SYNFIX Evolution 03.835.005Recalls & reported problems
Synthes GmbH · COUPLING FOR SYNFIX® EVOLUTION AIMING DEVICE HOLDER · Class 1
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034814124), received between May 3, 2019 and Sep 16, 2020 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility4
- Mechanical Jam1
Patient problems
- No Code Available3
- No Patient Involvement2
- Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2020Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2020, IN STERILE PROCESSING DEPARTMENT, WHILE ASSEMBLING THE SYNFIX EVOLUTION INSTRUMENT SET AFTER A CASE, AN AIMING DEVICE HOLDER WAS JAMMING. THERE IS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) COUPLING FOR SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10541866 ↗MalfunctionMar 17, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SYNFIX EVOLUTION AIMING DEVICE HOLDER WAS JAMMED WITH THE COUPLING FOR AIMING DEVICE HOLDER SYNFIX EVOLUTION. THE ISSUE WAS DISCOVERED AT THE STERILE PROCESSING DEPARTMENT. THERE WERE NO PATIENT AND SURGICAL INVOLVEMENT. CONCOMITANT DEVICE REPORTED: SYNFIX® EVOLUTION AIMING DEVICE HOLDER (PART# 03.835.004, LOT# 4L26073, QUANTITY 1). THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9842919 ↗InjuryDec 23, 2019Device-Device Incompatibility; Device-Device Incompatibility
InjuryDec 23, 2019Device-Device Incompatibility; Device-Device Incompatibility
InjuryMay 3, 2019Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT AN ANTERIOR SPINAL FUSION PROCEDURE USING WITH SYNFIX EVOLUTION SYSTEM. DURING THE PROCEDURE, AN UNKNOWN IMPLANT WAS INSERTED WITH A CORRECTLY ATTACHED SYNFIX EVOLUTION AIMING DEVICE HOLDER AND SYNFIX EVOLUTION AIMING DEVICE, BUT THE COUPLING FOR AIMING DEVICE HOLDER REMAINED INSIDE. WHEN THE COUPLING WAS REMOVED, THE COUPLING SCREWDRIVER WAS NOT REMOVED BY PULLING COUNTER COUNTERCLOCKWISE WHEN IT WAS STILL ENGAGED WITH THE HOLDER. IT WAS ONLY RELEASED WHEN THE SURGEON PULLED THE COUPLING SCREWDRIVER. THE SURGEON DID NOT RECOGNIZE DUE TO THE COUNTERCLOCKWISE TURNING, IT LOOSENED THE COUPLING SCREW AND AIMING DEVICE WHERE THE IMPLANT WAS NO LONGER FULLY FIXED. AS A RESULT, DURING SUBSEQUENT PILOT HOLE CREATION AND SCREW INSERTION, THE TRAJECTORY WAS NOT 100% CORRECT DUE TO THE LOOSENED AIMING DEVICE. ONCE THE SURGEON GOT AWARE OF THE ROOT CAUSE HE REMOVED ALL SCREWS, TIGHTENED THE AIMING DEVICE ON THE IMPLANT AGAIN AND RE-INSERTED ALL SCREWS WITHOUT ANY FURTHER PROBLEMS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNSPECIFIED NUMBER OF MINUTES OF SURGICAL
Patient problems: Injury; No Code Available
MAUDE report 8579353 ↗
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,295 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,539 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,498 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,456 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall875 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR716 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR704 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR337 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.