Skip to content

VERIFY S3069Recalls & reported problems

Steris Corporation · The VERIFY Disposable Biological Steam Test Pack is a single use test pack designed for monitoring the effectiveness of common steam sterilizing processes, including 250 degrees F (121 degrees C) gravity cycles and 270 degrees F (132 C) prevacuum cycles. · Class 2

All models of VERIFY (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

3 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
3

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

3 FDA recall records name this device's identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Recalls naming this device

  1. Aug 31, 2026Class IIOpen, Classified

    Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction

    Z-0075-2027 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Dec 12, 2022Class IIOpen, Classified

    Biological indicator (BI) used to promote growth of the G. stearothermophilus and B. atrophaeus spores was inconsistent in promoting growth throughout lot , some BIs within the lot performed as intended while others did not. When used to monitor steam and ethylene oxide sterilization cycles, the potential hazard is that the BI will show no growth indicating that the cycle was effective, even if it may not have been.

    Z-0781-2023 · Root cause: Employee error · Recall details · FDA record ↗

  3. May 25, 2022Class IIIOpen, Classified

    Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

    Z-1793-2022 · Root cause: Environmental control · Recall details · FDA record ↗

FDA clearance and classification

The FDA regulates product code FRC (indicator, biological sterilization process) under 21 CFR 880.2800 as a Class II device. 21 CFR 880.2800 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the VERIFY S3069

Has the VERIFY S3069 been recalled?

Yes. 3 FDA recall records name this device's identifier: 3 still open. The most recent, Z-0075-2027 (Class II), began on Aug 31, 2026: Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

Who makes the VERIFY S3069, and is it still sold?

The FDA's device database lists Steris Corporation as the company responsible for it (brand VERIFY, model S3069). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Indicator, biological sterilization process devices

All Indicator, biological sterilization process devices →

Sources and updates