Skip to content

NA 04.641.095Recalls & reported problems

Synthes GmbH · TI PROXIMAL EXTENSION SIZE 5/220MM RADIUS · Class 2 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10705034750392), received between Apr 1, 2019 and Apr 1, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Break” (1 report).

1 other recall cover devices cleared under the same FDA submission (K142587); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Break1

Patient problems

  • Discomfort1
  • No Code Available1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 1, 2019Break; Break

    04/23/2019: UPDATED EVENT DESCRIPTION: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT A REVISION SURGERY FOR A FAILED VERTICAL EXPANDABLE PROSTHETIC TITANIUM RIB (VEPTR) TITANIUM DISTAL EXTENSION, TITANIUM PROXIMAL EXTENSION AND TITANIUM DISTRACTION LOCK, THESE DEVICES HAVE BREAKAGE/FRACTURE. THE PATIENT PRESENTED WITH SOME DISCOMFORT. X-RAYS REVEALED SPLAYING OF THE RAILS OF THE PROXIMAL EXTENSION. THE PATIENT ALSO WAS ABOUT TO OUTGROW HIS EXTENSION ON THE CONTRA-LATERAL SIDE, SO SURGERY WAS SCHEDULE TO REPLACE BOTH SIDES. THERE WAS PATIENT CONSEQUENCE DUE TO MYELOMENIGOCELE. PROCEDURE OUTCOME IS UNKNOWN. THE PATIENT EXPOSED TO ADDITIONAL ANESTHESIA FOR ONE (1) HOUR. CONCOMITANT DEVICE REPORTED: TI DISTRACTION LOCK (PART # 497.125, LOT # H468144, QUANTITY # 1) UNK - RODS: SPINE (PART # UNKNOWN, LOT # UNKNOWN # QUANTITY # UNKNOWN). THIS COMPLAINT INVOLVES TWO (2) DEVICES.

    Patient problems: Discomfort; No Code Available

    MAUDE report 8470967 ↗

FDA clearance and classification

Cleared through 510(k) K142587 on Nov 18, 2014 (decision: Substantially Equivalent). Submitted by Depuy Synthes Spine, Inc. as “Vertical Expandable Prosthetic Titanium Rib”. The FDA review took 64 days. FDA record ↗

The FDA regulates product code MDI (prosthesis, rib replacement) under 21 CFR 888.3070 as a Class II device. 21 CFR 888.3070 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K142587. One submission often covers many models, so check the product and lot details in each record.

  • Z-2488-2015 · Jul 29, 2015 · Terminated
    VEPTR Vertical Expandable Prosthetic Titanium Rib, inferior cradles also known as caudal rib supports. The VEPTR/VEPTR II devices mechanically stabilize and/or correct thoracic de

Common questions about the NA 04.641.095

Has the NA 04.641.095 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034750392), based on the FDA recall file dated Nov 27, 2026. However, 1 recall involves devices cleared under the same FDA submission. It may or may not include this model, so check the product, model and lot details in that record.

What problems have been reported with the NA 04.641.095?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Apr 1, 2019. That report lists device problem “Break” (1); patient problem “Discomfort” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the NA 04.641.095?

None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the NA 04.641.095 FDA approved?

It was cleared through 510(k) K142587 on Nov 18, 2014. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the NA 04.641.095 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.

Who makes the NA 04.641.095, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand NA, model 04.641.095). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Prosthesis, rib replacement devices

All Prosthesis, rib replacement devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page