Skip to content

NA 03.617.902Recalls & reported problems

Synthes GmbH · STARDRIVE SCREWDRIVER SHAFT T8 SELF-RETAINING/QC · Class 2

Quick summary

This Synthes StarDrive screwdriver shaft is a surgical driver used in spine surgery. Its 102 FDA reports include 3 injuries and no deaths, and mostly describe worn, dull, or broken screwdriver tips, with some reports of bent parts and stripped drivers. Reports rose to 16 in the last 12 months from none in the 12 months before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
103
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 103 adverse event reports that list this exact device (DI 10705034717166), received between Dec 15, 2017 and Sep 21, 2026 — classified as 3 injuries, 100 malfunctions.

The most frequently reported specific device problem is “Material Integrity Problem” (57 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 3 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 2 reportsFeb 2026: 4 reportsMar 2026: 0 reportsApr 2026: 2 reportsMay 2026: 8 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Integrity Problem57
  • Material Twisted/Bent23
  • Compatibility Problem10
  • Device Operates Differently Than Expected6
  • Manufacturing, Packaging or Shipping Problem6
  • Mechanical Problem6
  • Naturally Worn6
  • Break5

Patient problems

  • No Clinical Signs, Symptoms or Conditions54
  • No Consequences Or Impact To Patient31
  • No Known Impact Or Consequence To Patient13
  • Foreign Body In Patient4
  • Insufficient Information3
  • No Code Available2
  • No Patient Involvement1

Recent adverse event reports

The 25 most recent of 103 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 21, 2026Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50839784 ↗
  • MalfunctionMay 29, 2026Material Integrity Problem

    IT WAS REPORTED THE TIPS OF THE SCREWDRIVERS ARE WORN DOWN. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25330690 ↗
  • MalfunctionMay 29, 2026Material Integrity Problem

    IT WAS REPORTED THE TIPS OF THE SCREWDRIVERS ARE WORN DOWN. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25330681 ↗
  • MalfunctionMay 29, 2026Material Integrity Problem

    IT WAS REPORTED THE TIPS OF THE SCREWDRIVERS ARE WORN DOWN. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25330648 ↗
  • MalfunctionMay 29, 2026Material Integrity Problem

    IT WAS REPORTED THE TIPS OF THE SCREWDRIVERS ARE WORN DOWN. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25330641 ↗
  • MalfunctionMay 6, 2026Material Integrity Problem

    IT WAS REPORTED THAT DRIVER STRIPPED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25096448 ↗
  • MalfunctionMay 5, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE SCREWDRIVERS HAD DULL/BROKEN TIPS. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25084360 ↗
  • MalfunctionMay 5, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE SCREWDRIVERS HAD DULL/BROKEN TIPS. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25084356 ↗
  • MalfunctionMay 5, 2026Material Integrity Problem

    IT WAS REPORTED THAT THE SCREWDRIVERS HAD DULL/BROKEN TIPS. THERE WAS NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25084354 ↗
  • MalfunctionApr 24, 2026Material Integrity Problem

    IT WAS REPORTED THAT WHILE TORQUING THE FOUR SCREWS ON THE ZERO-P HOUSING, IT WAS NOTICED THAT THE TIPS OF BOTH SCREWDRIVERS WERE SEVERELY WORN; CONSEQUENTLY, THEY LACKED SUFFICIENT GRIP TO PROPERLY APPLY TORQUE. THE SCREWS WERE TIGHTENED AS MUCH AS POSSIBLE WITHOUT CAUSING FURTHER DAMAGE TO THEIR HEADS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24984351 ↗
  • MalfunctionApr 24, 2026Material Integrity Problem

    IT WAS REPORTED THAT WHILE TORQUING THE FOUR SCREWS ON THE ZERO-P HOUSING, IT WAS NOTICED THAT THE TIPS OF BOTH SCREWDRIVERS WERE SEVERELY WORN; CONSEQUENTLY, THEY LACKED SUFFICIENT GRIP TO PROPERLY APPLY TORQUE. THE SCREWS WERE TIGHTENED AS MUCH AS POSSIBLE WITHOUT CAUSING FURTHER DAMAGE TO THEIR HEADS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24984332 ↗
  • MalfunctionFeb 6, 2026Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE DRIVERS HAD STRIPPED ON THE SET AND WOULD NOT TIGHTEN THE SCREW. TO COMPLETE THE PROCEDURE A T8 DRIVER WAS OPENED FROM ANOTHER TRAUMA SET. THERE WAS NO IMPACT TO PATIENT OR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24283913 ↗
  • MalfunctionFeb 6, 2026Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE DRIVERS HAD STRIPPED ON THE SET AND WOULD NOT TIGHTEN THE SCREW. TO COMPLETE THE PROCEDURE A T8 DRIVER WAS OPENED FROM ANOTHER TRAUMA SET. THERE WAS NO IMPACT TO PATIENT OR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24283911 ↗
  • MalfunctionFeb 4, 2026Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE DRIVERS HAD STRIPPED ON THE SET AND WOULD NOT TIGHTEN THE SCREW. THE T8 DRIVER FROM ANOTHER SET WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT OR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24261199 ↗
  • MalfunctionFeb 4, 2026Mechanical Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE DRIVERS HAD STRIPPED ON THE SET AND WOULD NOT TIGHTEN THE SCREW. THE T8 DRIVER FROM ANOTHER SET WAS OPENED TO COMPLETE THE PROCEDURE. THERE WAS NO IMPACT TO THE PATIENT OR PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24261191 ↗
  • MalfunctionJan 21, 2026Material Integrity Problem

    IT WAS REPORTED TWO (2) SCREWDRIVER SHAFTS WERE NOTED TO BE STRIPPED DURING AN ACDF. SURGEON STRUGGLED TO ADVANCE THE SCREWS WITH THE SCREWDRIVERS ON THE SET. AN ALTERNATIVE KIT WITH THE SAME DRIVE-FEATURE WAS QUICKLY OPENED AND THE CASE WENT AHEAD SUCCESSFULLY. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24134087 ↗
  • MalfunctionJan 21, 2026Material Integrity Problem

    IT WAS REPORTED TWO (2) SCREW DRIVER SHAFTS WERE NOTED TO BE STRIPPED DURING AN ACDF. SURGEON STRUGGLED TO ADVANCE THE SCREWS WITH THE SCREWDRIVERS ON THE SET. AN ALTERNATIVE KIT WITH THE SAME DRIVE-FEATURE WAS QUICKLY OPENED AND THE CASE WENT AHEAD SUCCESSFULLY. NO PATIENT HARM. NO SURGICAL DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24134078 ↗
  • MalfunctionSep 20, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON SEPTEMBER 12, 2024, THE TIP OF DRIVER WORE OUT. NO PATIENT HARM, DELAY, NO FRAGMENT. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER SHAFT T8 SELF-RETAINING/QC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20269331 ↗
  • MalfunctionSep 20, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2024 , THE TIP OF DRIVER WORE OUT. NO PATIENT HARM, DELAY, NO FRAGMENT. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER SHAFT T8 SELF-RETAINING/QC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20269322 ↗
  • MalfunctionSep 18, 2024Activation, Positioning or Separation Problem; Material Integrity Problem; Activation, Positioning or Separation Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON (B)(6) 2024 , THE SURGEON WAS DOING A 3 - LEVEL ACDF USING ZERO-P NATURAL PLATE. AFTER IMPLANTING A 7MM ZERO-P NATURAL PLATE AND SPACER AT C6-C7, THEY PLACED 14MM SCREWS. THE TORQUE LIMITING DEVICE WAS NOT PLACED ON THE STRAIGHT SCREW DRIVER RESULTING IN BREAKING OFF THE DRIVER TIP IN THE SCREW HEAD. THE SCREW WAS FULLY PLACED, BUT THE DRIVER TIP WOULD NOT COME OUT. AFTER SOME TIME TRYING TO REMOVE THE BROKEN DRIVER TIP WITH NO SUCCESS, THE SURGEON MOVED ON TO FINISH THE CASE SUCCESSFULLY. THERE WAS AN UNKNOWN SURGICAL DELAY. PATIENT IS DOING GREAT AND IS RECOVERING NICELY. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER SHAFT T8 SELF-RETAINING/QC.

    Patient problems: Foreign Body In Patient

    MAUDE report 20249126 ↗
  • MalfunctionJul 29, 2024Compatibility Problem; Material Integrity Problem; Compatibility Problem; Material Integrity Problem

    IT WAS REPORTED DURING THE SURGERY THE HOLDING SLEEVE PARTS OF THE T8 SELF-FASTENING SCREWDRIVER, ARE IN POOR CONDITION DO NOT ALLOW TORQUE TO THE LOCKED SCREWS OF THE PLATE OF THE ZERO-P SYSTEM. THE SURGEON COULD NOT PERFORM THE RESPECTIVE TORQUE TO THE SCREWS, HOWEVER THE SURGERY WAS SUCCESSFULLY COMPLETED, WITHOUT ALTERATIONS IN THE SURGICAL TIMES AND WITHOUT AFFECTING THE PATIENT. THIS REPORT IS FOR HOLDING SLEEVE FOR STARDRIVE SCREWDRIVER SHAFT T8 THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 19850123 ↗
  • MalfunctionFeb 6, 2024Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2024, THAT BOTH SCREWDRIVER TIPS WERE ROLLED, AND THAT MADE SURGERY DIFFICULT. THE DOCTOR WAS VERY UPSET; IT WAS NOT POSSIBLE TO TORQUE THE LAST SCREW, AND THE OTHERS WERE DIFFICULT BECAUSE OF THE BAD GRIP OF THE TIPS. PLEASE REVIEW AND CHANGE IT SO AS NOT TO HAVE DIFFICULTY WITH IT. THIS REPORT IS FOR ONE (1) SCRDRIVER SHAFT T8 SELF-HOLDING. THIS IS REPORT 1 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18649499 ↗
  • MalfunctionFeb 6, 2024Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN COLOMBIA AS FOLLOWS: IT WAS REPORTED ON (B)(6)2024, THAT BOTH SCREWDRIVER TIPS WERE ROLLED, AND THAT MADE SURGERY DIFFICULT. THE DOCTOR WAS VERY UPSET; IT WAS NOT POSSIBLE TO TORQUE THE LAST SCREW, AND THE OTHERS WERE DIFFICULT BECAUSE OF THE BAD GRIP OF THE TIPS. PLEASE REVIEW AND CHANGE IT SO AS NOT TO HAVE DIFFICULTY WITH IT. THIS REPORT IS FOR ONE (1) SCRDRIVER SHAFT T8 SELF-HOLDING THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 18649498 ↗
  • MalfunctionOct 19, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2023, THE TIPS OF THE INSTRUMENTS IN QUESTION WERE WORN DOWN. THERE WAS NO REPORTED PATIENT OR PROCEDURE INTERACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER SHAFT T8 SELF-RETAINING/QC. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17965743 ↗
  • MalfunctionOct 19, 2023Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2023, THE TIPS OF THE INSTRUMENTS IN QUESTION WERE WORN DOWN. THERE WAS NO REPORTED PATIENT OR PROCEDURE INTERACTION. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A STARDRIVE SCREWDRIVER SHAFT T8 SELF-RETAINING/QC. THIS IS REPORT 4 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17965742 ↗

Other Intervertebral fusion device with integrated fixation, cervical devices

All Intervertebral fusion device with integrated fixation, cervical devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.