SYMPHONY 2020-00-244Recalls & reported problems
Medos International Sàrl · SYMPHONY OCT SYSTEM SHORT DRILL GUIDE ADJUSTABLE BODY 2.8MM 0-50mm +/- 0.43 percent · Class 1
All models of SYMPHONY (113)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034530758), received between Sep 13, 2022 and Sep 19, 2023 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility2
- Break1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 19, 2023Break; Break
DEVICE REPORT FROM DEPUY SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2023 DURING A POSTERIOR CERVICAL FUSION (C1-C2) FOR CERVICAL MYELOPATHY PROCEDURE, A NURSE ASSEMBLED NECESSARY INSTRUMENTS INCLUDING THE DRILL GUIDE BODY ACCORDING TO PROCEDURE MANUALS. DURING THE DISASSEMBLING THE DRILL GUIDE BODY AFTER FINISHING THE FINAL FIXATION, WHEN ATTEMPTING TO SEPARATE THE DRILL GUIDE SLEEVE FROM THE DRILL GUIDE GRADUATION ADJUSTMENT PORTION, THE WELDING ON THE DRILL STOPPER PORTION PEELED OFF AND THE DRILL GUIDE GRADUATION ADJUSTMENT PORTION WAS DAMAGED. THE DRILL GUIDE BODY HAD BEEN READY BUT NOT BEEN USED IN THE SURGERY, THEREFORE THERE WAS NO IMPACT ON THE PATIENT NOR THE SURGICAL TIME. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) ADJ DRILL GUIDE BODY 2.8MM. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17775011 ↗MalfunctionSep 13, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A POSTERIOR CERVICAL FUSION PERFORMED ON (B)(6) 2022. THE NURSE ASSEMBLED THE NAVIGATION DRILL GUIDE BODY IN QUESTION, BUT THE SILVER KNOB PART WAS PUSHED DOWN WHILE THE SCALE OF THE DRILL GUIDE WAS NOT SET TO "A". SO, THE DOCTOR PUSHED THE SILVER KNOB IN TO MAKE AN ADJUSTMENT, THEN INSTRUMENTS GOT STUCK WHERE IT SHOULDN'T HAVE BEEN. THE NAVIGATION DRILL GUIDE BODY COULD NO LONGER BE DISASSEMBLED. THERE WAS NO SURGICAL DELAY. NO FRAGMENTS WAS REPORTED. PATIENT STATUS/ OUTCOME? :STABLE NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) ADJ DRILL GUIDE BODY 2.8MM. THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15411650 ↗MalfunctionSep 13, 2022Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN JAPAN AS FOLLOWS: IT WAS REPORTED THAT THIS WAS A POSTERIOR CERVICAL FUSION PERFORMED ON (B)(6) 2022. THE NURSE ASSEMBLED THE NAVIGATION DRILL GUIDE BODY IN QUESTION, BUT THE SILVER KNOB PART WAS PUSHED DOWN WHILE THE SCALE OF THE DRILL GUIDE WAS NOT SET TO "A". SO, THE DOCTOR PUSHED THE SILVER KNOB IN TO MAKE AN ADJUSTMENT, THEN INSTRUMENTS GOT STUCK WHERE IT SHOULDN'T HAVE BEEN. THE NAVIGATION DRILL GUIDE BODY COULD NO LONGER BE DISASSEMBLED. THERE WAS NO SURGICAL DELAY. NO FRAGMENTS WAS REPORTED. PATIENT STATUS/ OUTCOME? :STABLE NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR ONE (1) ADJ DRILL GUIDE BODY 2.8MM THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15411649 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page