ECHELON ECH45SRecalls & reported problems

Ethicon Endo-Surgery, LLC · ECHELON 3000 45 STANDARD · Class 2

All models of ECHELON (10)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Quick summary

This Ethicon Echelon surgical stapler appears in 265 adverse event reports, with no deaths and 19 injuries. The reports mostly describe the stapler failing to fire, being hard to open or close, failing to form staples, and articulating or positioning unexpectedly, with some reports describing a stapler that failed to fire while inside a patient. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
261
Deaths reported
0
Injuries reported
18
Recalls naming it
0

Summary

The FDA's MAUDE database contains 261 adverse event reports that list this exact device (DI 10705036035008), received between Feb 23, 2023 and Sep 24, 2026 — classified as 18 injuries, 243 malfunctions.

The most frequently reported specific device problem is “Failure to Fire” (100 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 8 reportsJan 2024: 7 reportsFeb 2024: 2 reportsMar 2024: 8 reportsApr 2024: 3 reportsMay 2024: 3 reportsJun 2024: 2 reportsJul 2024: 6 reportsAug 2024: 2 reportsSep 2024: 18 reportsOct 2024: 4 reportsNov 2024: 5 reportsDec 2024: 7 reportsJan 2025: 2 reportsFeb 2025: 6 reportsMar 2025: 5 reportsApr 2025: 3 reportsMay 2025: 9 reportsJun 2025: 7 reportsJul 2025: 9 reportsAug 2025: 4 reportsSep 2025: 5 reportsOct 2025: 7 reportsNov 2025: 6 reportsDec 2025: 8 reportsJan 2026: 16 reportsFeb 2026: 11 reportsMar 2026: 18 reportsApr 2026: 7 reportsMay 2026: 7 reportsJun 2026: 11 reportsJul 2026: 3 reportsAug 2026: 9 reportsSep 2026: 2 reportsOct 2026: 0 reportsNov 2026: 0 reports20242025202618
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Fire100
  • Difficult to Open or Close88
  • Positioning Failure34
  • Failure to Form Staple28
  • Material Integrity Problem18
  • Ejection Problem16
  • Mechanical Jam16
  • Adverse Event Without Identified Device or Use Problem14

Patient problems

  • No Clinical Signs, Symptoms or Conditions240
  • Hemorrhage/Blood Loss/Bleeding10
  • Unspecified Tissue Injury7
  • Insufficient Information6
  • Failure to Anastomose5
  • Pneumothorax5
  • Pain4
  • Vomiting2

Recent adverse event reports

The 25 most recent of 261 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884368 ↗
  • MalfunctionSep 17, 2026Failure to Cut

    It was reported that during a Whipple procedure, the surgeon used the stapler to transect pancreatic tissue. On the first firing, the device formed staples on both sides of the tissue as intended; however, the device failed to cut the tissue. Suspecting that tissue thickness may have contributed to the issue, the user replaced the initial blue reload with a thicker green reload and attempted to fire the device again. The device again stapled but failed to cut the tissue. After three unsuccessful firing attempts with different reloads, a new stapler was opened and used to continue the procedure. No patient consequences were reported. Additional information requested.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803544 ↗
  • MalfunctionAug 26, 2026Unintended Ejection

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50459722 ↗
  • MalfunctionAug 20, 2026Failure to Form Staple; Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50347388 ↗
  • MalfunctionAug 19, 2026Positioning Failure; Difficult to Open or Close; Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50324420 ↗
  • MalfunctionAug 14, 2026Failure to Form Staple

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50260388 ↗
  • MalfunctionAug 14, 2026Failure to Form Staple

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50260163 ↗
  • MalfunctionAug 14, 2026Positioning Failure; Failure to Fire; Difficult to Open or Close; No Apparent Adverse Event

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury

    MAUDE report 50254714 ↗
  • MalfunctionAug 14, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50252908 ↗
  • MalfunctionAug 6, 2026Failure to Fire; Difficult to Open or Close

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50161564 ↗
  • MalfunctionAug 4, 2026Failure to Fire

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50131258 ↗
  • MalfunctionJul 27, 2026Failure to Fire; Difficult to Open or Close; Ejection Problem

    It was reported that during an Appendectomy, the first fire was white reload, second fire was blue reload and the third reload locked out open and couldn't reload. The tech said she did clean it properly. There was no patient consequence nor surgical delay reported. No additional information provided.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50019365 ↗
  • MalfunctionJul 9, 2026Failure to Form Staple; Difficult to Open or Close

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE SURGEON WAS USING THE STAPLER ON A LUNG. STAPLER FIRED AND LOCKED. THEY WERE NOT ABLE TO OPEN. HAD TO MANUAL OVERRIDE TO OPEN. MULTIPLE ROWS OF STABLES STANDING STRAIGHT UP. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED..

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25762882 ↗
  • MalfunctionJul 1, 2026Failure to Fire

    IT WAS REPORTED THAT DURING THE LAPAROSCOPIC APPENDECTOMY PROCEDURE, THE DEVICE MISFIRED AND A NEW DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25678351 ↗
  • MalfunctionJun 25, 2026Positioning Failure; Activation Problem

    IT WAS REPORTED THAT DURING A PEDIATRIC LAP APPENDECTOMY, THE APPENDIX WAS SUCCESSFULLY STAPLED AND FREED BUT THE STAPLER WOULD NOT STRAIGHTEN OUT AND PUSHING THE HOME BUTTON OR ARROW BUTTON. THEY USED THE MANUAL OVERRIDE, BUT BLACK LEVER WAS VERY STUCK. NO PATIENT CONSEQUENCES REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25622248 ↗
  • MalfunctionJun 24, 2026Positioning Failure

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE DEVICE WOULD NOT ARTICULATE BEFORE FIRING. CHANGED THE NEW ONE TO COMPLETE SURGERY. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION CAN BE PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25612749 ↗
  • MalfunctionJun 23, 2026Failure to Form Staple; Failure to Fire; Difficult to Open or Close; No Apparent Adverse Event; Noise, Audible

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE OF THE LIVER, SURGEON WENT TO USE ECHELON 3000 STAPLER AND THE BATTERY DIED - DID NOT HAVE ENOUGH POWER TO FIRE THROUGH ENTIRE ORGAN (LIVER) AND AS A RESULT LEFT A HOLE. REP WAS ON SITE IN THE DAYS FOLLOWING AND SPOKE WITH SURGEON - THE SURGEON SAID THE BATTERY DID NOT SOUND RIGHT - NOT A SLOWING DOWN IN THICK TISSUE BUT A VERY SLUGGISH SOUND AND THEN DIED. SURGEON WAS FIRING ON A VESSEL IN THE LIVER AND WAS VERY ANXIOUS ABOUT THE OUTCOME - USED THE MANUAL OVERRIDE TO REVERSE IT. RESIDENT TOLD ME SEPARATELY BUT DID NOT FORM PROPERLY. SURGEON OPENED A NEW STAPLER AND COMPLETED THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury

    MAUDE report 25595609 ↗
  • MalfunctionJun 19, 2026Positioning Failure

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE ARTICULATION WAS NOT WORKING PROPERLY ON THE DEVICE. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25566181 ↗
  • MalfunctionJun 8, 2026Failure to Fire; No Apparent Adverse Event

    IT WAS REPORTED THAT THE DURING A ROBOTIC ASSISTED NEPHRECTOMY, DEVICE WAS FIRED ACROSS THE KIDNEY AND SURGEON WAS NOT HAPPY WITH PERFUSION COMING FROM DISTAL END OF STAPLE LINE. SURGEON OPENED NEW ECH45S AND CASE CONTINUED WITH NO ADVERSE EVENTS OR ISSUES. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions

    MAUDE report 25431032 ↗
  • MalfunctionJun 8, 2026Failure to Fire

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE DEVICE DID NOT WORK EVEN WITH THE BATTERY WAS ASSEMBLED. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25427239 ↗
  • MalfunctionJun 3, 2026Ejection Problem

    IT WAS REPORTED THAT DURING AN THORACIC PROCEDURE, AN ENDOSCOPIC STAPLER WAS USED TO REMOVE A BLEB FROM A PATIENT, AND THE STAPLE CARTRIDGE WAS STUCK IN THE STAPLER AFTER FIRING. THE STAPLER WAS UNABLE TO BE RELOADED RENDERING IT USELESS. A NEW STAPLER HAD TO BE OPENED IN ORDER TO FINISH THE SURGERY. THERE WAS NO PATIENT CONSEQUENCE NOR SURGICAL DELAY REPORTED. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25376178 ↗
  • MalfunctionJun 1, 2026Ejection Problem

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE ECH45S STAPLER WAS SUCCESSFULLY LOADED AND FIRED FOUR TIMES. UPON ATTEMPTING TO LOAD THE STAPLER FOR A FIFTH FIRING, THE RELOAD WOULD NOT FIT OR SNAP INTO PLACE. A DIFFERENT RELOAD WAS ATTEMPTED BUT ALSO WOULD NOT LOAD PROPERLY, CONFIRMING THAT THE ISSUE WAS WITH THE STAPLER ITSELF. A SECOND ECH45S STAPLER WAS OPENED. ONCE AGAIN, AFTER 4 SUCCESSFUL FIRINGS, THE FIFTH RELOAD FOR THE SECOND STAPLER WOULD NOT LOAD PROPERLY. THE MEDICAL STAFF REPEATED THE TROUBLE SHOOTING STEP OF ATTEMPTING A NEW RELOAD, WHICH WAS ALSO UNSUCCESSFUL. A THIRD ECH45S STAPLER WAS OPENED TO COMPLETE THE FINAL 2-3 FIRINGS FOR THE PROCEDURE. THE FIRST TWO ECH45S STAPLERS INVOLVED IN THE EVENT ARE FROM THE SAME LOT. PROCEDURE WAS DELAYED BY 20 MINUTES DUE TO THE REPORTED EVENT. NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25355375 ↗
  • MalfunctionJun 1, 2026Failure to Fire; Difficult to Open or Close; Noise, Audible

    IT WAS REPORTED THAT DURING AN UNKNOWN THORACIC PROCEDURE, AFTER OPENING THE NEW DEVICE AND 2 FIRINGS, THE MOTOR SOUND BECAME STRANGE AT THE 3RD FIRING. THE KNIFE STOPPED MIDWAY, SO THE HOSPITAL USED THE MANUAL OVERRIDE LEVER TO DEAL WITH THIS ISSUE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25351022 ↗
  • MalfunctionJun 1, 2026Ejection Problem

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, THE ECH45S STAPLER WAS SUCCESSFULLY LOADED AND FIRED FOUR TIMES. UPON ATTEMPTING TO LOAD THE STAPLER FOR A FIFTH FIRING, THE RELOAD WOULD NOT FIT OR SNAP INTO PLACE. A DIFFERENT RELOAD WAS ATTEMPTED BUT ALSO WOULD NOT LOAD PROPERLY, CONFIRMING THAT THE ISSUE WAS WITH THE STAPLER ITSELF. A SECOND ECH45S STAPLER WAS OPENED. ONCE AGAIN, AFTER 4 SUCCESSFUL FIRINGS, THE FIFTH RELOAD FOR THE SECOND STAPLER WOULD NOT LOAD PROPERLY. THE MEDICAL STAFF REPEATED THE TROUBLE SHOOTING STEP OF ATTEMPTING A NEW RELOAD, WHICH WAS ALSO UNSUCCESSFUL. A THIRD ECH45S STAPLER WAS OPENED TO COMPLETE THE FINAL 2-3 FIRINGS FOR THE PROCEDURE. THE FIRST TWO ECH45S STAPLERS INVOLVED IN THE EVENT ARE FROM THE SAME LOT. PROCEDURE WAS DELAYED BY 20 MINUTES DUE TO THE REPORTED EVENT. NO PATIENT CONSEQUENCE REPORTED. NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25350865 ↗
  • MalfunctionJun 1, 2026Failure to Fire; Difficult to Open or Close; Mechanical Jam

    IT WAS REPORTED THAT DURING AN UNSPECIFIED SURGICAL PROCEDURE, IT WAS OBSERVED THAT THE KNIFE DID NOT MOVE FORWARD NOR RETURNED UPON FIRING. SO THE MANUAL OVERRIDE LEVER WAS USED, AND THE KNIFE WAS RETURNED AND RELEASED MANUALLY. THE FIRING NUMBER AT WHICH THE ISSUE OCCURRED WAS UNKNOWN. THE CARTRIDGE COLOR WAS GOLD. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348448 ↗

FDA clearance and classification

The FDA regulates product code GAG (stapler, surgical) under 21 CFR 878.4740 as a Class II device. 21 CFR 878.4740 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: A surgical stapler for internal use is a specialized prescription device used to deliver compatible staples to internal tissues during surgery for resection, transection, and creating anastomoses.

Common questions about the ECHELON ECH45S

Has the ECHELON ECH45S been recalled?

No FDA recall record in our data names this exact device identifier (DI 10705036035008), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the ECHELON ECH45S?

The FDA's MAUDE database lists 261 adverse event reports for this device, received between Feb 23, 2023 and Sep 24, 2026. In those reports, the most common specific device problems are “Failure to Fire” (100), “Difficult to Open or Close” (88) and “Positioning Failure” (34); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (10), “Unspecified Tissue Injury” (7) and “Failure to Anastomose” (5). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the ECHELON ECH45S?

None of the reports are classified as deaths, 18 as injuries and 243 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the ECHELON ECH45S, and is it still sold?

The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON, model ECH45S). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Stapler, surgical devices

All Stapler, surgical devices →