Click'X 498.989Recalls & reported problems
Synthes (U.S.A.) LP · 6.2MM TI CLICK'X® PEDICLE SCR PREASSEMBLED 45MM THRD LENGTH · Class 2 · Implant
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10705034792385), received between Aug 9, 2018 and May 7, 2019 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Unintended Movement” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Unintended Movement2
- Detachment of Device or Device Component1
- Material Twisted/Bent1
Patient problems
- No Code Available2
- Pain2
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryMay 7, 2019Unintended Movement; Unintended Movement
InjuryMay 7, 2019Unintended Movement; Unintended Movement
ON 06/18/2019: UPDATED EVENT DESCRIPTION: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT AN ELECTIVE REVISION SURGERY SYSTEM INSTABILITY DUE TO THE CLICK COLUMN SYSTEM HAD LOOSENED, SPECIFICALLY, THE 3D HEADS, LOCKING SCREWS AND FOUR (4) PEDICLE SCREWS. THE PATIENT PRESENTED PAIN BEFORE DETECTING THE MALFUNCTION. THE CLICK COLUMN SYSTEM MALFUNCTION WAS DETECTED (B)(6) 2019. IT WAS MENTIONED THAT THE SCREWS THAT WERE USED IN THE SURGERY WERE PRE-ASSEMBLED FROM CLICK COLUMN SYSTEM THAT IS TO SAY THAT IN THIS SURGERY THEY DID NOT USE THE 3D HEADS. THE DEVICES WERE ORIGINALLY IMPLANTED ON (B)(6) 2019. THE RECOVERED IMPLANTS WERE, TWO (2) CLICKX PRE-ASSEMBLED SCREWS 6.2 X 45, TWO (2) CLICKX PRE-ASSEMBLED SCREWS 6.2 X 50, TWO (2) SOFT CURVED RODS AND FOUR (4) CLICKX LOCKING CAPS. THE SURGEON WILL MONITOR THE EFFECTIVENESS OF THE CLICK COLUMN SYSTEM THROUGH POSTOPERATIVE RADIOGRAPHIC CONTROLS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. CURRENTLY, THE PATIENT WAS IN GOOD CONDITION. THIS COMPLAINT INVOLVES TEN (10) DEVICES.
Patient problems: No Code Available; Pain
MAUDE report 8587444 ↗MalfunctionAug 9, 2018Detachment of Device or Device Component; Material Twisted/Bent; Detachment of Device or Device Component; Material Twisted/Bent
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2018, THE PATIENT UNDERWENT A LUMBAR ARTHRODESIS PROCEDURE. THE CLICK X PEDICLE SCREW DUSTED THE THREAD IN THE FINAL TIGHTENING. DURING THE PROCEDURE THERE WAS LOSS OF THE THREAD OF 2 PEDICLE SCREWS, MAKING THE SURGICAL PROCEDURE DIFFICULT. IT WAS NECESSARY TO USE PRESSURE TO REMOVE THE SCREW. IT WAS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO PATIENT OUTCOME REPORTED. DURING MANUFACTURER¿S INVESTIGATION OF THE RETURNED DEVICE IT WAS IDENTIFIED THAT THE LOCKING HEAD PORTION OF THE IMPLANT FELL APART IN ADDITION THERE IS DAMAGE TO THE TOP OF THE SCREW PORTION (MOSTLY LIKELY DONE BY PLIERS TO REMOVE THE SCREW, AS MATERIAL IS BENT AND ROLLED OVER) AND THE TOP PROXIMAL THREADS SHOWINGS SIGNS OF ROLLING. THIS DEVICE CONDITION WAS RE-EVALUATED AND DETERMINED TO BE REPORTABLE ON (B)(6) 2018. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7769953 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.