SYNFIX Evolution 03.835.005Recalls & reported problems
Synthes GmbH · COUPLING FOR SYNFIX® EVOLUTION AIMING DEVICE HOLDER · Class 1
All models of SYNFIX Evolution (21)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10705034814124), received between May 3, 2019 and Sep 16, 2020 — classified as 3 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device-Device Incompatibility” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device-Device Incompatibility4
- Mechanical Jam1
Patient problems
- No Code Available3
- No Patient Involvement2
- Injury1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2020Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT ON (B)(6) 2020, IN STERILE PROCESSING DEPARTMENT, WHILE ASSEMBLING THE SYNFIX EVOLUTION INSTRUMENT SET AFTER A CASE, AN AIMING DEVICE HOLDER WAS JAMMING. THERE IS NO PATIENT INVOLVEMENT. THIS REPORT INVOLVES ONE (1) COUPLING FOR SYNFIX® EVOLUTION AIMING DEVICE HOLDER. THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 10541866 ↗MalfunctionMar 17, 2020Device-Device Incompatibility; Device-Device Incompatibility
IT WAS REPORTED THAT ON AN UNKNOWN DATE, THE SYNFIX EVOLUTION AIMING DEVICE HOLDER WAS JAMMED WITH THE COUPLING FOR AIMING DEVICE HOLDER SYNFIX EVOLUTION. THE ISSUE WAS DISCOVERED AT THE STERILE PROCESSING DEPARTMENT. THERE WERE NO PATIENT AND SURGICAL INVOLVEMENT. CONCOMITANT DEVICE REPORTED: SYNFIX® EVOLUTION AIMING DEVICE HOLDER (PART# 03.835.004, LOT# 4L26073, QUANTITY 1). THIS IS REPORT 2 OF 2 FOR (B)(4).
Patient problems: No Patient Involvement
MAUDE report 9842919 ↗InjuryDec 23, 2019Device-Device Incompatibility; Device-Device Incompatibility
InjuryDec 23, 2019Device-Device Incompatibility; Device-Device Incompatibility
InjuryMay 3, 2019Device-Device Incompatibility; Device-Device Incompatibility
DEVICE REPORT FROM SYNTHES REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THAT ON (B)(6) 2019, THE PATIENT UNDERWENT AN ANTERIOR SPINAL FUSION PROCEDURE USING WITH SYNFIX EVOLUTION SYSTEM. DURING THE PROCEDURE, AN UNKNOWN IMPLANT WAS INSERTED WITH A CORRECTLY ATTACHED SYNFIX EVOLUTION AIMING DEVICE HOLDER AND SYNFIX EVOLUTION AIMING DEVICE, BUT THE COUPLING FOR AIMING DEVICE HOLDER REMAINED INSIDE. WHEN THE COUPLING WAS REMOVED, THE COUPLING SCREWDRIVER WAS NOT REMOVED BY PULLING COUNTER COUNTERCLOCKWISE WHEN IT WAS STILL ENGAGED WITH THE HOLDER. IT WAS ONLY RELEASED WHEN THE SURGEON PULLED THE COUPLING SCREWDRIVER. THE SURGEON DID NOT RECOGNIZE DUE TO THE COUNTERCLOCKWISE TURNING, IT LOOSENED THE COUPLING SCREW AND AIMING DEVICE WHERE THE IMPLANT WAS NO LONGER FULLY FIXED. AS A RESULT, DURING SUBSEQUENT PILOT HOLE CREATION AND SCREW INSERTION, THE TRAJECTORY WAS NOT 100% CORRECT DUE TO THE LOOSENED AIMING DEVICE. ONCE THE SURGEON GOT AWARE OF THE ROOT CAUSE HE REMOVED ALL SCREWS, TIGHTENED THE AIMING DEVICE ON THE IMPLANT AGAIN AND RE-INSERTED ALL SCREWS WITHOUT ANY FURTHER PROBLEMS. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN UNSPECIFIED NUMBER OF MINUTES OF SURGICAL
Patient problems: Injury; No Code Available
MAUDE report 8579353 ↗
FDA clearance and classification
The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗
Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the SYNFIX Evolution 03.835.005
Has the SYNFIX Evolution 03.835.005 been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705034814124), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the SYNFIX Evolution 03.835.005?
The FDA's MAUDE database lists 5 adverse event reports for this device, received between May 3, 2019 and Sep 16, 2020. In those reports, the most common specific device problems are “Device-Device Incompatibility” (4) and “Mechanical Jam” (1); the most common patient problem is “Injury” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SYNFIX Evolution 03.835.005?
None of the reports are classified as deaths, 3 as injuries and 2 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SYNFIX Evolution 03.835.005 FDA approved?
Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Synthes GmbH.
Who makes the SYNFIX Evolution 03.835.005, and is it still sold?
The FDA's device database lists Synthes GmbH as the company responsible for it (brand SYNFIX Evolution, model 03.835.005). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,276 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,528 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,491 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,454 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall873 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR715 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR703 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR334 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page