Skip to content

Parker Flex-Tip H-PFTVPU-80Recalls & reported problems

Salter Labs · Parker Trach Vac with Flex-Tip and ThinCuff (H-PFTVPU), 8.0mm · Class 2

All models of Parker Flex-Tip (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10749065101808), received between Jun 10, 2025 and Sep 14, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Gas/Air Leak” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak4

Patient problems

  • Low Oxygen Saturation2
  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 14, 2026Gas/Air Leak

    It was reported that one week after intubation, the automatic cuff pressure monitor alarm sounded continuously. The tube was extubated, and the patient was immediately reintubated with a new tube. There was no reported delay in therapy, harm, or injury associated with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50756313 ↗
  • MalfunctionMar 19, 2026Gas/Air Leak

    IT WAS REPORTED THAT INTUBATION WAS PERFORMED ON (B)(6) 2026, AT 10:00 PM. ON (B)(6) 2026, THE ENDOTRACHEAL TUBE WAS EXCHANGED DUE TO DECREASED VENTILATION VOLUME. FOLLOWING EXTUBATION, THE MEDICAL STAFF INSPECTED THE DEVICE AND NOTED THAT THE CUFF WAS UNABLE TO BE INFLATED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 24638800 ↗
  • MalfunctionJun 20, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED, ON (B)(6) 2025 IN THE MORNING, INTUBATION OF THE PATIENT WAS PERFORMED. SUCTIONING WAS PERFORMED EVERY HOUR AFTER THE START OF USE. ON (B)(6) 2025 AT 10:00 A.M., SUCTIONING WAS PERFORMED DUE TO THE PATIENT CHOKING; A CUFF LEAK OCCURRED AFTER SUCTIONING. THE TIDAL VOLUME AND SPO2 OF THE PATIENT DROPPED. THE DOCTOR REINTUBATED THE PATIENT WITH A NEW TUBE. IT WAS ADDITIONALLY REPORTED, THE CUFF HAS CONTINUALLY INFLATED FOR OVER 100 HOURS WHEN THE SUCTION END OF THE TUBE WAS CAPPED.

    Patient problems: Low Oxygen Saturation

    MAUDE report 22302957 ↗
  • MalfunctionJun 10, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED, AFTER INTUBATION IN THE ER, THE PATIENT WAS TRANSFERRED TO THE WARD. IMMEDIATELY AFTER PLACING THE PATIENT ON THE VENTILATOR IN THE WARD, A LOW PRESSURE ALARM ON THE VENTILATOR WENT OFF AND A CUFF LEAK OCCURRED (WITHIN THE FIRST HOUR OF USE). ADDITIONAL AIR WAS INJECTED, BUT THE LEAK WAS NOT RESOLVED. THE PATIENT WAS RE-INTUBATED; THERE WAS NO REPORTED INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22177068 ↗

Other Tube, Tracheal (W/Wo Connector) devices

All Tube, Tracheal (W/Wo Connector) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page