VITROS 6801428Recalls & reported problems
Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products Anti-HBc Reagent Pack · Class 2
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 10758750001217), received between Jan 4, 2023 and Aug 7, 2026 — classified as 11 malfunctions.
The most frequently reported specific device problem is “False Positive Result” (8 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- False Positive Result8
- False Negative Result4
- Incorrect Measurement2
- Non Reproducible Results1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50174146 ↗MalfunctionAug 7, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50173294 ↗MalfunctionJul 8, 2026False Positive Result
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25750059 ↗MalfunctionApr 10, 2026False Positive Result; Non Reproducible Results
A FIELD APPLICATION SPECIALIST (FAS), ON BEHALF OF A CUSTOMER, CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTIONS CENTER (TSC) TO REPORT A DISCORDANT, REACTIVE (POSITIVE) VITROS AHBC RESULT OBTAINED FROM A SINGLE PATIENT SAMPLE PROCESSED USING VITROS IMMUNODIAGNOSTIC PRODUCTS ANTI-HBC (AHBC) REAGENT PACK LOT 2910 ON A VITROS 5600 INTEGRATED SYSTEM. THE RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO THE REPEAT RESULT OBTAINED FROM THE SAME SAMPLE. PATIENT SAMPLE RESULT OF 0.1 S/C (REACTIVE) VERSUS THE EXPECTED RESULT OF 3.2 S/C (NON-REACTIVE) BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE DISCORDANT, REACTIVE VITROS AHBC RESULT WAS NOT REPORTED FROM THE LABORATORY. THERE HAVE BEEN NO ALLEGATIONS OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT 614102.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24844146 ↗MalfunctionMar 30, 2026False Positive Result; Incorrect Measurement
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT DISCORDANT REACTIVE ANTI-HBC (AHBC) RESULTS OBTAINED WHEN TWO DIFFERENT SAMPLES FROM THE SAME PATIENT WERE TESTED USING VITROS IMMUNODIAGNOSTIC PRODUCTS ANTI-HBC (AHBC) REAGENT PACK LOT 2910, ON A VITROS XT 7600 INTEGRATED SYSTEM. THE RESULTS WERE CONSIDERED DISCORDANT WHEN COMPARED TO A NEGATIVE RESULT OBTAINED FROM A NON-VITROS ABBOTT ALINITY METHOD. PATIENT 1, SAMPLE 1 VITROS AHBC RESULT OF 0.84 S/C (REACTIVE) VS THE NON-VITROS AHBC RESULT OF 0.47 S/C (NON-REACTIVE). PATIENT 1, SAMPLE 2 VITROS AHBC RESULT OF 0.56 S/C (REACTIVE) VS THE NON-VITROS AHBC RESULT OF 0.47 S/C (NON-REACTIVE). BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE DISCORDANT, FALSE REACTIVE VITROS AHBC RESULTS WERE NOT REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF TWO 3500A FORMS BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24713027 ↗MalfunctionMar 30, 2026False Positive Result; Incorrect Measurement
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT DISCORDANT REACTIVE ANTI-HBC (AHBC) RESULTS OBTAINED WHEN TWO DIFFERENT SAMPLES FROM THE SAME PATIENT WERE TESTED USING VITROS IMMUNODIAGNOSTIC PRODUCTS ANTI-HBC (AHBC) REAGENT PACK LOT: 2910, ON A VITROS XT 7600 INTEGRATED SYSTEM. THE RESULTS WERE CONSIDERED DISCORDANT WHEN COMPARED TO A NEGATIVE RESULT OBTAINED FROM A NON-VITROS ABBOTT ALINITY METHOD. PATIENT 1, SAMPLE 1 VITROS AHBC RESULT OF 0.84 S/C (REACTIVE) VS THE NON-VITROS AHBC RESULT OF 0.47 S/C (NON-REACTIVE). PATIENT 1, SAMPLE 2 VITROS AHBC RESULT OF 0.56 S/C (REACTIVE) VS THE NON-VITROS AHBC RESULT OF 0.47 S/C (NON-REACTIVE). BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE DISCORDANT, FALSE REACTIVE VITROS AHBC RESULTS WERE NOT REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF TWO 3500A FORMS BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER:
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24712999 ↗MalfunctionNov 19, 2025False Negative Result; False Negative Result
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A DISCORDANT, FALSE NEGATIVE ANTI-HBC (AHBC) RESULT WAS OBTAINED FROM A SINGLE PATIENT SAMPLE TESTED USING VITROS AHBC LOT 2870 ON A VITROS 5600 INTEGRATED SYSTEM. THE RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO A REACTIVE RESULT OBTAINED FROM AN UNKNOWN NON-VITROS ANTI-HBS METHOD AND A REACTIVE VITROS AHBC RESULT. PATIENT 1, SAMPLE 2 VITROS AHBC RESULT OF 2.62 S/C (NON-REACTIVE) VS THE VITROS AHBC RESULT OF 0.17 S/C (REACTIVE). BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE DISCORDANT, FALSE REACTIVE VITROS AHBC RESULT WAS NOT REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23594376 ↗MalfunctionApr 25, 2024False Negative Result; False Negative Result
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A FALSE NEGATIVE ANTI-HBC (AHBC) RESULT WAS OBTAINED WHEN A VITROS AHBC LEVEL 2 REACTIVE CONTROL (LOT 1240) WAS TESTED USING VITROS AHBC REAGENT LOT 2780 ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS AHBC RESULT OF 1.22 S/C (NEGATIVE) VERSUS THE EXPECTED RESULT OF REACTIVE BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE FALSE NEGATIVE VITROS AHBC RESULT WAS OBTAINED WHEN THE CUSTOMER WAS PROCESSING A NON-PATIENT FLUID AND WAS NOT REPORTED FROM THE LABORATORY. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19179975 ↗MalfunctionJan 10, 2024False Negative Result; False Negative Result
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A DISCORDANT, NON-REACTIVE VITROS ANTI-HBC (AHBC) RESULT FROM A SINGLE PATIENT SAMPLE WAS OBTAINED WHEN TESTED USING VITROS AHBC LOT 2710 ON A VITROS XT 7600 INTEGRATED SYSTEM. THE VITROS RESULT WAS CONSIDERED DISCORDANT WHEN COMPARED TO A NON-VITROS ABBOTT RESULT. VITROS AHBC RESULT OF 1.93 S/C (NON-REACTIVE) VS. THE EXPECTED RESULT OF POSITIVE. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE DISCORDANT, NON-REACTIVE VITROS AHBC RESULT WAS NOT REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(6) AND REPORTABILITY ASSESSMENT 604644.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18486984 ↗MalfunctionSep 28, 2023False Positive Result; False Positive Result
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT DISCORDANT, REACTIVE ANTI-HBC (AHBC) RESULTS WERE OBTAINED WHEN A SINGLE PATIENT SAMPLE WAS TESTED USING VITROS AHBC LOT 2730 ON A VITROS XT7600 INTEGRATED SYSTEM. THE RESULTS WERE CONSIDERED DISCORDANT WHEN COMPARED TO A NEGATIVE NON-VITROS ROCHE RESULT FOR THE SAME PATIENT SAMPLE. PATIENT 1, VITROS AHBC RESULTS OF 0.58, 0.56 AND 0.51 S/C (REACTIVE) VERSUS THE ROCHE RESULT OF NEGATIVE BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE DISCORDANT, REACTIVE VITROS AHBC RESULTS WERE NOT REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4). AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17831885 ↗MalfunctionJan 4, 2023False Negative Result; False Positive Result; False Negative Result; False Positive Result
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT DISCORDANT, NON-REACTIVE VITROS ANTI-HBC (AHBC) RESULTS AND DISCORDANT, POSITIVE VITROS ANTI-HBS (AHBS) SAMPLES FROM ONE PATIENT WERE TESTED USING VITROS AHBC LOT 2680 AND VITROS AHBS LOT 3990 ON A VITROS 5600 IMMUNODIAGNOSTIC SYSTEM. THE VITROS RESULTS WERE CONSIDERED DISCORDANT WHEN COMPARED TO RESULTS FROM A NON-VITROS (ABBOTT) METHOD. PATIENT 1, VITROS AHBS RESULTS OF 40.76 AND 48.27 MIU/L (POSITIVE) VERSUS THE EXPECTED RESULT OF NEGATIVE PATIENT 1, VITROS AHBC RESULT OF 1.98 AND 1.87 S/C (NON-REACTIVE) VERSUS THE EXPECTED RESULT OF REACTIVE THE DISCORDANT VITROS AHBC AND VITROS AHBS RESULTS WERE NOT REPORTED FROM THE LABORATORY. ORTHO HAS NOT BEEN MADE AWARE OF ANY ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF TWO MDR¿S FOR THIS EVENT. TWO 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16097139 ↗
Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices
- Elecsys HBsAg II 08814848190Roche Diagnostics GmbH · Test, hepatitis b (b core, be antigen, be antibody, b core igm)93 reports
- ARCHITECT 4P53-25Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall16 reports
- VITROS 6801925Ortho Clinical Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall15 reports
- ARCHITECT 1L82-27Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall4 reports
- VITROS 6801926Ortho Clinical Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)1 recall0 reports
- ARCHITECT 4P54-25Abbott Ireland Diagnostics · Test, hepatitis b (b core, be antigen, be antibody, b core igm)27 reports
- ADVIA Centaur® HBsII 10492138Siemens Healthcare Diagnostics Inc. · Test, Hepatitis B (B Core, Be Antigen, Be Antibody, B Core Igm)18 reports
- Atellica™ IM HBsII 10995604Siemens Healthcare Diagnostics Inc. · Test, hepatitis b (b core, be antigen, be antibody, b core igm)12 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.