VITROS 8298671Recalls & reported problems
Ortho-Clinical Diagnostics, Inc. · VITROS Chemistry Products PHYT Slides · Class 2
All models of VITROS (71)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Ortho-Clinical Diagnostics VITROS PHYT slide is used on VITROS analyzers to measure phenytoin (PHYT), an anti-seizure drug, in patient samples. The 53 reports mostly describe test results coming out lower or higher than expected, and in some cases results that could not be reproduced on repeat testing. All53 describe malfunctions, with no deaths and no injuries. One recall, started in June 2017 and since closed, concerned the possibility of biased results when the slides were run with citrate plasma samples.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 48
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 48 adverse event reports that list this exact device (DI 10758750004690), received between Apr 25, 2023 and Sep 15, 2025 — classified as 48 malfunctions.
The most frequently reported specific device problem is “Low Test Results” (38 reports).
1 FDA recall record name this device's identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 12, 2017Class IITerminated
The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.
Z-0273-2018 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Low Test Results38
- High Test Results30
- Non Reproducible Results2
Patient problems
- No Clinical Signs, Symptoms or Conditions48
Recent adverse event reports
The 25 most recent of 48 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A LOWER THAN EXPECTED, NON-REPRODUCIBLE PHENYTOIN (PHYT) RESULT FROM A SINGLE PATIENT SAMPLE USING VITROS PHYT SLIDE LOT 2627-0194-4131 PROCESSED ON A VITROS XT 7600 INTEGRATED SYSTEM. PATIENT SAMPLE VITROS PHYT RESULT OF 56.7 UMOL/L VS. THE EXPECTED VITROS PHYT RESULT OF 85.0 UMOL/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE LOWER-THAN-EXPECTED VITROS PHYT RESULT WAS NOT REPORTED OUTSIDE OF THE LABORATORY AND THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23057000 ↗MalfunctionJan 29, 2025Low Test Results; Non Reproducible Results; Low Test Results; Non Reproducible Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT A NON-REPRODUCIBLE, LOWER THAN EXPECTED PHENYTOIN (PHYT) RESULT FROM A SINGLE PATIENT SAMPLE USING VITROS PHYT SLIDE LOT 2627-0190-6032 PROCESSED ON A VITROS XT 7600 INTEGRATED SYSTEM. PATIENT SAMPLE VITROS PHYT RESULT OF 76.13 UMOL/L VS. THE EXPECTED VITROS PHYT RESULT OF 143.17 UMOL/L BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED. THE NON-REPRODUCIBLE, LOWER THAN EXPECTED PHYT RESULT WAS REPORTED OUTSIDE OF THE LABORATORY; HOWEVER, NO TREATMENT WAS GIVEN, CHANGED, OR WITHHELD BASED ON THE REPORTED RESULT. A CORRECTED REPORT WAS ISSUED TO THE PHYSICIAN AND THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBER (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21264581 ↗MalfunctionApr 22, 2024High Test Results; High Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING VITROS TDM PERFORMANCE VERIFIERS AND NON-VITROS BIORAD QUALITY CONTROL FLUIDS ON A VITROS XT7600 SYSTEM. VITROS PHYT SLIDE LOT 2627-0187-7862:(B)(6) 2024: VITROS TDM PV3 RESULT 32.6 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. VITROS PHYT SLIDE LOT 2627-0187-0219: (B)(6) 2024 VITROS PHYT RESULTS 18.98 AND 18.86 UG/ML VERSUS THE BASELINE MEAN OF 12.23 UG/ML. (B)(6) 2024 VITROS TDM PV3 RESULT OF 31.8 UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML.(B)(6) 2024: VITROS TDM PV3 RESULT 31.4 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. (B)(6) 2024: VITROS TDM PV3 RESULTS >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, 31.9, 34.8, 34.3, 33.5, AND 34.3 UG/ML UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. (B)(6) 2024: VITROS BIORAD MULTIQUAL UNASSAYED QC LOT 92963 VITROS PHYT RESULTS 32.0, 31.4, 31.6, 32.0, 31.7, 31.1, 32.0, 32.1, 30.8, 31.7, 32.5, 30.6, 31.7, 32.1, 31.1, 31.2, AND 31.6 UG/ML VERSUS THE CUSTOMER BASELINE MEAN 25.0 UG/ML. BIA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19159945 ↗MalfunctionApr 22, 2024High Test Results; High Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING VITROS TDM PERFORMANCE VERIFIERS AND NON-VITROS BIORAD QUALITY CONTROL FLUIDS ON A VITROS XT7600 SYSTEM. VITROS PHYT SLIDE LOT 2627-0187-7862: (B)(6) 2024: VITROS TDM PV3 RESULT 32.6 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. VITROS PHYT SLIDE LOT 2627-0187-0219: (B)(6) 2024 VITROS PHYT RESULTS 18.98 AND 18.86 UG/ML VERSUS THE BASELINE MEAN OF 12.23 UG/ML. (B)(6) 2024 VITROS TDM PV3 RESULT OF 31.8 UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. (B)(6) 2024: VITROS TDM PV3 RESULT 31.4 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. (B)(6) 2024: VITROS TDM PV3 RESULTS >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, 31.9, 34.8, 34.3, 33.5, AND 34.3 UG/ML UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. (B)(6) 2024: VITROS BIORAD MULTIQUAL UNASSAYED QC LOT 92963 VITROS PHYT RESULTS 32.0, 31.4, 31.6, 32.0, 31.7, 31.1, 32.0, 32.1, 30.8, 31.7, 32.5, 30.6, 31.7, 32.1, 31.1, 31.2, AND 31.6 UG/ML VERSUS THE CUSTOMER BASELINE MEAN 25.0 UG/ML. BI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19159840 ↗MalfunctionApr 22, 2024High Test Results; High Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING VITROS TDM PERFORMANCE VERIFIERS AND NON-VITROS BIORAD QUALITY CONTROL FLUIDS ON A VITROS XT7600 SYSTEM. VITROS PHYT SLIDE LOT 2627-0187-7862: (B)(6) 2024: VITROS TDM PV3 RESULT 32.6 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. VITROS PHYT SLIDE LOT 2627-0187-0219: (B)(6) 2024 VITROS PHYT RESULTS 18.98 AND 18.86 UG/ML VERSUS THE BASELINE MEAN OF 12.23 UG/ML. (B)(6) 2024 VITROS TDM PV3 RESULT OF 31.8 UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. (B)(6) 2024: VITROS TDM PV3 RESULT 31.4 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. (B)(6) 2024: VITROS TDM PV3 RESULTS >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, 31.9, 34.8, 34.3, 33.5, AND 34.3 UG/ML UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. (B)(6) 2024: VITROS BIORAD MULTIQUAL UNASSAYED QC LOT 92963 VITROS PHYT RESULTS 32.0, 31.4, 31.6, 32.0, 31.7, 31.1, 32.0, 32.1, 30.8, 31.7, 32.5, 30.6, 31.7, 32.1, 31.1, 31.2, AND 31.6 UG/ML VERSUS THE CUSTOMER BASELINE MEAN 25.0 UG/ML. BI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19159815 ↗MalfunctionApr 22, 2024High Test Results; High Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING VITROS TDM PERFORMANCE VERIFIERS AND NON-VITROS BIORAD QUALITY CONTROL FLUIDS ON A VITROS XT7600 SYSTEM. VITROS PHYT SLIDE LOT 2627-0187-7862: 4 MARCH 2024: VITROS TDM PV3 RESULT 32.6 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. VITROS PHYT SLIDE LOT 2627-0187-0219: 23 MARCH 2024 VITROS PHYT RESULTS 18.98 AND 18.86 UG/ML VERSUS THE BASELINE MEAN OF 12.23 UG/ML. 24 MARCH 2024 VITROS TDM PV3 RESULT OF 31.8 UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. 25 MARCH 2024: VITROS TDM PV3 RESULT 31.4 UG/ML VERSUS THE MIDPOINT OF THE ROM 25.9 UG/ML. 25 MARCH 2024: VITROS TDM PV3 RESULTS >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, >40, 31.9, 34.8, 34.3, 33.5, AND 34.3 UG/ML UG/ML VERSUS THE MIDPOINT OF THE RANGE OF MEANS 25.9 UG/ML. 01 APRIL 2024: VITROS BIORAD MULTIQUAL UNASSAYED QC LOT 92963 VITROS PHYT RESULTS 32.0, 31.4, 31.6, 32.0, 31.7, 31.1, 32.0, 32.1, 30.8, 31.7, 32.5, 30.6, 31.7, 32.1, 31.1, 31.2, AND 31.6 UG/ML VERSUS THE CUSTOMER BASELINE MEAN 25.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19159786 ↗MalfunctionApr 10, 2024Low Test Results; Low Test Results
THE INVESTIGATION CONCLUDES THAT A DISCORDANT LOW VITROS PHENYTOIN (PHYT) RESULT WAS OBTAINED WHEN PROCESSING A COLLEGE OF AMERICAN PATHOLOGISTS (CAP) PROFICIENCY FLUIDS USING TWO DIFFERENT VITROS PHYT SLIDE LOTS ON A VITROS XT 7600 SYSTEM. CAP WAS STILL COMPILING DATA AND WAS UNABLE TO PROVIDE THE EXPECTED RESULT FOR THIS PROFICIENCY FLUID. THEREFORE, THE EXPECTED RESULT WAS THE PHYT RESULT OBTAINED FROM THE IN USE PHYT SLIDE LOT 2626-0186-6657. VITROS PHYT LOT 2627-0187-9131 RESULT OF 24.2 UG/ML VERSUS THE VITROS PHYT LOT 2626-0186-6657 RESULT OF 31.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE DISCORDANT LOW VITROS PHYT RESULT WAS OBTAINED WHEN PROCESSING PROFICIENCY FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19080171 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER THREE OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4), (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369378 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4), (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369373 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER SIX OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369371 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER SEVEN OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369365 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FIVE OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369358 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FOUR OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPORTABILITY ASSESSMENT (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369350 ↗MalfunctionDec 20, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED WHEN PROCESSING BIORAD IMMUNOASSAY PLUS QC LEVEL 3, LOT 85310 TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. VITROS PHYT SLIDES, LOT 2626-0186-4393 VITROS PHYT RESULTS 27.9, 29.5, 29.6, 12.8, 13.4, 8.8, AND 14.1 UG/ML VERSUS THE BASELINE MEAN 23.2 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE HIGHER AND LOWER THAN EXPECTED VITROS PHYT RESULTS WERE OBTAINED WHEN PROCESSING QUALITY CONTROL FLUIDS. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ONE OF SEVEN MDR¿S FOR THIS EVENT. SEVEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS SEVEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPORTABILITY ASSESSMENT 604381.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18369329 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER THIRTEEN OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187416 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER NINE OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REP
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187390 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER ELEVEN OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND R
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187389 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TEN OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187374 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER SIX OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187304 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER EIGHT OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187302 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER SEVEN OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187290 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FOUR OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187285 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER FIVE OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REP
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187275 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER TWO OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND REPO
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187245 ↗MalfunctionNov 21, 2023High Test Results; Low Test Results; High Test Results; Low Test Results
A CUSTOMER CONTACTED THE ORTHO CLINICAL DIAGNOSTICS (ORTHO) TECHNICAL SOLUTION CENTER (TSC) TO REPORT LOWER AND HIGHER THAN EXPECTED VITROS PHENYTOIN (PHYT) RESULTS WERE OBTAINED FROM A SINGLE LEVEL NON-VITROS MAS QUALITY CONTROLS, LOT OCR25113, USING TWO LOTS OF VITROS PHYT TESTED ON A VITROS XT7600 INTEGRATED SYSTEM. PHYT LOT 2626-0186-5242 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 16.09, 19.35, 18.18, 20.54, 35.40, 19.78, 20.83, 18.75, 18.43, 15.22, AND 19.06 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML PHYT LOT 2625-0185-2782 MAS LEVEL 3 LOT 25113 FLUID RESULTS OF 35.03 AND 35.25 UG/ML VERSUS THE MAS PEER MEAN OF 28.6 UG/ML BIASED RESULTS OF THE MAGNITUDE AND DIRECTION OBSERVED MAY LEAD TO INAPPROPRIATE PHYSICIAN ACTION IF THEY WERE TO OCCUR UNDETECTED ON PATIENT SAMPLES. THE LOWER AND HIGHER THAN EXPECTED RESULTS WERE OBTAINED FROM A NON-PATIENT QUALITY CONTROL FLUID. THERE WAS NO ALLEGATION OF PATIENT HARM AS A RESULT OF THIS EVENT. THIS REPORT IS NUMBER THREE OF THIRTEEN MDR¿S FOR THIS EVENT. THIRTEEN 3500A FORMS ARE BEING SUBMITTED FOR THIS EVENT AS THIRTEEN DEVICES WERE INVOLVED. THIS REPORT CORRESPONDS TO ORTHO CLINICAL DIAGNOSTICS INC (ORTHO) COMPLAINT NUMBERS (B)(4) AND RE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18187179 ↗
FDA clearance and classification
The FDA regulates product code DIP (enzyme immunoassay, diphenylhydantoin) under 21 CFR 862.3350 as a Class II device. 21 CFR 862.3350 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the VITROS 8298671
Has the VITROS 8298671 been recalled?
Yes. 1 FDA recall record names this device's identifier, and the FDA has closed (terminated) it. The recall, Z-0273-2018 (Class II), began on Jun 12, 2017: The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the VITROS 8298671?
The FDA's MAUDE database lists 48 adverse event reports for this device, received between Apr 25, 2023 and Sep 15, 2025. In those reports, the most common specific device problems are “Low Test Results” (38), “High Test Results” (30) and “Non Reproducible Results” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the VITROS 8298671?
None of the reports are classified as deaths, 0 as injuries and 48 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the VITROS 8298671, and is it still sold?
The FDA's device database lists Ortho-Clinical Diagnostics, Inc. as the company responsible for it (brand VITROS, model 8298671). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other ENZYME IMMUNOASSAY, DIPHENYLHYDANTOIN devices
- ONLINE TDM Phenytoin 08445567190Roche Diagnostics GmbH · ENZYME IMMUNOASSAY, DIPHENYLHYDANTOIN2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page