Skip to content

NA 04.633.401Recalls & reported problems

Synthes GmbH · TI SNAP-ON OPEN PARALLEL CONNECTOR 5.5MM/6.0MM · Class 2 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10705034746753), received between May 25, 2017 and Oct 6, 2017 — classified as 2 injuries.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Device Dislodged or Dislocated1
  • Unintended Movement1

Patient problems

  • Failure of Implant2
  • No Code Available1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 6, 2017Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT REVISION SURGERY WAS COMPLETED ON (B)(6) 2017. THE PATIENT WAS IMPLANTED WITH SYNTHES CLICK X AND PANGEA SCREWS AS WELL AS DEPUY SCREWS MORE THAN 5 YEARS AGO. THE ENTIRE CONSTRUCT IS FROM T-11 TO L-5. THE PATIENT WAS THEN IMPLANTED WITH MATRIX CONNECTORS AND RODS TO JOIN TOGETHER THE SYNTHES AND DEPUY SCREWS ON (B)(6) 2017. THE PATIENT WAS INVOLVED IN A CAR ACCIDENT ON AN UNKNOWN DATE AND PRESENTED WITH PAIN. HE FOLLOWED UP WITH THE SURGEON IN THE BEGINNING OF (B)(6) AND IT WAS DISCOVERED THAT THE CONNECTORS DISCONNECTED FROM THE RODS ON THE L-2 TO L-3. THE PATIENT WAS REVISED WITH A SYNTHES TRANSITIONAL ROD AND THE LOCKING CAPS WERE REPLACED. THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO REPORT OF PATIENT HARM. THERE ARE 2 DEVICES IN THIS COMPLAINT CONCOMITANT DEVICES REPORTED: RODS (PART# 04.620.160, LOT# 1526511, QUANTITY 2), DEPUY SPINE SCREWS (PART # 1791-71-085, LOT# BDHT1SY, QUANTITY 1). THIS REPORT IS 1 OF 2 FOR (B)(4).

    Patient problems: Failure of Implant

    MAUDE report 6924007 ↗
  • InjuryMay 25, 2017Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNKNOWN PROCEDURE IN (B)(6) 2016 AT MULTIPLE LEVELS OF THE THORACIC AND LUMBAR SPINE AND WAS IMPLANTED WITH UNKNOWN SYNTHES DEVICES. DURING A ROUTINE FOLLOW-UP VISIT 6 MONTHS LATER, IT WAS IDENTIFIED VIA X-RAY THAT THE ROD TO ROD CONNECTOR HAD COME LOOSE AT THE L2-L3 LEVEL. ON (B)(6) 2017 THE PATIENT WAS RETURNED TO THE OPERATING ROOM WHERE THE ROD TO ROD CONNECTOR WAS EXPLANTED, AND A NEW ROD TO ROD CONNECTOR WAS IMPLANTED IN ITS PLACE. THERE WAS NO DELAY IN SURGERY REPORTED AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THE PATIENT STATUS IS REPORTED AS STABLE.CONCOMITANT DEVICES REPORTED: UNKNOWN SYNTHES 6.0MM / 5.5MM RODS (PART #: UNKNOWN, LOT #UNKNOWN, QTY. 2), UNKNOWN SYNTHES SCREWS (PART #: UNKNOWN, LOT #: UNKNOWN, QTY. UNKNOWN), UNKNOWN SYNTHES LOCKING CAPS (PART #: UNKNOWN, LOT #: UNKNOWN, QTY. UNKNOWN).THIS REPORT IS FOR ONE (1) ROD CONNECTOR.THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Failure of Implant; No Code Available

    MAUDE report 6590351 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page