Skip to content

NA 387.286Recalls & reported problems

Synthes GmbH · SINGLE LOCKING DRILL GUIDE · Class 1

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10705034770994), received between Jan 16, 2017 and Jan 31, 2024 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Component Missing2
  • Material Integrity Problem2
  • Corroded1
  • Material Fragmentation1
  • Material Twisted/Bent1

Patient problems

  • No Patient Involvement3
  • No Clinical Signs, Symptoms or Conditions2
  • No Code Available1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 31, 2024Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM CANADA REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE TOOTH OF A DRILL GUIDE IN THE CERVICAL SPINE LOCKING PLATE (CSLP) SET HAD BROKEN OFF; DRILL GUIDE WAS NOT WORKING PROPERLY. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR A SINGLE LOCKING DRILL GUIDE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18619086 ↗
  • MalfunctionOct 1, 2021Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, WHILE INSPECTING THE SINGLE LOCKING DRILL GUIDE, THE TECHNICIAN NOTED THE TIP HAD BROKEN COMPONENT. THERE WAS NO PROCEDURE OR PATIENT INVOLVEMENT REPORTED. THIS REPORT IS FOR A SINGLE LOCKING DRILL GUIDE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12565215 ↗
  • MalfunctionJun 25, 2019Break; Material Twisted/Bent; Break; Material Twisted/Bent

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SELF-HOLDING DRILL SLEEVE WAS BENT AND UNABLE TO ALLOW FULL PASSAGE OF THE DRILL BIT. IT IS UNKNOWN IF WHEN THE ISSUE WAS DISCOVERED. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SINGLE LOCKING DRILL GUIDE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 8732742 ↗
  • MalfunctionAug 30, 2018Break; Component Missing; Break; Component Missing

    IT WAS REPORTED ON (B)(6) 2018 THAT A SINGLE LOCKING DRILL GUIDE IS MISSING A SET SCREW DURING A ROUTINE INSPECTION. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SINGLE LOCKING DRILL GUIDE. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 7832862 ↗
  • MalfunctionFeb 8, 2018Break; Break

    IT WAS REPORTED THAT WHILE INSPECTING THE DRILL GUIDE IN CENTRAL STERILE, IT WAS DISCOVERED THAT ONE OF THE FLANGES WAS BROKEN. PART WAS NOT USED IN SURGERY. THERE WAS NO PATIENT OR PROCEDURE INVOLVEMENT. THIS REPORT IS FOR ONE (1) SINGLE LOCKING DRILL GUIDE. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Code Available

    MAUDE report 7251251 ↗
  • MalfunctionJan 16, 2017Corroded; Material Fragmentation; Component Missing; Corroded; Material Fragmentation; Component Missing

    DEVICE REPORT FROM SYNTHES EUROPE REPORTED AN EVENT IN (B)(6) AS FOLLOWS: IT WAS INITIALLY REPORTED THAT RUST WAS DETECTED ON THE SURFACE AT THE TIP OF A DRILL SLEEVE DURING CLEANING PROCESS. THERE WAS NO REPORTED PATIENT OR PROCEDURAL INVOLVEMENT ASSOCIATED WITH THE REPORTED EVENT. UPON INITIAL INSPECTION OF THE RETURNED DEVICE ON DECEMBER 23, 2016 IT WAS OBSERVED THAT A SMALL PORTION OF THE TONGS WAS BROKEN OFF AND MISSING. IT IS UNKNOWN WHEN THE REPORTED ISSUES OCCURRED. THIS EVENT WAS REASSESSED FOR REPORTABILITY BASED ON THIS ADDITIONAL FINDING. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 6253359 ↗

Other Guide, surgical, instrument devices

All Guide, surgical, instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.