TRIALTIS 5161-00-5069Recalls & reported problems
Medos International Sàrl · TRIALTIS Spine System Screwdriver,T27 Navigated Cannulated · Class 2
All models of TRIALTIS (46)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034589282), received between Aug 15, 2024 and Jun 16, 2025 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Twisted/Bent3
- Break2
- Material Deformation1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2025Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent
IT WAS REPORTED THAT THE DRIVER TIP WAS STRIPPED WHEN ASSEMBLING TRAYS AFTER A SURGERY. NO COMPLAINTS WERE MADE DURING THE CASE AND PATIENT WAS NOT AFFECTED IN ANY WAY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22214126 ↗MalfunctionJun 2, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT DURING A SPINAL FUSION PROCEDURE, WHILE THE INSERTING AN 8.0MM ILIAC SCREW, THE OUTER SLEEVE WAS ACCIDENTALLY DEPRESSED AND DISENGAGED FROM THE SCREW HEAD. THE NAVIGATED SCREWDRIVER THEN STRIPPED AT THE TEETH OF THE INSTRUMENT. SURGERY WAS NOT DELAYED AND WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCE. THE ITEM WAS REMOVED FROM USE. THE SCREW WAS NOT DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22124350 ↗MalfunctionJan 30, 2025Material Twisted/Bent; Material Twisted/Bent
IT WAS REPORTED THAT ON AN UNKNOWN DATE THE ATRIALIS NAV DRIVER WAS STRIPPED. A REPLACEMENT IS REQUIRED AT THE HOSPITAL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21272536 ↗MalfunctionDec 31, 2024Break; Break
IT WAS REPORTED THAT THE TIP OF SCREWDRIVER BROKE WHEN ATTEMPTING TO INSERT A SCREW FURTHER DURING A SPINE PROCEDURE. A 10X90MM SCREW WAS INSERTED. SURGEON ATTEMPTED TO GO BACK WITH SCREWDRIVER TO INSERT SCREW FURTHER, AND THE TIP BROKE. THE TIP WAS REMOVED FROM SCREW. THERE WAS A SURGICAL DELAY OF 2 MINUTES. FRAGMENT WAS GENERATED AND WAS EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME WAS UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21046730 ↗MalfunctionAug 15, 2024Break; Break
IT WAS REPORTED THAT ON (B)(6) 2024, THE T27 DRIVER SHAFT TIP BROKE OFF WHILE INSERTING A 9.0MM X 55MM SCREW IN THE PELVIS. THIS WAS A REVISION OF A PREVIOUSLY PLACED CONSTRUCT. A 8.0MM X 55MM SCREW WAS REMOVED AND THE TRIALTIS 9.0MM X 55MM SCREW WAS INSERTED IN THE SAME TRAJECTORY. WHEN THE SCREW WAS IN 75% OF THE WAY, THE TIP OF THE DRIVER SHAFT BROKE OFF. THE SCREW WAS LEFT IN AT THIS POINT AND THE CONSTRUCT WAS COMPLETED WITHOUT ANY ADDITIONAL INCIDENTS. THERE WAS SURGICAL DELAY OF 5 MINS. THE SCREW WAS LEFT PROUD AND ATTACHED A CONNECTOR TO COMPLETE THE CONSTRUCT. NO PATIENT STATUS/ OUTCOME / CONSEQUENCES OCCURRED. THIS REPORT IS FOR ONE (1) SCREWDRIVER, T27 NAV CANN. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19996325 ↗
Other Orthopedic stereotaxic instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page