ECHELON ECH60SRecalls & reported problems
Ethicon Endo-Surgery, LLC · ECHELON 3000 60MM STANDARD · Class 2
All models of ECHELON (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This is an Ethicon Echelon surgical stapler. The 615 reports mostly describe the stapler failing to fire, being hard to open or close, and failing to form staples, with some noting jams and bleeding. The reports mention 1 death and 54 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 608
- Deaths reported
- 1
- Injuries reported
- 55
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 608 adverse event reports that list this exact device (DI 10705036035015), received between Feb 17, 2023 and Sep 24, 2026 — classified as 1 death, 55 injuries, 552 malfunctions.
The most frequently reported specific device problem is “Failure to Fire” (219 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Failure to Fire219
- Difficult to Open or Close185
- Failure to Form Staple80
- Positioning Failure65
- Mechanical Jam53
- Adverse Event Without Identified Device or Use Problem50
- Material Integrity Problem45
- Ejection Problem42
Patient problems
- No Clinical Signs, Symptoms or Conditions551
- Hemorrhage/Blood Loss/Bleeding35
- Failure to Anastomose23
- Unspecified Tissue Injury11
- Insufficient Information9
- Pneumothorax6
- Cardiac Arrest3
- Low Blood Pressure/ Hypotension3
Recent adverse event reports
The 25 most recent of 608 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 24, 2026Difficult to Open or Close
It was reported that, during an unknown surgery, the jaws were closed to attach the SLR, but the jaws did not open when the scrub nurse was preparing the device during setting before use. The cartridge color was Gold. Another device was used to complete the case. There were no adverse consequences to the patient. No further information is available.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883541 ↗InjurySep 17, 2026Failure to Form Staple
It was reported that the doctor performed a pulmonary segmentectomy on (b)(6) 2026 and used our stapler and reloads, as well as our staple line reinforcement. There were no complications on the day of the procedure. The staple line was intact, and no bleeding was observed. The procedure was completed successfully without any issues. The doctor also used a hemostatic product from another manufacturer. Two weeks after the procedure, on (b)(6) 2026, at another hospital, the doctor had to perform a surgical reintervention on the patient. During this reintervention, the physician reported the presence of cystic fibrosis and indicated that the staple line was open. I shared some photographs with our team. During this subsequent procedure, the physician stated that no additional medical devices were used and that no suturing was performed because the tissue was extremely friable. The doctor would like to better understand what may have occurred that resulted in the staple line opening. There was no delay reported. Additional information requested.
Patient problems: Fibrosis
MAUDE report 50808678 ↗MalfunctionSep 17, 2026Failure to Fire
It was reported that during an unspecified surgical procedure, it was observed that the blade on the device would not come out during firing. There was no patient consequence nor surgical delay reported. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50804368 ↗MalfunctionSep 15, 2026Failure to Fire
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50778208 ↗MalfunctionAug 24, 2026Failure to Form Staple
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50410804 ↗MalfunctionAug 19, 2026Failure to Form Staple; Failure to Fire; Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50324203 ↗MalfunctionAug 19, 2026Positioning Failure
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50324176 ↗MalfunctionAug 17, 2026Insufficient Device Problem Information
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50278351 ↗MalfunctionAug 14, 2026Product Quality Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50260433 ↗MalfunctionAug 14, 2026Failure to Form Staple; No Apparent Adverse Event
Patient problems: No Clinical Signs, Symptoms or Conditions; Pneumothorax; Unspecified Tissue Injury
MAUDE report 50260202 ↗MalfunctionAug 14, 2026Unintended Ejection; Failure to Fire; Difficult to Open or Close
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50256142 ↗MalfunctionAug 14, 2026Ejection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50255523 ↗MalfunctionAug 14, 2026Ejection Problem
It was reported that during an unknown procedure, it was observed that the cartridge could not be loaded to the main body. Another device was used to complete the case. There were no adverse consequences to the patient. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50255288 ↗MalfunctionAug 14, 2026Positioning Failure; Difficult to Open or Close; Noise, Audible
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50254859 ↗MalfunctionAug 14, 2026Material Integrity Problem; Firing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50254122 ↗InjuryAug 6, 2026Adverse Event Without Identified Device or Use Problem
It was reported that on July 3, during VATS (left upper lobectomy)for lung cancer, a GST60T with an SLR was loaded into an ECH60S, and the interlobar fissure was divided with two firings. The procedure was completed without issue. The patient was discharged on July 11. However, on July 13, the patient was emergently readmitted with cardiac tamponade (shock). The examination was performed after that, and it was confirmed that the blood accumulated in the pericardial sac. Reoperation was performed on July 16, and the patient recovered well and was discharged. During the reoperation, inspection of the pericardium and the staple line (lung parenchyma) revealed one malformed staple line that remained in a U-shape. The doctor explained that this malformed staple, found on the staple line (lung parenchyma) created during the second firing across the interlobar fissure using the EC3000, most likely injured the pericardium and myocardium, resulting in cardiac tamponade. Patient background information is unknown. The cartridge color was Black. No device will be returning. No additional information provided.
Patient problems: Cardiac Tamponade; Hemorrhage/Blood Loss/Bleeding; Shock; Unspecified Tissue Injury
MAUDE report 50164510 ↗MalfunctionAug 6, 2026Appropriate Device Problem Term/Code Not Available
Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Tissue Injury
MAUDE report 50161468 ↗MalfunctionAug 4, 2026Ejection Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50133428 ↗MalfunctionAug 4, 2026Difficult to Open or Close
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50131218 ↗MalfunctionAug 3, 2026Failure to Fire; Difficult to Open or Close; Noise, Audible
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50115910 ↗MalfunctionJul 27, 2026Difficult to Open or Close
It was reported that during a Laparoscopic Rectal case, when closing the jaws, the ratchet mechanism did not engage. Another device was used to complete the case. There were no adverse consequences to the patient. No additional information provided.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50038415 ↗InjuryJul 27, 2026Adverse Event Without Identified Device or Use Problem
It was reported that an uneventful Sleeve Gastrectomy was performed on (b)(6) and the patient was discharged the next day. On (b)(6) the patient had nausea and abdomen pain. Seen by GP had a CT, showed sleeve leak with gas in abdomen and more pain on (b)(6) and presented to ED (b)(6). The patient had CT and endoscoped, 4x2cm cavity noted on proximal staple line, just below GOJ. Patient is stable, on ward, not septic. The last firing on the staple last was GST60G with SLR. Intraoperatively the staple line looked complete, with good hemostasis, no additional intervention required intra op. Patient closed and discharged home the next day. Additional information requested.
Patient problems: Abdominal Pain; Dyspnea; Failure to Anastomose; Nausea
MAUDE report 50019113 ↗MalfunctionJul 9, 2026Failure to Fire; Difficult to Open or Close; Mechanical Jam
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, IT WAS OBSERVED THAT THE CLOSING MOTION FELT DIFFERENT FROM USUAL WHEN CLAMPING BUT SINCE IT OPERATED, FIRING WAS PERFORMED. HOWEVER, THE KNIFE DID NOT RETURN, SO THE KNIFE WAS RETURNED USING THE HOME BUTTON. AFTER THAT, WHEN ATTEMPTING THE SECOND FIRING, THE KNIFE DID NOT MOVE. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO ADDITIONAL INFORMATION PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25769150 ↗InjuryJul 1, 2026Failure to Form Staple; Failure to Fire; Difficult to Open or Close; Material Integrity Problem; Mechanical Jam; Adverse Event Without Identified Device or Use Problem; Noise, Audible
IT WAS REPORTED THAT DURING A LEFT UPPER LOBECTOMY PROCEDURE, THE PATIENT¿S LOBE WAS DESCRIBED AS THICK, PREVENTING THE STAPLER FROM CLOSING INITIALLY. THE SURGEON SUBSEQUENTLY FORCED THE DEVICE TO CLOSE AND HELD COMPRESSION FOR TWO MINUTES. UPON FIRING, A LOUD POP WAS HEARD, AND THE KNIFE DID NOT ADVANCE FURTHER. TWO PLEE60A DEVICES REQUIRED MANUAL OVERRIDE TO OPEN. ADDITIONALLY, BOTH PLEE60A DEVICES RAN OUT OF BATTERY CHARGE DURING FIRING, AND THE ANVIL ON THE ECH45L AND ECH60 DEVICES WAS NOTED TO BE BENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25681501 ↗MalfunctionJul 1, 2026Difficult to Open or Close; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE ARTICULATION JOINT BENDS AND THE JAW DOES NOT CLOSE WELL ARE THERE ANY REPORTED CASES OF HARM TO PATIENTS OR USERS? UNKNOWN WAS THERE A SIGNIFICANT DELAY? UNKNOWN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25678280 ↗
FDA clearance and classification
The FDA regulates product code GAG (stapler, surgical) under 21 CFR 878.4740 as a Class II device. 21 CFR 878.4740 ↗
Devices of this type normally need 510(k) clearance before sale.
FDA definition of this device type: A surgical stapler for internal use is a specialized prescription device used to deliver compatible staples to internal tissues during surgery for resection, transection, and creating anastomoses.
Common questions about the ECHELON ECH60S
Has the ECHELON ECH60S been recalled?
No FDA recall record in our data names this exact device identifier (DI 10705036035015), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the ECHELON ECH60S?
The FDA's MAUDE database lists 608 adverse event reports for this device, received between Feb 17, 2023 and Sep 24, 2026. In those reports, the most common specific device problems are “Failure to Fire” (219), “Difficult to Open or Close” (185) and “Failure to Form Staple” (80); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (35), “Failure to Anastomose” (23) and “Unspecified Tissue Injury” (11). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the ECHELON ECH60S?
1 report is classified as a death, 55 as injuries and 552 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the ECHELON ECH60S, and is it still sold?
The FDA's device database lists Ethicon Endo-Surgery, LLC as the company responsible for it (brand ECHELON, model ECH60S). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Stapler, surgical devices
- ECHELON ECH60LEthicon Endo-Surgery, LLC · Stapler, surgical463 reports
- ECHELON ECH45SEthicon Endo-Surgery, LLC · Stapler, surgical261 reports
- EEA TRIEEA28XTCovidien LP · Stapler, surgical218 reports
- ECHELON ECH60CEthicon Endo-Surgery, LLC · Stapler, surgical194 reports
- ECHELON ECH45CEthicon Endo-Surgery, LLC · Stapler, surgical132 reports
- EEA TRIEEA31XTCovidien LP · Stapler, surgical73 reports
- ETHICON EC3D60LEthicon Endo-Surgery, LLC · Stapler, surgical39 reports
- Echelon Linear Cutter GLC100Ethicon Endo-Surgery, LLC · Stapler, surgical36 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page