Skip to content

Celerity LCB044Recalls & reported problems

Steris Corporation · The Celerity 20 HP Biological Indicator is used with the V-PRO Low Temperature Sterilization Systems and STERRAD NX, 100S and 100NX sterilizers. · Class 2

All models of Celerity (5)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10724995154230), received between Aug 31, 2022 and Aug 28, 2023 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. May 25, 2022Class IIIOpen, Classified

    Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

    Z-1793-2022 · Root cause: Environmental control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Use of Device Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Insufficient Device Problem Information1
  • Leak/Splash1

Patient problems

  • Burn(s)2
  • No Clinical Signs, Symptoms or Conditions1
  • Skin Discoloration1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT AN EMPLOYEE WAS IN THE PROCESS OF ACTIVATING THE CELERITY 20 HP BIOLOGICAL INDICATOR AND OBSERVED "WHITE SPOTS" ON THEIR FINGERS. THE EMPLOYEE SOUGHT TREATMENT AND PLACED THEIR FINGERS IN WATER. THE EMPLOYEE WAS ALSO ADVISED TO APPLY OINTMENT (VASELINE).

    Patient problems: Skin Discoloration

    MAUDE report 17631820 ↗
  • MalfunctionFeb 10, 2023Leak/Splash; Leak/Splash

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT MW5114138 THAT THERE WAS NO MEDIA IN THE CAP FOR BIOLOGICAL TESTING. NO REPORT OF INJURY OR PROCEDURE DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16358303 ↗
  • MalfunctionOct 28, 2022Use of Device Problem; Use of Device Problem

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A "BURN" AFTER HANDLING THEIR CELERITY 20 HP BIOLOGICAL INDICATOR. THE EMPLOYEE WASHED THEIR HANDS AND WENT TO THE ER.

    Patient problems: Burn(s)

    MAUDE report 15693195 ↗
  • MalfunctionAug 31, 2022Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A "BURN" WHILE HANDLING THEIR BIOLOGICAL INDICATOR. THE EMPLOYEE SELF-TREATED BY RINSING THE BURN. IT IS UNKNOWN IF ADDITIONAL MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Burn(s)

    MAUDE report 15328003 ↗

FDA clearance and classification

Cleared through Special 510(k) K251452 on Jun 6, 2025 (decision: Substantially Equivalent). Submitted by Steris as “Celerity 5 HP Biological Indicator (LCB052); Celerity 5 HP Challenge Pack (LCB059); Celerity 20 HP Biological Indicator (LCB044); Celerity 20 HP Challenge Pack (LCB045)”. The FDA review took 28 days. FDA record ↗

Cleared through Special 510(k) K250044 on Feb 6, 2025 (decision: Substantially Equivalent). Submitted by STERIS Corporation as “Celerity 5 HP Biological Indicator (LCB052); Celerity 5 HP Challenge Pack (LCB059); Celerity 20 HP Biological Indicator (LCB044); Celerity 20 HP Challenge Pack (LCB045)”. The FDA review took 27 days. FDA record ↗

Cleared through 510(k) K231490 on Aug 7, 2023 (decision: Substantially Equivalent). Submitted by STERIS Corporation as “Celerity 20 HP Biological Indicator; VERIFY V24 Self-Contained Biological Indicator”. The FDA review took 76 days. FDA record ↗

…and 5 more FDA submissions listed for this device.

The FDA regulates product code FRC (indicator, biological sterilization process) under 21 CFR 880.2800 as a Class II device. 21 CFR 880.2800 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Celerity LCB044

Has the Celerity LCB044 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-1793-2022 (Class III), began on May 25, 2022: Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the Celerity LCB044?

The FDA's MAUDE database lists 4 adverse event reports for this device, received between Aug 31, 2022 and Aug 28, 2023. In those reports, the most common specific device problems are “Use of Device Problem” (2) and “Leak/Splash” (1); the most common patient problems are “Burn(s)” (2) and “Skin Discoloration” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Celerity LCB044?

None of the reports are classified as deaths, 0 as injuries and 4 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Celerity LCB044 FDA approved?

It was cleared through 510(k) K171504 on Aug 22, 2017; its record lists 8 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Who makes the Celerity LCB044, and is it still sold?

The FDA's device database lists Steris Corporation as the company responsible for it (brand Celerity, model LCB044). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Indicator, biological sterilization process devices

All Indicator, biological sterilization process devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page