TRIALTIS 5161-00-5069Recalls & reported problems

Medos International Sàrl · TRIALTIS Spine System Screwdriver,T27 Navigated Cannulated · Class 2

All models of TRIALTIS (46)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034589282), received between Aug 15, 2024 and Jun 16, 2025 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent3
  • Break2
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 16, 2025Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT THE DRIVER TIP WAS STRIPPED WHEN ASSEMBLING TRAYS AFTER A SURGERY. NO COMPLAINTS WERE MADE DURING THE CASE AND PATIENT WAS NOT AFFECTED IN ANY WAY. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22214126 ↗
  • MalfunctionJun 2, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A SPINAL FUSION PROCEDURE, WHILE THE INSERTING AN 8.0MM ILIAC SCREW, THE OUTER SLEEVE WAS ACCIDENTALLY DEPRESSED AND DISENGAGED FROM THE SCREW HEAD. THE NAVIGATED SCREWDRIVER THEN STRIPPED AT THE TEETH OF THE INSTRUMENT. SURGERY WAS NOT DELAYED AND WAS SUCCESSFULLY COMPLETED WITH NO PATIENT CONSEQUENCE. THE ITEM WAS REMOVED FROM USE. THE SCREW WAS NOT DAMAGED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22124350 ↗
  • MalfunctionJan 30, 2025Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT ON AN UNKNOWN DATE THE ATRIALIS NAV DRIVER WAS STRIPPED. A REPLACEMENT IS REQUIRED AT THE HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21272536 ↗
  • MalfunctionDec 31, 2024Break; Break

    IT WAS REPORTED THAT THE TIP OF SCREWDRIVER BROKE WHEN ATTEMPTING TO INSERT A SCREW FURTHER DURING A SPINE PROCEDURE. A 10X90MM SCREW WAS INSERTED. SURGEON ATTEMPTED TO GO BACK WITH SCREWDRIVER TO INSERT SCREW FURTHER, AND THE TIP BROKE. THE TIP WAS REMOVED FROM SCREW. THERE WAS A SURGICAL DELAY OF 2 MINUTES. FRAGMENT WAS GENERATED AND WAS EASILY REMOVED WITHOUT REQUIRING ADDITIONAL INTERVENTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. PATIENT OUTCOME WAS UNKNOWN.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21046730 ↗
  • MalfunctionAug 15, 2024Break; Break

    IT WAS REPORTED THAT ON (B)(6) 2024, THE T27 DRIVER SHAFT TIP BROKE OFF WHILE INSERTING A 9.0MM X 55MM SCREW IN THE PELVIS. THIS WAS A REVISION OF A PREVIOUSLY PLACED CONSTRUCT. A 8.0MM X 55MM SCREW WAS REMOVED AND THE TRIALTIS 9.0MM X 55MM SCREW WAS INSERTED IN THE SAME TRAJECTORY. WHEN THE SCREW WAS IN 75% OF THE WAY, THE TIP OF THE DRIVER SHAFT BROKE OFF. THE SCREW WAS LEFT IN AT THIS POINT AND THE CONSTRUCT WAS COMPLETED WITHOUT ANY ADDITIONAL INCIDENTS. THERE WAS SURGICAL DELAY OF 5 MINS. THE SCREW WAS LEFT PROUD AND ATTACHED A CONNECTOR TO COMPLETE THE CONSTRUCT. NO PATIENT STATUS/ OUTCOME / CONSEQUENCES OCCURRED. THIS REPORT IS FOR ONE (1) SCREWDRIVER, T27 NAV CANN. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19996325 ↗

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →