NA SD312.025Recalls & reported problems
Synthes GmbH · DRILL/TAP AND SCREW GUIDE WITH POST-SHORT · Class 1
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034705804), received between Dec 12, 2016 and Dec 12, 2016 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break1
- Material Fragmentation1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2016Break; Material Fragmentation; Break; Material Fragmentation
IT WAS REPORTED THAT DURING THE INITIAL ANTERIOR CERVICAL FUSION SURGERY PERFORMED ON (B)(6) 2016 FOR C5-6 & C6-7 LEVELS TO TREAT CERVICAL STENOSIS THE CUSTOM DRILL/TAP AND SCREW (DTS) GUIDE WITH POST SHORT BROKE OFF. THE GUIDE HAS A SMALL POST ON ITS TIP AND IT GOES IN THE PLATE HOLE TO ALIGN THE PLATE FOR SCREW INSERTION. AFTER SCREW INSERTION WHEN SURGEON WAS PULLING OUT THE GUIDE FROM THE PLATE, 1MM TINY LITTLE POST OF THE GUIDE BROKE OFF. THE BROKEN PIECE WAS EASILY RETRIEVED. NO FRAGMENTS REMAINED IN THE PATIENT WAS CONFIRMED THROUGH THE ROUTINE X-RAYS TAKEN DURING THE PROCEDURE. SURGEON USED THE LONG DRILL GUIDE TO COMPLETE THE PROCEDURE. THERE WAS ABOUT 1 MINUTE SURGICAL DELAY DUE TO REPORTED EVENT. SURGERY WAS COMPLETED SUCCESSFULLY WITH NO PATIENT HARM, NO ADDITIONAL MEDICAL INTERVENTION. PATIENT OUTCOME POST-SURGERY REPORTED AS STABLE. CONCOMITANT DEVICE REPORTED AS: TITANIUM ANTERIOR CERVICAL COMPRESSION PLATE (PART # 450.536, LOT # UNKNOWN, QUANTITY 1), UNKNOWN SCREW (PART # UNKNOWN, LOT # UNKNOWN, QUANTITY UNKNOWN). THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6166001 ↗
Other Guide, surgical, instrument devices
- NA 03.037.017Synthes GmbH · Guide, surgical, instrument346 reports
- EVOS 71174934Smith & Nephew, Inc. · Guide, surgical, instrument149 reports
- LCP 323.042Synthes GmbH · Guide, surgical, instrument129 reports
- DJO SURGICAL 342-10-705Encore Medical, L.P. · Guide, Surgical, Instrument1 recall51 reports
- DJO SURGICAL 341-10-707Encore Medical, L.P. · Guide, Surgical, Instrument81 reports
- NA 03.010.404Synthes GmbH · Guide, surgical, instrument73 reports
- DJO SURGICAL 341-14-708Encore Medical, L.P. · Guide, Surgical, Instrument1 recall22 reports
- DJO SURGICAL 341-10-711Encore Medical, L.P. · Guide, Surgical, Instrument1 recall16 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.