMBAR CANAL STENOSIS. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH 60 MINUTES DELAY. SINCE THE PATIENT HAD NEUROLOGICAL SYMPTOMS CAUSED BY THE IMPLANTS AFTER SURGERY, AN REOPERATION WAS PERFORMED SUBSEQUENTLY. IN THE REOPERATION, 2 SCREWS IN QUESTION WERE REMOVED, NEW SCREWS WERE INSERTED WITH A CHANGE IN ENTRY POINT AND DIRECTION, AND RE-FIXATION WAS PERFORMED. AS THERE ARE NO SCREWDRIVERS THAT CAN APPLY TO VIPER PRIME SCREWS AFTER BREAKING OF THEIR TABS, NEW SCREWS WERE INSERTED AFTER REMOVING THE SCREWS IN QUESTION. NO FURTHER INFORMATION WAS PROVIDED.

    Patient problems: Unspecified Nervous System Problem

    MAUDE report 21123255 ↗
  • MalfunctionJul 19, 2022Device Slipped; Material Deformation; Device Slipped; Material Deformation

    DEVICE REPORT FROM JAPAN REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022 DURING TLIF FOR TREATING SPONDYLOLISTHESIS, WHEN THE SCREW IN QUESTION WAS INSERTED IN L5 LEFT, THE TIP OF THE GUIDEWIRE BECAME DEFORMED AND WAS NOT ABLE TO BE REMOVED FROM THE INCISION. DURING TROUBLESHOOTING, THE SCREW BECAME LOOSE. WHEN THE SCREW WAS REMOVED, IT WAS FOUND THAT THE TIP OF THE SCREW HAD BEEN DEFORMED. THE PROCEDURE WAS COMPLETED USING A REPLACEMENT SCREW AFTER A DELAY OF 10 MINUTES. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS A VIPER PRIME CFX X-TAB POLYAXIAL SCREW 5.5 7 X 40MM. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15057170 ↗
  • MalfunctionApr 20, 2022Break; Break

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT A SPINAL FUSION (L3-5) FOR TREATING BURST FRACTURE OCCURRED ON (B)(6) 2022. THE SURGEON DID NOT SUCCEED IN FINAL SETSCREW TIGHTENING AT L4 RIGHT. THE PROCEDURE WAS COMPLETED WITH A REPLACEMENT WHICH RESULTED IN LESS THAN 30-MINUTE SURGICAL DELAY. CONCOMITANT DEVICE REPORTED: UNKNOWN TIGHTENER (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS IS REPORT 1 OF 1 FOR (B)(4). THIS REPORT IS FOR VIPER PRIME SCREW.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14162366 ↗
  • InjuryFeb 16, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON, (B)(6) 2021 THE PATIENT UNDERWENT THE PRIMARY PROCEDURE (FOREIGN OBJECT REMOVAL AND SPINAL FUSION) RANGING L2/3. AFTER THE CAGE WAS DEPLOYED, THE SCREWS WERE PERCUTANEOUSLY INSERTED. THE PATIENT FELL OVER TWICE, WHICH RESULTED IN SCREW LOOSENING AND THE CAGE WAS BACKED OUT. ALSO, THE PATIENT HAD SUFFERED FROM PARALYSIS IN THE LOWER LIMB(S). ON (B)(6) 2022 THE PATIENT UNDERWENT A PROCEDURE TO REMOVE THE CAGE AND THE SCREW. THIS REPORT IS FOR ONE (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13545905 ↗
  • InjuryFeb 16, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON, (B)(6) 2021 THE PATIENT UNDERWENT THE PRIMARY PROCEDURE (FOREIGN OBJECT REMOVAL AND SPINAL FUSION) RANGING L2/3. AFTER THE CAGE WAS DEPLOYED, THE SCREWS WERE PERCUTANEOUSLY INSERTED. THE PATIENT FELL OVER TWICE, WHICH RESULTED IN SCREW LOOSENING AND THE CAGE WAS BACKED OUT. ALSO, THE PATIENT HAD SUFFERED FROM PARALYSIS IN THE LOWER LIMB(S). ON (B)(6) 2022 THE PATIENT UNDERWENT A PROCEDURE TO REMOVE THE CAGE AND THE SCREW. THIS REPORT IS FOR ONE (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS IS REPORT 4 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13545904 ↗
  • InjuryFeb 16, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON, (B)(6) 2021 THE PATIENT UNDERWENT THE PRIMARY PROCEDURE (FOREIGN OBJECT REMOVAL AND SPINAL FUSION) RANGING L2/3. AFTER THE CAGE WAS DEPLOYED, THE SCREWS WERE PERCUTANEOUSLY INSERTED. THE PATIENT FELL OVER TWICE, WHICH RESULTED IN SCREW LOOSENING AND THE CAGE WAS BACKED OUT. ALSO, THE PATIENT HAD SUFFERED FROM PARALYSIS IN THE LOWER LIMB(S). ON (B)(6) 2022 THE PATIENT UNDERWENT A PROCEDURE TO REMOVE THE CAGE AND THE SCREW. THIS REPORT IS FOR ONE (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS IS REPORT 1 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13545899 ↗
  • InjuryFeb 16, 2022Device Slipped; Device Slipped

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON, (B)(6) 2021 THE PATIENT UNDERWENT THE PRIMARY PROCEDURE (FOREIGN OBJECT REMOVAL AND SPINAL FUSION) RANGING L2/3. AFTER THE CAGE WAS DEPLOYED, THE SCREWS WERE PERCUTANEOUSLY INSERTED. THE PATIENT FELL OVER TWICE, WHICH RESULTED IN SCREW LOOSENING AND THE CAGE WAS BACKED OUT. ALSO, THE PATIENT HAD SUFFERED FROM PARALYSIS IN THE LOWER LIMB(S). ON (B)(6) 2022 THE PATIENT UNDERWENT A PROCEDURE TO REMOVE THE CAGE AND THE SCREW. THIS REPORT IS FOR ONE (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS IS REPORT 2 OF 4 FOR COMPLAINT (B)(4).

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 13545897 ↗
  • MalfunctionMar 16, 2021Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A PPS IN LATERAL POSITION AT L4-5 TREATING SPONDYLOLISTHESIS ON (B)(6) 2021. WHEN THE SURGEON INSERTED A SCREW ON THE VIPER PRIME, THE SCREW IDLY TURNED. THE SURGEON CONTINUED THE PROCEDURE WITH A REPLACEMENT STYLET AND SCREW. WHEN THE SURGEON EXPERIENCED DIFFICULTY APPLYING THE REPLACEMENT STYLET AND SCREW AGAINST THE BONE, THEY CHANGED THE SURGICAL TECHNIQUE TO J-PROBE+GUIDE WIRING AND COMPLETED INSERTION. THE PROCEDURE WAS COMPLETED WITH LESS THAN 30-MINUTE SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. CONCOMITANT MEDICAL PRODUCTS: UNKNOWN INSERTER (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY UNKNOWN). THIS REPORT IS FOR ONE (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11495299 ↗
  • InjuryDec 5, 2020Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT FOR PLIF ( POSTERIOR LUMBAR INTERBODY FUSION) AT L4-5 TREATING SPONDYLOLISTHESIS. DURING THE SURGERY, IT WAS FOUND THAT TWO SCREWS HAD COME OFF THE PEDICLE. THE LESION WAS DEPLOYED AGAIN, AND THE 1ST SCREW WAS REMOVED WITHOUT DIFFICULTY BUT DIFFICULT TO PICK UP THE 2ND SCREW. FINALLY, THE 2ND SCREW WAS REMOVED, AND THE SURGERY COMPLETED SUCCESSFULLY WITH 30 MINUTES DELAY. THE PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS REPORT IS 2 OF 2 FOR (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 10952296 ↗
  • InjuryDec 5, 2020Break; Break

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE PATIENT UNDERWENT FOR PLIF ( POSTERIOR LUMBAR INTERBODY FUSION) AT L4-5 TREATING SPONDYLOLISTHESIS. DURING THE SURGERY, IT WAS FOUND THAT TWO SCREWS HAD COME OFF THE PEDICLE. THE LESION WAS DEPLOYED AGAIN, AND THE 1ST SCREW WAS REMOVED WITHOUT DIFFICULTY BUT DIFFICULT TO PICK UP THE 2ND SCREW. FINALLY, THE 2ND SCREW WAS REMOVED, AND THE SURGERY COMPLETED SUCCESSFULLY WITH 30 MINUTES DELAY. THE PATIENT OUTCOME WAS UNKNOWN. THIS COMPLAINT INVOLVES TWO (2) DEVICES. THIS REPORT IS FOR (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS REPORT IS 1 OF 2 FOR (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 10952108 ↗
  • MalfunctionJun 17, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020 DURING A TLIF (TRANSFORAMINAL LUMBAR INTERBODY FUSION) IN THE L4 RIGHT TREATING LUMBAR SPINAL CANAL STENOSIS THAT AFTER THE SCREW (186770340S) WAS INSERTED, A ROD AND A SETSCREW WERE DEPLOYED. THE SCREW HAD COME OFF THE PEDICLE AND THE SCREWHEAD CAME OFF THE REST OF THE SCREW SECTION. THE PROCEDURE WAS COMPLETED LESS THAN 30-MINUTE SURGICAL DELAY. CONCOMITANT DEVICES: UNKNOWN SET SCREW (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). UNKNOWN RODS (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1) . THIS REPORT IS FOR ONE (1) VIPER PRIME CFX XTAB 7X40MM TI. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10164128 ↗

Other Thoracolumbosacral pedicle screw system devices

All Thoracolumbosacral pedicle screw system devices →