Parker Flex-Tip H-PFTVVC-70Recalls & reported problems

Salter Labs · Parker Trach Vac with Flex-Tip and PVC Cuff (H-PFTVVC), 7.0mm · Class 2

All models of Parker Flex-Tip (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10749065191700), received between Jul 2, 2024 and Aug 1, 2025 — classified as 2 injuries, 1 malfunction.

The most frequently reported specific device problem is “Gas/Air Leak” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Gas/Air Leak2
  • Leak/Splash1
  • Material Split, Cut or Torn1

Patient problems

  • Insufficient Information1
  • Respiratory Insufficiency1
  • Swelling/ Edema1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 1, 2025Gas/Air Leak; Gas/Air Leak

    IT WAS REPORTED, AUTOMATIC CUFF PRESSURE GAUGE ALARMED FREQUENTLY AND RE-INTUBATION WAS PERFORMED. AFTER CHECKING FOR LEAKS BY DIPPING THE CUFF IN WATER, A PINHOLE-SIZED HOLE WAS FOUND IN THE CUFF. IT WAS ADDITIONALLY REPORTED; AN ANCHOR FAST WAS USED DURING INTUBATION. POSITION CHANGED EVERY 2-3 HOURS TO LEFT AND RIGHT. THE TUBE POSITION WAS CHANGED LEFT OR RIGHT WITH ANCHOR FAST IN USE. THERE WAS NO REPORTED INJURY.

    Patient problems: Insufficient Information

    MAUDE report 22691740 ↗
  • InjurySep 27, 2024Material Split, Cut or Torn; Material Split, Cut or Torn

    THE TUBE WAS INTUBATED INTO THE PATIENT ON (B)(6). ON (B)(6), THE VENTILATOR WAS REMOVED, AND AN ATTEMPT WAS MADE TO EXTUBATE THE PATIENT, BUT THE TUBE COULD NOT BE REMOVED. AT THAT TIME, THE TUBE COULD NOT BE PUSHED IN. SUSPECTING EDEMA, A STEROID WAS ADMINISTERED, AND AFTER A DAY OF OBSERVATION, ANOTHER ATTEMPT WAS MADE ON (B)(6), BUT THE PATIENT COULD NOT BE EXTUBATED. ON (B)(6), THE SURGEON WAS ABLE TO PUSH THE TUBE IN.THE ANESTHESIOLOGY DEPARTMENT CONSULTED WITH THE SURGEON AND OTOLARYNGOLOGY DEPARTMENT, AND ON (B)(6), A TRACHEOTOMY WAS PERFORMED IN THE OP ROOM, AND ONLY THE TIP OF THE TUBE WAS REMOVED FROM THE TRACHEOSTOMY SITE. IN ORDER TO REMOVE THE TUBE TIP, A LARGER INCISION WAS MADE THAN A NORMAL TRACHEOTOMY. WHEN THE TUBE WAS REMOVED, THE PATIENT WAS VENTILATED WITH THE INTUBATED TUBE UP TO THE TRACHEOSTOMY. FROM THE TRACHEOSTOMY, THE END OF THE TUBE WAS FIRST CUT OFF, AND ONLY THE END OF THE TUBE, INCLUDING THE CUFF, WAS REMOVED FROM THE TRACHEOSTOMY SITE. THE TUBE COULD NOT BE PUSHED IN.,

    Patient problems: Swelling/ Edema

    MAUDE report 20328133 ↗
  • InjuryJul 2, 2024Leak/Splash; Gas/Air Leak; Leak/Splash; Gas/Air Leak

    THE INFLATING TUBE WAS TORN AND A LEAK WAS OBSERVED FROM THERE. THE PATIENT WAS MOVED TO ICU WHEN A GURGLING SOUND WAS HEARD COMING FROM THE PATIENTS' MOUTH. THE INFLATING TUBE WAS TORN AND A LEAK WAS OBSERVED FROM THERE. THEY CHECKED THE ORAL CAVITY AND FOUND THAT THE SOUND WAS COMING FROM THE TRACHEA.

    Patient problems: Respiratory Insufficiency

    MAUDE report 19658985 ↗

Other Tube, Tracheal (W/Wo Connector) devices

All Tube, Tracheal (W/Wo Connector) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page