Skip to content

XRL 03.807.310Recalls & reported problems

Synthes GmbH · XRL MEDIUM SHAFT · Class 2

All models of XRL (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034723129), received between Aug 24, 2021 and Jun 23, 2022 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (2 reports).

1 other recall cover devices cleared under the same FDA submission (K103320); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Compatibility Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 23, 2022Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING THE PREPARATION FOR IMPLANTATION OF A XRL CAGE, THE ASSEMBLY OF THE IMPLANT HOLDER AND IT WAS NOTICED THAT THE XRL RELEASE TOOL DID NOT HOLD TIGHT IN THE SHAFT AND THE TACTILE FEEDBACK OF THE RESTING POSITION AND THE END POSITION WAS VERY WEAK. AFTER CONNECTING TO THE IMPLANT, THE IMPLANT WAS EXPANDED FOR TEST OVER THE BACK TABLE (NOT IN SITU). IT WAS THEN NOTICED THAT THE RELEASE TOOL WAS SLIPPING BACK OUT OF THE SHAFT AND HAD TO BE HOLD IN PLACE BY HAND. AS THE IMPLANT HAD TO BE COLLAPSED, IT WAS NOT POSSIBLE (HANDLE ON OF, RELEASE TOOL IN PLACE) TO TURN THE HANDLE COUNTERCLOCKWISE AND COLLAPSE THE IMPLANT. THERE WAS A SURGICAL DELAY OF 10 MINUTES. A SECOND IMPLANT WAS UNWRAPPED AN IMPLANTED WITH CONSTANT PRESSURE ON THE RELEASE TOOL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) XRL MEDIUM SHAFT. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14789712 ↗
  • MalfunctionAug 24, 2021Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE INSTRUMENT WAS DIFFICULT TO WORK WITH AND BECAME STUCK. THE ISSUE WAS DISCOVERED PRIOR TO A PROCEDURE. THIS REPORT INVOLVES ONE (1) XRL MEDIUM SHAFT. THIS IS REPORT 3 OF 3 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12361670 ↗

FDA clearance and classification

Cleared through 510(k) K103320 on Oct 20, 2011 (decision: Substantially Equivalent). Submitted by Synthes Spine Co.Lp as “Synthes Xrl”. The FDA review took 342 days. FDA record ↗

The FDA regulates product code MQP (spinal vertebral body replacement device) under 21 CFR 888.3060 as a Class II device. 21 CFR 888.3060 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K103320. One submission often covers many models, so check the product and lot details in each record.

  • Z-1681-2014 · Apr 8, 2014 · Terminated
    Synthes XRL Medium Trial - Central Body, an instrument used as part of the Synthes XRL System, a vertebral body replacement device. For spinal surgery. The Synthes XRL device is

Common questions about the XRL 03.807.310

Has the XRL 03.807.310 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034723129), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the XRL 03.807.310?

The FDA's MAUDE database lists 2 adverse event reports for this device, received between Aug 24, 2021 and Jun 23, 2022. In those reports, the most common specific device problem is “Compatibility Problem” (2). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the XRL 03.807.310?

None of the reports are classified as deaths, 0 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the XRL 03.807.310 FDA approved?

It was cleared through 510(k) K103320 on Oct 20, 2011. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the XRL 03.807.310, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand XRL, model 03.807.310). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices

All SPINAL VERTEBRAL BODY REPLACEMENT DEVICE devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page