Skip to content

VITROS recallZ-0273-2018 · Ortho-Clinical Diagnostics · initiated Jun 12, 2017

Terminated

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm identified the potential for biased results to be generated when processing VITROS PHYT Slide product with citrate plasma specimen samples. The current Instructions for Use (IFU) and Assay Summary Chart state serum, heparin plasma, and citrate plasma as acceptable specimen types for use with the PHYT assay.

FDA root cause category: Device Design

What the company is doing

Ortho-Clinical Diagnostic sent an Urgent Product Correction Letter dated June 12, 2017, sent via FedEx overnight courier and/ or ORTHO PLUS e-Communications and/ or US Postal Service Priority Mail (for PO Boxes only) to all customers who have a VITROS analyzer that can process PHYT Slides to inform them that negatively biased results may occur when processing citrate plasma samples using VITROS Chemistry Products PHYT Slides. Customers were instructed to immediately discontinue using citrate plasma samples to process VITROS PHYT Slides. They were also informed that it is still acceptable to use serum and heparin plasma samples to process VITROS PHYT Slides. Foreign affiliates were informed by e-mail on June 12, 2017 of the issue and instructed to notify their consignees of the issue and the appropriate actions to take. For further questions, please call (908) 218-8776.

Recall details

Recall number
Z-0273-2018
Status
Terminated
Initiated by the company
Jun 12, 2017
Recalling firm
Ortho-Clinical Diagnostics
Quantity
US; 4680; Foreign: 1286
Product code
DIP
FDA submissions
K081543, K941142
Distribution
Worldwide Distribution - US Distribution including Puerto Rico and the following countries: Bermuda, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, Russia and Venezuela.

Product description

VITROS Chemistry Products PHYT Slides, REF/CATALOG 829 8671, Rx ONLY, IVD; Ortho-Clinical Diagnostics, Inc. For in vitro diagnostic use only. VITROS Chemistry Products PHYT Slides quantitatively measure phenytoin (PHYT) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Code information (models, lots, serial numbers)

Unique Identifier Number ( UDI) 10758750004690; This issue affects expired, current (in date) and future product GENs.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page