EXPEDIUM 3010-19-003Recalls & reported problems

Medos International Sàrl · NAVIGATION ENABLED INSTRUMENTS EXPEDIUM DRIVER - SHAFT T20 5.5 · Class 2

All models of EXPEDIUM (291)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Expedium navigated surgical driver has 132 reports, most describing parts that break, bend or wear, such as shafts, sleeves and handles. Some describe a driver tip breaking off inside a screw. The reports mention 8 injuries and no deaths, and no recalls name this device. Reports fell to 7 in the last 12 months from 25 in the previous 12.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
133
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 133 adverse event reports that list this exact device (DI 10705034561387), received between Jan 15, 2021 and Sep 23, 2026 — classified as 8 injuries, 125 malfunctions.

The most frequently reported specific device problem is “Break” (95 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 2 reportsJan 2024: 6 reportsFeb 2024: 2 reportsMar 2024: 2 reportsApr 2024: 5 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 6 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 3 reportsDec 2024: 6 reportsJan 2025: 6 reportsFeb 2025: 0 reportsMar 2025: 8 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 4 reportsDec 2025: 2 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break95
  • Material Twisted/Bent37
  • Device-Device Incompatibility16
  • Entrapment of Device14
  • No Apparent Adverse Event2
  • Device Slipped1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions118
  • Foreign Body In Patient14
  • Insufficient Information4

Recent adverse event reports

The 25 most recent of 133 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 23, 2026Break

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50873873 ↗
  • MalfunctionFeb 23, 2026Break

    IT WAS REPORTED THAT WHEN SURGEON, WORKING WITH VERY SCLEROTIC BONE, USED A MALLET ON NAVIGATED INSTRUMENTS, RESULTING IN THE HANDLE BREAKING. WHILE INSERTING A 100MM ILIAC CREST BOLT, HIGH SHEAR FORCES CAUSED THE TIP OF THE DRIVER TO BREAK OFF IN THE SCREW. THE SCREWDRIVER WAS REPLACED, AND THE INSERTED SCREW WAS REMOVED AND REPLACED WITH A NEW ONE. THE PROCEDURE WAS DELAYED BY 10 MINUTES DUE TO THE EVENT, BUT WAS SUCCESSFULLY COMPLETED. FRAGMENTS WERE GENERATED AND REMOVED EASILY WITHOUT ADDITIONAL INTERVENTION. THERE WAS NO REPORTED CONSEQUENCE OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24421002 ↗
  • MalfunctionDec 30, 2025Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED AND STATES THAT: THE SHAFT BROKE AS WELL AS THE SLEEVE, THE MISSING CODE SHOULD HAVE BEEN ADDED AS WELL SINCE THE 2 PARTS GOES TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23929314 ↗
  • MalfunctionDec 30, 2025Break; Break

    ADDITIONAL INFORMATION WAS RECEIVED: THE SHAFT BROKE AS WELL AS THE SLEEVE, THE MISSING CODE SHOULD HAVE BEEN ADDED AS WELL SINCE THE 2 PARTS GOES TOGETHER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23929300 ↗
  • MalfunctionNov 18, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE SCREW DRIVERS APPEAR TO BE WEARING AT THE DISTAL TIP AND NOT ENGAGING WITH SCREW HEAD FULLY. BUTTONS ON THE SLEEVES ARE BECOMING WORN AS WELL AND DRIVER T HANDLES ARE SQUEAKING MORE THAN USUAL AND DIFFICULT TO CONNECT TO SHAFT. ENTIRE DRIVER TOOL NEEDS REPLACING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23587059 ↗
  • MalfunctionNov 18, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE SCREW DRIVERS APPEAR TO BE WEARING AT THE DISTAL TIP AND NOT ENGAGING WITH SCREW HEAD FULLY. BUTTONS ON THE SLEEVES ARE BECOMING WORN AS WELL AND DRIVER T HANDLES ARE SQUEAKING MORE THAN USUAL AND DIFFICULT TO CONNECT TO SHAFT. ENTIRE DRIVER TOOL NEEDS REPLACING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23587057 ↗
  • MalfunctionNov 18, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE SCREW DRIVERS APPEAR TO BE WEARING AT THE DISTAL TIP AND NOT ENGAGING WITH SCREW HEAD FULLY. BUTTONS ON THE SLEEVES ARE BECOMING WORN AS WELL AND DRIVER T HANDLES ARE SQUEAKING MORE THAN USUAL AND DIFFICULT TO CONNECT TO SHAFT. ENTIRE DRIVER TOOL NEEDS REPLACING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586989 ↗
  • MalfunctionNov 18, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT THE SCREW DRIVERS APPEAR TO BE WEARING AT THE DISTAL TIP AND NOT ENGAGING WITH SCREW HEAD FULLY. BUTTONS ON THE SLEEVES ARE BECOMING WORN AS WELL AND DRIVER T HANDLES ARE SQUEAKING MORE THAN USUAL AND DIFFICULT TO CONNECT TO SHAFT. ENTIRE DRIVER TOOL NEEDS REPLACING. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23586987 ↗
  • MalfunctionOct 27, 2025Break; Break

    IT WAS REPORTED THAT THE EQUIPMENT WAS BROKEN. DURING A SPINE PROCEDURE. SURGERY WAS NOT DELAYED DUE TO THE EVENT AND WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23389486 ↗
  • MalfunctionJun 16, 2025Device-Device Incompatibility; Device-Device Incompatibility

    IT WAS REPORTED THAT THE STARS ON ONE INSTRUMENTS WAS NOT MOVING FREELY. A FEW THINGS WERE TRIED, BUT WAS NOTICED THAT IT WAS THE SCREWDRIVERS AND NOT THE STARS. THE STAR WAS PLACED ON ANOTHER INSTRUMENT AND IT WORKED PERFECTLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22213006 ↗
  • MalfunctionMar 28, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NAVIGATION DRIVER 5.5MM 301019003 X2 HAS BECAME DULL AT THE TIP (REDACTED), NO CASE OR PATIENT IMPACTED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21719973 ↗
  • MalfunctionMar 28, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE NAVIGATION DRIVER 5.5MM 301019003 X2 HAS BECAME DULL AT THE TIP (REDACTED), NO CASE OR PATIENT IMPACTED. NO PATIENT INVOLVEMENT,

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21719970 ↗
  • MalfunctionMar 17, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2025, DURING A HARDWARE REMOVAL IT WAS NOTICED THAT THE DRIVER TIP WAS ROUNDED OFF AND WAS SLIPPING IN THE HEAD OF THE SCREW. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE DRIVER TIP WAS IN TACK. JUST ROUNDED AND WOULD NOT GRIP THE INSIDE OF THE SCREW. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21620703 ↗
  • MalfunctionMar 17, 2025Device-Device Incompatibility; Material Twisted/Bent; Device-Device Incompatibility; Material Twisted/Bent

    IT WAS REPORTED THAT ON (B)(6) 2025, DURING A HARDWARE REMOVAL IT WAS NOTICED THAT THE DRIVER TIP WAS ROUNDED OFF AND WAS SLIPPING IN THE HEAD OF THE SCREW. THERE WAS NO SURGERY DELAYED DUE TO THE REPORTED EVENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE DRIVER TIP WAS IN TACK. JUST ROUNDED AND WOULD NOT GRIP THE INSIDE OF THE SCREW. THERE WAS NO PATIENT CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21620653 ↗
  • MalfunctionMar 7, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DRIVER HAS BECOME BLUNT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21553619 ↗
  • MalfunctionMar 7, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DRIVER HAS BECOME BLUNT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21553605 ↗
  • MalfunctionMar 7, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DRIVER HAS BECOME BLUNT. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21553600 ↗
  • MalfunctionMar 6, 2025Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED THAT THE DOCTOR WAS INSERTING AN S2 ALAR SCREW CROSSING INTO THE PELVIS AND ON HER FINAL TURN OF THE DRIVER, TIP OF DRIVER SHEARED OFF INTO THE SCREW. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. THE FRAGMENT GENERATED WAS RETAINED WITHIN THE SCREW. THE SURGEONS ATTEMPTED TO RETRIEVE THE FRAGMENT FROM THE SCREW BUT WERE UNSUCCESSFUL. THERE WAS NO PATIENT CONSEQUENCE. NO ADDITIONAL MEDICAL INTERVENTION WAS REQUIRED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21535136 ↗
  • MalfunctionJan 21, 2025Break; Break

    IT WAS REPORTED THAT DUE TO THE CANNULATION, THE DRIVER'S BREAK OFF AT THE END EASILY. MULTIPLE DRIVERS WERE BROKEN DURING THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21196647 ↗
  • MalfunctionJan 21, 2025Break; Break

    IT WAS REPORTED THAT DUE TO THE CANNULATION, THE DRIVER'S BREAK OFF AT THE END EASILY. MULTIPLE DRIVERS WERE BROKEN DURING THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21196550 ↗
  • MalfunctionJan 21, 2025Break; Break

    IT WAS REPORTED THAT DUE TO THE CANNULATION, THE DRIVER'S BREAK OFF AT THE END EASILY. MULTIPLE DRIVERS WERE BROKEN DURING THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21196512 ↗
  • MalfunctionJan 21, 2025Break; Break

    IT WAS REPORTED THAT DUE TO THE CANNULATION, THE DRIVER'S BREAK OFF AT THE END EASILY. MULTIPLE DRIVERS WERE BROKEN DURING THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21196473 ↗
  • MalfunctionJan 21, 2025Break; Break

    IT WAS REPORTED THAT DUE TO THE CANNULATION, THE DRIVER'S BREAK OFF AT THE END EASILY. MULTIPLE DRIVERS WERE BROKEN DURING THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21196421 ↗
  • MalfunctionJan 8, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE ITEMS WERE WORN AND STRIPPED. SURGERY WAS COMPLETED SUCCESSFULLY USING THE INSTRUMENTS, AND THERE WAS NO IMPACT TO PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21105963 ↗
  • MalfunctionDec 12, 2024Break; Entrapment of Device; Break; Entrapment of Device

    IT WAS REPORTED WHEN THE SURGEON WAS INSERTING THE S2 AI SCREW INTO THE PATIENT, THE DISTAL TIP OF THE SCREWDRIVER SHEARED OFF IN THE HEAD OF THE PEDICLE SCREW. THE SHEARED OFF FRAGMENT REMAINED IN THE HEAD OF THE SCREW AND THE SURGEON DID NOT REMOVE THE SCREW. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY AND NO CONSEQUENCE TO THE PATIENT.

    Patient problems: Foreign Body In Patient

    MAUDE report 20916241 ↗

Other Orthopedic stereotaxic instrument devices