Skip to content

NA 03.600.006Recalls & reported problems

Synthes GmbH · PLATE HOLDING DRILL GUIDE- SINGLE BARREL/FIXED ANGLE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034711799), received between Jan 29, 2019 and Jan 29, 2019 — classified as 1 malfunction.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 29, 2019Break; Break

    IT WAS REPORTED THAT DURING A ROUTINE INSPECTION ON (B)(6) 2019, A PLATE HOLDING DRILL GUIDE BROKE APART WHILE TESTING IT. THE SALES CONSULTANT STATED THAT THERE WAS NOT ANY FORCE APPLIED TO IT. THEY DO NOT KNOW IF IT SUSTAINED A BLOW AND JUST BARELY CLUNG TOGETHER UNTIL THAT MOMENT. THE SALES CONSULTANT SQUEEZED IT WHEN TESTING THE LOCK FUNCTION OF THIS INSTRUMENT. ALSO, IT WAS NOTICED THAT THE THREADS ON THE DISTAL END OF THE INNER SHAFT OF THE SCREW REMOVAL TOOL WERE DAMAGED. THE INNER SHAFT WOULD NOT SCREW INTO A SCREW. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE: EXTRACTION SCREWDRIVER (PART #: 352.311, LOT #: UNKNOWN, QUANTITY: 1) THIS REPORT IS FOR ONE (1) PLATE HOLDING DRILL GUIDE- SINGLE BARREL/FIXED ANGLE THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8286372 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page