XRLRecalls & reported problems across all models
Synthes GmbH · 11 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 52
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls
- 0
We found 11 models of XRL by Synthes GmbH in the FDA's device registry (GUDID). Together they are named in 52 adverse event reports in the FDA's MAUDE database, received between May 29, 2018 and Sep 3, 2025 — classified as 4 injuries, 48 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (34 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Most reported XRL problems
Device problems
- Calibration Problem34
- Compatibility Problem10
- Adverse Event Without Identified Device or Use Problem3
- Material Integrity Problem3
- Device-Device Incompatibility1
- Failure to Calibrate1
Patient problems
- No Clinical Signs, Symptoms or Conditions30
- No Patient Involvement14
- No Consequences Or Impact To Patient4
- No Code Available2
- Nonunion/Delayed-union Bone Fracture2
XRL models
11 models, most reported first. Each model has its own page with its reports, recalls and device details.
- XRL 03.807.357ORTHOPEDIC MANUAL SURGICAL INSTRUMENT20 reports
- XRL 03.807.358ORTHOPEDIC MANUAL SURGICAL INSTRUMENT15 reports
- XRL 03.807.348SPINAL VERTEBRAL BODY REPLACEMENT DEVICE5 reports
- XRL 03.807.300SPINAL VERTEBRAL BODY REPLACEMENT DEVICE4 reports
- XRL 03.807.310SPINAL VERTEBRAL BODY REPLACEMENT DEVICE2 reports
- XRL 03.807.313SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 report
- XRL 03.807.315SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 report
- XRL 03.807.330SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 report
- XRL 03.807.382SPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 report
- XRL 08.807.234SSPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 report
- XRL 08.807.235SSPINAL VERTEBRAL BODY REPLACEMENT DEVICE1 report
Recent XRL adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT THE ITEM WAS FOUND TO BE OUT OF TOLERANCE WHEN DOING AN ON SITE TEST. EVENT DID NOT INVOLVE ANY PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22954205 ↗MalfunctionMay 16, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT AT A JNJ FACILITY ON AN UNKNOWN DATE, SEVERAL INSTRUMENTS FAILED A TORQUE LIMITING TEST. THE DEVICES ARE OUT OF CALIBRATION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22038856 ↗MalfunctionAug 22, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT DURING AN ON-SITE TEST ON (B)(6) 2024, A TORQUE LIMITING HANDLE WAS FOUND TO BE OUT OF TOLERANCE. THE DEVICE TESTED LOW. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A XRL MEDIUM TORQUE LIMITING HANDLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20025501 ↗MalfunctionJan 31, 2024Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN PORTUGAL AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6)2024, THE XRL RELEASE TOOL BROKE WHEN THE SURGEON WAS TRYING TO DETACH IT. THE TOP METAL PART THAT IS MADE FOR HAMMERING BROKE FROM THE REST OF THE INSTRUMENT. THE SURGERY WAS DELAYED BY FIVE MINUTES, AND WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCE. THIS REPORT IS FOR AN XRL RELEASE TOOL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18614820 ↗MalfunctionDec 14, 2023Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN ISRAEL AS FOLLOWS: IT WAS REPORTED THAT ON ( B)(6) 2023, DURING THE SURGERY THAT TOOK PLACE AT THE HOSPITAL. A FAULTY ITEM WAS DISCOVERED IN THE XRL SPREADER CAGE HOLDER SET. THE SURGEON CONNECTED TO THE CAGE, BUT THE CAGE HOLDER WAS UNABLE TO OPEN THE CAGE. THERE WAS NO CLAIM AGAINST THE CAGE. NO FRAGMENTS WERE GENERATED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THIS REPORT INVOLVES ONE (1) XRL RELEASE TOOL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18329096 ↗MalfunctionAug 2, 2023Material Integrity Problem; Material Integrity Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE UNITED KINGDOM AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE END HANDLE/KNOB DETACHED FROM RELEASE TOOL FOLLOWING USE IN THEATRE. THIS REPORT INVOLVES ONE (1) XRL RELEASE TOOL. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17443672 ↗MalfunctionOct 25, 2022Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING EVALUATION AT SERVICE AND REPAIR, IT WAS FOUND OUT THAT XRL LARGE TORQUE LIMITING HANDLE HAD FAILED IN CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE XRL LARGE TORQUE LIMITING HANDLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15666056 ↗MalfunctionOct 24, 2022Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING EVALUATION AT SERVICE AND REPAIR, THE REPAIR TECHNICIANS FOUND THAT THIS DEVICE FAILED CALIBRATION. ISSUE WAS FOUND DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE XRL LARGE TORQUE LIMITING HANDLE. THIS IS REPORT 1 OF 1 FOR (B)(4)..
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15659642 ↗MalfunctionAug 19, 2022Calibration Problem; Calibration Problem
IT WAS REPORTED THAT ON AN UNKNOWN DATE IN 2022, DURING EVALUATION AT SERVICE AND REPAIR, THE TORQUE LIMITING HANDLE IN QUESTION FAILED CALIBRATION. THERE WAS NO REPORTED PATIENT INVOLVEMENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT IS FOR A XRL MEDIUM TORQUE LIMITING HANDLE. THIS IS REPORT 1 OF 1 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15261677 ↗MalfunctionAug 18, 2022Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING THE PREPARATION FOR IMPLANTATION OF A XRL CAGE, THE ASSEMBLY OF THE IMPLANT HOLDER AND IT WAS NOTICED THAT THE XRL RELEASE TOOL DID NOT HOLD TIGHT IN THE SHAFT AND THE TACTILE FEEDBACK OF THE RESTING POSITION AND THE END POSITION WAS VERY WEAK. AFTER CONNECTING TO THE IMPLANT, THE IMPLANT WAS EXPANDED FOR TEST OVER THE BACK TABLE (NOT IN SITU). IT WAS THEN NOTICED THAT THE RELEASE TOOL WAS SLIPPING BACK OUT OF THE SHAFT AND HAD TO BE HOLD IN PLACE BY HAND. AS THE IMPLANT HAD TO BE COLLAPSED, IT WAS NOT POSSIBLE (HANDLE ON OF, RELEASE TOOL IN PLACE) TO TURN THE HANDLE COUNTERCLOCKWISE AND COLLAPSE THE IMPLANT. THERE WAS A SURGICAL DELAY OF 10 MINUTES. A SECOND IMPLANT WAS UNWRAPPED AN IMPLANTED WITH CONSTANT PRESSURE ON THE RELEASE TOOL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) XRL MEDIUM SPREADER TOP WITH 8MM DISTRACTION. THIS IS REPORT 4 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15248332 ↗MalfunctionAug 18, 2022Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING THE PREPARATION FOR IMPLANTATION OF A XRL CAGE, THE ASSEMBLY OF THE IMPLANT HOLDER AND IT WAS NOTICED THAT THE XRL RELEASE TOOL DID NOT HOLD TIGHT IN THE SHAFT AND THE TACTILE FEEDBACK OF THE RESTING POSITION AND THE END POSITION WAS VERY WEAK. AFTER CONNECTING TO THE IMPLANT, THE IMPLANT WAS EXPANDED FOR TEST OVER THE BACK TABLE (NOT IN SITU). IT WAS THEN NOTICED THAT THE RELEASE TOOL WAS SLIPPING BACK OUT OF THE SHAFT AND HAD TO BE HOLD IN PLACE BY HAND. AS THE IMPLANT HAD TO BE COLLAPSED, IT WAS NOT POSSIBLE (HANDLE ON OF, RELEASE TOOL IN PLACE) TO TURN THE HANDLE COUNTERCLOCKWISE AND COLLAPSE THE IMPLANT. THERE WAS A SURGICAL DELAY OF 10 MINUTES. A SECOND IMPLANT WAS UNWRAPPED AN IMPLANTED WITH CONSTANT PRESSURE ON THE RELEASE TOOL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) XRL SPREADER. THIS IS REPORT 3 OF 4 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15248305 ↗MalfunctionJun 23, 2022Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING THE PREPARATION FOR IMPLANTATION OF A XRL CAGE, THE ASSEMBLY OF THE IMPLANT HOLDER AND IT WAS NOTICED THAT THE XRL RELEASE TOOL DID NOT HOLD TIGHT IN THE SHAFT AND THE TACTILE FEEDBACK OF THE RESTING POSITION AND THE END POSITION WAS VERY WEAK. AFTER CONNECTING TO THE IMPLANT, THE IMPLANT WAS EXPANDED FOR TEST OVER THE BACK TABLE (NOT IN SITU). IT WAS THEN NOTICED THAT THE RELEASE TOOL WAS SLIPPING BACK OUT OF THE SHAFT AND HAD TO BE HOLD IN PLACE BY HAND. AS THE IMPLANT HAD TO BE COLLAPSED, IT WAS NOT POSSIBLE (HANDLE ON OF, RELEASE TOOL IN PLACE) TO TURN THE HANDLE COUNTERCLOCKWISE AND COLLAPSE THE IMPLANT. THERE WAS A SURGICAL DELAY OF 10 MINUTES. A SECOND IMPLANT WAS UNWRAPPED AN IMPLANTED WITH CONSTANT PRESSURE ON THE RELEASE TOOL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) XRL MEDIUM SHAFT. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14789712 ↗MalfunctionJun 23, 2022Compatibility Problem; Compatibility Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN SWITZERLAND AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2022, DURING THE PREPARATION FOR IMPLANTATION OF A XRL CAGE, THE ASSEMBLY OF THE IMPLANT HOLDER AND IT WAS NOTICED THAT THE XRL RELEASE TOOL DID NOT HOLD TIGHT IN THE SHAFT AND THE TACTILE FEEDBACK OF THE RESTING POSITION AND THE END POSITION WAS VERY WEAK. AFTER CONNECTING TO THE IMPLANT, THE IMPLANT WAS EXPANDED FOR TEST OVER THE BACK TABLE (NOT IN SITU). IT WAS THEN NOTICED THAT THE RELEASE TOOL WAS SLIPPING BACK OUT OF THE SHAFT AND HAD TO BE HOLD IN PLACE BY HAND. AS THE IMPLANT HAD TO BE COLLAPSED, IT WAS NOT POSSIBLE (HANDLE ON OF, RELEASE TOOL IN PLACE) TO TURN THE HANDLE COUNTERCLOCKWISE AND COLLAPSE THE IMPLANT. THERE WAS A SURGICAL DELAY OF 10 MINUTES. A SECOND IMPLANT WAS UNWRAPPED AN IMPLANTED WITH CONSTANT PRESSURE ON THE RELEASE TOOL. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WERE NO PATIENT CONSEQUENCES. THIS REPORT INVOLVES ONE (1) XRL RELEASE TOOL. THIS IS REPORT 1 OF 2 FOR (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14789594 ↗MalfunctionJun 23, 2022Calibration Problem; Calibration Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, DURING EVALUATION AT SERVICE AND REPAIR IT WAS FOUND OUT THAT XRL LARGE TORQUE LIMITING HANDLE HAD FAILED IN CALIBRATION. THERE WAS NO PATIENT INVOLVEMENT. THIS COMPLAINT INVOLVES ONE(1) DEVICE XRL LARGE TORQUE LIMITING HANDLE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14786468 ↗MalfunctionJun 6, 2022Calibration Problem; Calibration Problem
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN THE USA AS FOLLOWS: FAILED CALIBRATION. THIS COMPLAINT INVOLVES ONE (1) DEVICE XRL LARGE TORQUE LIMITING HANDLE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14604960 ↗
Other Synthes GmbH products
All Synthes GmbH devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page