directCHECK Whole Blood Controls DCJLR-NRecalls & reported problems
Accriva Diagnostics Holdings, Inc. · directCHECK ACT-LR - Normal 15/box · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10711234170243), received between Apr 21, 2017 and May 11, 2018 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Human Factors Issue” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Human Factors Issue1
- Improper or Incorrect Procedure or Method1
Patient problems
- Laceration(s)2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 11, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method
HEALTHCARE PROFESSIONAL REPORTED THAT AN END USER SUSTAINED AN INJURY WHILE HANDLING A DIRECTCHECK QUALITY CONTROL. THIS CONTROL IS PACKAGED IN A GLASS AMPULE INSIDE A CRUSHABLE PLASTIC DROPPER VIAL CONTAINING DILUENT. THE END USER CRUSHED THE VIAL TO RECONSTITUTE THE CONTROL AND SUSTAINED A SMALL CUT TO HER RIGHT INDEX FINGER WHICH WAS CAUSED BY A SHARD OF GLASS. THE END-USER WORE GLOVES BUT FAILED TO UTILIZE THE PROTECTIVE SLEEVE PROVIDED WITH THE PRODUCT . THE PURPOSE OF THE SLEEVE IS TO SAFEGUARD THE END USER AGAINST POTENTIAL INJURY WHILE HANDLING THE CONTROL. THE END USER FILED A REPORT WITH HER SUPERVISOR AND WAS SEEN BY OCCUPATIONAL HEALTH. NO LAB TESTS WERE DRAWN AND GENERAL SITE CARE WAS PERFORMED. THE NEXT DAY, THE END USER FOLLOWED UP WITH OCCUPATIONAL HEALTH. THE PUNCTURE SITE SHOWED NO PROBLEMS AND WAS HEALING.
Patient problems: Laceration(s)
MAUDE report 7509440 ↗MalfunctionApr 21, 2017Human Factors Issue; Adverse Event Without Identified Device or Use Problem; Human Factors Issue; Adverse Event Without Identified Device or Use Problem
HEALTHCARE PROFESSIONAL REPORTED THAT AN END USER SUSTAINED AN INJURY WHILE DISPENSING A DIRECT CHECK QUALITY CONTROL. THIS CONTROL IS PACKAGED IN A GLASS AMPULE INSIDE A CRUSHABLE PLASTIC DROPPER VIAL CONTAINING DILUENT. WHEN RECONSTITUTING THE CONTROL, THE END-USER WORE GLOVES BUT FAILED TO UTILIZE THE PROTECTIVE SLEEVE PROVIDED WITH THE PRODUCT . THE PURPOSE OF THE SLEEVE IS TO SAFEGUARD THE END USER AGAINST POTENTIAL INJURY WHILE HANDLING THE CONTROL. THE END USER SQUEEZED THE VIAL AND SUSTAINED A SMALL CUT TO HER RIGHT INDEX FINGER WHICH WAS CAUSED BY A SHARP PIECE OF PLASTIC. THE END USER WASHED THE AFFECTED AREA WITH SOAP AND WATER AND APPLIED A BAND AID. NO OTHER MEDICAL CARE WAS REQUIRED AND NO COMPLICATIONS WERE DOCUMENTED.
Patient problems: Laceration(s)
MAUDE report 6513172 ↗
Other Plasma, Coagulation Control devices
- STA® - System Control N + P 00678Diagnostica Stago · PLASMA, COAGULATION CONTROL1 recall0 reports
- directCHECK ACT-LR Whole Blood Control, Level 2 000DCGLR-2Accriva Diagnostics Holdings, Inc. · Plasma, Coagulation Control1 recall0 reports
- 5-PLASMA Bivalirudin Control 5D-449055-Diagnostics Ltd · Plasma, Coagulation Control1 recall0 reports
- BIOPHEN UFH Control Plasma 223101Hyphen Biomed · Plasma, Coagulation Control1 recall0 reports
- directCHECK Whole Blood Controls DCJLR-AAccriva Diagnostics Holdings, Inc. · Plasma, Coagulation Control5 reports
- Control Plasma N 10446235Siemens Healthcare Diagnostics Products GmbH · PLASMA, COAGULATION CONTROL1 report
- directCHECK Whole Blood Controls DCJACT-AAccriva Diagnostics Holdings, Inc. · Plasma, Coagulation Control1 report
- directCHECK Whole Blood Controls DCJCPT-AAccriva Diagnostics Holdings, Inc. · Plasma, Coagulation Control1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page