Skip to content

VITROS 6801425Recalls & reported problems

Ortho Clinical Diagnostics · VITROS Immunodiagnostic Products Anti-HBc IgM Reagent Pack · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10758750002795), received between Mar 22, 2023 and Mar 22, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “False Positive Result” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • False Positive Result1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 22, 2023False Positive Result; False Positive Result

    A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT DISCORDANT REACTIVE VITROS ANTI-HBC IGM (AHBCM) AND ANTI-HAV IGM (AHAVM) RESULTS OBTAINED FROM TWO SAMPLES COLLECTED FROM THE SAME PATIENT COMPARED TO ANTI-HBC TOTAL AND ANTI-HAV TOTAL NEGATIVE RESULTS WERE OBTAINED FROM THE SAME PATIENT SAMPLES WHEN TESTED ON A VITROS 3600 IMMUNODIAGNOSTIC SYSTEM. SAMPLE 1 VITROS AHAVM RESULT OF 10.4 S/C VS. A VITROS AHAV TOTAL NEGATIVE RESULT. SAMPLE 1 VITROS AHBCM RESULT OF 2.87 S/C VS. A VITROS AHBC TOTAL NEGATIVE RESULT. SAMPLE 2 VITROS AHAVM RESULT OF 10.4 S/C VS. A VITROS AHAV TOTAL NEGATIVE RESULT. SAMPLE 2 VITROS AHBCM RESULT OF 2.91 S/C VS. A VITROS AHBC TOTAL NEGATIVE RESULT. BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE VITROS AHBCM AND AHAVM REACTIVE RESULTS WERE REPORTED OUTSIDE THE LABORATORY BASED ON THE CONCORDANT RESULTS OBTAINED AT THE MAYO CLINIC. THERE WAS NO REPORTED ALLEGATION OF HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF TWO MDR¿S FOR THIS EVENT. TWO 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS TWO DEVICES WERE IN

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16591918 ↗

FDA clearance and classification

The FDA regulates product code LOM (test, hepatitis b (b core, be antigen, be antibody, b core igm)) under 21 CFR 866.3172 as a Class II device. 21 CFR 866.3172 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: In vitro diagnostic device intended for aid in the diagnosis of chronic or acute HBV infection. HBV surface antigen (HbsAg) is also used for screening of HBV infection in pregnant women to identify neonates who are at risk of acquiring hepatitis B during perinatal period.

Common questions about the VITROS 6801425

Has the VITROS 6801425 been recalled?

No FDA recall record in our data names this exact device identifier (DI 10758750002795), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the VITROS 6801425?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Mar 22, 2023. That report lists device problem “False Positive Result” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the VITROS 6801425?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the VITROS 6801425, and is it still sold?

The FDA's device database lists Ortho Clinical Diagnostics as the company responsible for it (brand VITROS, model 6801425). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices

All Test, hepatitis b (b core, be antigen, be antibody, b core igm) devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page