Skip to content

STA® - System Control N + P 00678Recalls & reported problems

Diagnostica Stago · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

1 other recall cover devices cleared under the same FDA submission (K943518); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Nov 18, 2019Class IITerminated

    A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also observed on several External Quality Assessment (EQA) results.

    Z-0738-2020 · Root cause: Process control · Recall details · FDA record ↗

FDA clearance and classification

Cleared through 510(k) K943518 on Mar 21, 1995 (decision: Substantially Equivalent). Submitted by American Bioproducts Co. as “Sta System Control [N] + [P] Kit”. The FDA review took 243 days. FDA record ↗

The FDA regulates product code GGN (plasma, coagulation control) under 21 CFR 864.5425 as a Class II device. 21 CFR 864.5425 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K943518. One submission often covers many models, so check the product and lot details in each record.

  • Z-1044-2012 · Aug 23, 2011 · Terminated
    STA - System Control N + P Product Usage: The STA -System Control N+P kit provides a normal plasma and an abnormal plasma intended for use as two control levels for the following

Common questions about the STA® - System Control N + P 00678

Has the STA® - System Control N + P 00678 been recalled?

Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0738-2020 (Class II), began on Nov 18, 2019: A positive bias was confirmed to exists on the STA - Unicalibrator levels with STA - Deficient IX and STA - Deficient VIII when compared to the International Standard (IS) for factor VIII or IX. This bias is also… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Is the STA® - System Control N + P 00678 FDA approved?

It was cleared through 510(k) K943518 on Mar 21, 1995. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the STA® - System Control N + P 00678, and is it still sold?

The FDA's device database lists Diagnostica Stago as the company responsible for it (brand STA® - System Control N + P, model 00678). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other PLASMA, COAGULATION CONTROL devices

All PLASMA, COAGULATION CONTROL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page