Skip to content

SYNFIX Evolution 03.835.004Recalls & reported problems

Synthes GmbH · SYNFIX® EVOLUTION AIMING DEVICE HOLDER · Class 1

All models of SYNFIX Evolution (21)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Synthes SYNFIX Evolution aiming device holder is a surgical instrument used to hold and position implants during bone fixation procedures. The 46 reports mostly describe the instrument jamming, seizing, bending or breaking, and parts not attaching properly to other equipment, so the shaft sometimes would not retract and could not be separated. Three injuries were reported and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
46
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 10705034814100), received between Oct 4, 2018 and Jul 21, 2025 — classified as 3 injuries, 43 malfunctions.

The most frequently reported specific device problem is “Compatibility Problem” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 2 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 1 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Compatibility Problem11
  • Mechanical Problem11
  • Mechanical Jam8
  • Break6
  • Device-Device Incompatibility6
  • Material Integrity Problem6
  • Detachment of Device or Device Component1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • No Patient Involvement13
  • No Consequences Or Impact To Patient8
  • No Code Available3
  • Insufficient Information2
  • Injury1
  • No Information1

Recent adverse event reports

The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 21, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THE INSERTION INSTRUMENT AND SLEEVE DID NOT FUNCTION PROPERLY. THE OUTER SLEEVE DOES NOT RETRACT AND IS VERY DIFFICULT TO SEPARATE. THERE WAS NO CONSEQUENCES FOR THE PATIENT. THERE WAS NO DELAY IN THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22554377 ↗
  • MalfunctionJun 4, 2025Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT UPON CAGE ATTACHMENT, THE COUPLING DEVICE DIDN'T ATTACH TO SAID ATTACHMENT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO GENERATION OF FRAGMENTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22141827 ↗
  • MalfunctionMay 28, 2025Compatibility Problem; Compatibility Problem

    IT WAS REPORTED WHEN CHECKING IN INSTRUMENT AND IMPLANT TRAYS, INSTRUMENT 03.835.004 DID NOT FUNCTION PROPERLY. THE QUICK COUPLE MECHANISM TO ATTACH AND DETACH FROM AIMING GUIDE WAS SEIZED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095998 ↗
  • MalfunctionMay 28, 2025Compatibility Problem; Compatibility Problem

    IT WAS REPORTED WHEN CHECKING IN INSTRUMENT AND IMPLANT TRAYS, INSTRUMENT 03.835.004 DID NOT FUNCTION PROPERLY. THE QUICK COUPLE MECHANISM TO ATTACH AND DETACH FROM AIMING GUIDE WAS SEIZED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22095218 ↗
  • MalfunctionMar 7, 2025Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT THE DEVICE DOES NOT FUNCTION OR HOLD. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21553323 ↗
  • MalfunctionMar 7, 2025Compatibility Problem; Compatibility Problem

    IT WAS REPORTED THAT THE DEVICE DOES NOT FUNCTION OR HOLD. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH NO SURGICAL DELAY. THERE WAS NO PATIENT CONSEQUENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21553320 ↗
  • MalfunctionOct 31, 2024Material Integrity Problem; No Apparent Adverse Event; Material Integrity Problem; No Apparent Adverse Event

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SYNFIX EVOLUTION IMPLANT HOLDER SHAFT BENT AND WAS NOT USABLE. THERE WAS NO PATIENT CONSEQUENCE. NO OTHER INFORMATION WAS PROVIDED AT THE TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20577679 ↗
  • MalfunctionOct 11, 2024Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT ON OCTOBER 1 2024, THE AIMING DEVICE HOLDER JAMMED ON INSPECTION. THE SHAFT REQUIRED FORCE TO RETRACT AND THEN BECAME STUCK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20430826 ↗
  • MalfunctionJun 10, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON UNKNOWN DATE THE SYNFIX EVOLUTION SET AND THE INSERTER, AS WELL AS THE SYNFIX EVOLUTION AIMING DEVICE HOLDER IN THE SET BROKE. THIS REPORT IS FOR ONE SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19498351 ↗
  • MalfunctionJun 8, 2024Material Integrity Problem; Material Integrity Problem

    IT WAS REPORTED THAT THE INSERTER IN THE SET IS BROKEN. NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: Insufficient Information

    MAUDE report 19494256 ↗
  • MalfunctionDec 21, 2023Compatibility Problem; Material Integrity Problem; Compatibility Problem; Material Integrity Problem

    IT WAS REPORTED THAT ON NOVEMBER 20, 2023, AFTER AN A&P LUMBAR FUSION PROCEDURE, THE TAB RESTRICTING SLEEVE WAS BENT FROM THE ENGAGING AIMING DEVICE DUE TO GETTING STUCK WHEN THE OUTER SLEEVE WAS RETRACTED. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18380350 ↗
  • MalfunctionMay 2, 2023Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6)2023, THE SCREW IN THE SYNFIX EVO AIMING DEVICE 10.5 PLUS12 BECAME STICKY AND REQUIRED MORE FORCE THAN USUAL. THE SYNFIX EVOLUTION AIMING DEVICE HOLDER CAME IN THE SET PULLED APART, AND IT COULD NOT BE PUT BACK TOGETHER AS IT WAS VERY STIFF. THE SYNFIX EVOLUTION SCREWDRIVER HAD WEAR AND TEAR ON THE DRIVER TIPS, AND FRAYED EDGES. THERE WAS NO IMPACT TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE. THIS REPORT INVOLVES ONE SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 3 OF 4 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16857022 ↗
  • MalfunctionFeb 21, 2023Mechanical Problem; Mechanical Problem

    IT WAS REPORTED DURING AN INSPECTION AT THE LOANER SET DEPARTMENT AT LOC UMKIRCH ON (B)(6) 2022, IT WAS NOTICED THAT THE SYNFIX EVOLUTION AIMING DEVICE HOLDER WAS JAMMED. THE ITEMS HAS ONLY BEEN USED 1 TIMES. NO SURGERY IMPACT. NO PATIENT IMPACT. THIS IS REPORT 1 OF 1 FOR (B)(4). THIS REPORT IS FOR THE SYNFIX AIMING DEVICE HOLDER

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16409229 ↗
  • MalfunctionDec 28, 2022Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN GERMANY AS FOLLOWS: IT WAS REPORTED THAT RATCH HANDLE W/AO QUICK COUPL SM: DURING AN INSPECTION AT THE LOANER SET DEPARTMENT AT LOC UMKIRCH IT WAS NOTICED THAT THE RUBBER OF THE HANDPIECE COMES OFF. NO SURGERY IMPACT. NO PATIENT IMPACT. SYNFIX® EVOLUTION AIMING DEVICE HOLDER: DURING AN INSPECTION AT THE LOANER SET DEPARTMENT AT LOC UMKIRCH IT WAS NOTICED THAT INNER SHAFT RUNS TOO 'HEAVY. NO SURGERY IMPACT. NO PATIENT IMPACT. THIS REPORT IS FOR ONE (1)SYNFIX EVOLUTION AIMING DEVICE HOLDER THIS IS REPORT 2 OF 2 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16062649 ↗
  • MalfunctionJun 29, 2022Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN AUSTRALIA AS FOLLOWS: IT WAS REPORTED ON (B)(6) 2022, THAT THE SYNFIX EVOLUTION AIMING DEVICE HOLDER WAS NOT CONNECTING TO THE SYNFIX EVOLUTION AIMING DEVICE. ANOTHER AIMING DEVICE HOLDER WAS PUT TOGETHER SO WE COULD USE THAT. THAT HOLDER WORKED. THERE WAS SURGICAL DELAY FOR TWO MINUTES AND MORE. PROCEDURE WAS COMPLETED SUCCESSFULLY AND NO FRAGMENTS WERE GENERATED. THERE WAS NO PATIENT OUTCOME/ CONSEQUENCES. CONCOMITANT DEVICE REPORTED: UNK - GUIDES/SLEEVES/AIMING: SPINE (PART# UNKNOWN; LOT# UNKNOWN; QUANTITY: 1). THIS REPORT IS FOR ONE (1) SYNFIX EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14858568 ↗
  • MalfunctionApr 22, 2022Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED INTRA - OPERATIVELY WHILST TRYING TO LOAD THE SYNFIX EVOLUTION CAGE ONTO THE INSERTER FOR INSERTION, THE INSERTER WOULDN'T RETAIN THE IMPLANT. UPON CLOSER INSPECTION AND COMPARISON WITH A SECOND INSERTER ON THE SET, IT WAS OBVIOUS THAT TWO LITTLE NIBS WERE MISSING FROM THE DISTAL END OF THE INSERTER. THIS COMPLAINT INVOLVES ONE (1) DEVICE SYNFIX EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14186454 ↗
  • MalfunctionDec 28, 2021Compatibility Problem; Compatibility Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, DURING LOAN KIT INSPECTION THE AIMING DEVICE DID NOT MOVE FREELY, BECAME CAUGHT ON POINTS DURING ACTUATION, AND WAS DIFFICULT TO DISASSEMBLE. THIS REPORT INVOLVES (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13098474 ↗
  • MalfunctionJun 9, 2021Mechanical Problem; Mechanical Problem

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE THAT AFTER ASSEMBLING THE INSTRUMENT AS INTENDED, IT WAS IMPOSSIBLE TO TAKE APART OR USE AND WAS COMPLETELY STUCK. THIS WAS DISCOVERED DURING CLEANING AND STERILIZATION. THIS REPORT IS FOR (1) SYNFIX EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11967541 ↗
  • MalfunctionMar 30, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SYNFIX EVOLUTION AIMING DEVICE HOLDER SET STICKS, WILL NOT ATTACH EASILY TO THE AIM DEVICES. IT WAS TESTED PRIOR TO A CASE AND IT WAS NOT WORKING CORRECTLY. THERE WAS NO PATIENT INVOLVEMENT. CONCOMITANT DEVICE REPORTED. UNKNOWN SYNFIX EVOLUTION AIMING DEVICE (PAR T# UNKNOWN, LOT # UNKNOWN, QTY UNKNOWN). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS REPORT IS 1 OF 1 (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11592307 ↗
  • MalfunctionFeb 26, 2021Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2021, THE PATIENT UNDERWENT FOR AN UNKNOWN PROCEDURE. IT WAS DISCOVERED THAT THE AIMING DEVICE IS DEFECTIVE. IT WAS UNKNOWN WHEN AND HOW IT¿S OCCURRED. THE PROCEDURE WAS COMPLETED WITH THE SECOND SET. THERE WAS NO PATIENT IMPACT REPORTED. THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS REPORT IS 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11388846 ↗
  • MalfunctionDec 1, 2020Material Integrity Problem; Material Integrity Problem

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE DURING AN ANTERIOR LUMBAR INTERBODY FUSION (ALIF), ANTERIOR SPINE FIXATION, THAT THE INNER SHAFT OF THE INSERTER BENT. OPERATING THE TOOL IS DIFFICULT, IMPRECISE AND IT IS PRONE TO COMING APART DUE TO THE FORCES YOU HAVE TO USE ON IT. THERE WAS NO PATIENT INVOLVEMENT. THIS REPORT IS FOR ONE (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 FOR (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10927635 ↗
  • MalfunctionSep 22, 2020Mechanical Jam; Mechanical Jam

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THAT THE LOAN SET JAMS. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 2 OF 2 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10567033 ↗
  • MalfunctionSep 22, 2020Mechanical Jam; Mechanical Jam

    DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON AN UNKNOWN DATE, THAT THE LOAN SET JAMS. THERE WAS NO PATIENT CONSEQUENCE. THIS REPORT INVOLVES ONE (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10566996 ↗
  • MalfunctionSep 18, 2020Device-Device Incompatibility; Device-Device Incompatibility

    DEVICE REPORT FROM (B)(6) REPORTS AN EVENT AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2020, THE SYNFIX EVOLUTION AIMING DEVICE HOLDER WAS FAULTY NOT ABLE TO RETRACT AND HOLD AN UNKNOWN AIMING DEVICE. ANOTHER DEVICE WAS USED. THERE WAS A FIFTEEN (15) MINUTES SURGICAL DELAY. PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WAS NO PATIENT CONSEQUENCE. CONCOMITANT DEVICE REPORTED: UNKNOWN AIMING DEVICE (PART# UNKNOWN, LOT# UNKNOWN, QUANTITY 1). THIS COMPLAINT INVOLVES ONE (1) DEVICE. THIS REPORT IS FOR (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 1 (B)(4).

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10554604 ↗
  • MalfunctionSep 16, 2020Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT ON (B)(6) 2020, IN STERILE PROCESSING DEPARTMENT, WHILE ASSEMBLING THE SYNFIX EVOLUTION INSTRUMENT SET AFTER A CASE, AN AIMING DEVICE HOLDER WAS JAMMING. THERE IS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 1 OF 2 FOR (B)(4).

    Patient problems: No Patient Involvement

    MAUDE report 10541814 ↗

FDA clearance and classification

The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the SYNFIX Evolution 03.835.004

Has the SYNFIX Evolution 03.835.004 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705034814100), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the SYNFIX Evolution 03.835.004?

The FDA's MAUDE database lists 46 adverse event reports for this device, received between Oct 4, 2018 and Jul 21, 2025. In those reports, the most common specific device problems are “Compatibility Problem” (11), “Mechanical Problem” (11) and “Mechanical Jam” (8); the most common patient problem is “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SYNFIX Evolution 03.835.004?

None of the reports are classified as deaths, 3 as injuries and 43 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SYNFIX Evolution 03.835.004 FDA approved?

Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.

Who makes the SYNFIX Evolution 03.835.004, and is it still sold?

The FDA's device database lists Synthes GmbH as the company responsible for it (brand SYNFIX Evolution, model 03.835.004). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPACTOR devices

All IMPACTOR devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page